Operational Guide
53
The following table lists the temperature dependence of Hanna EC calibration
standards. edge™ uses these values during calibration and their temperature
coefficients.
ºC
ºF
HI7030
HI8030
(μS/cm)
HI7031
HI8031
(μS/cm)
HI7033
HI8033
(μS/cm)
HI7034
HI8034
(μS/cm)
HI7035
HI8035
(μS/cm)
HI7039
HI8039
(μS/cm)
0
32
7150
776
64
48300
65400
2760
5
41
8220
896
65
53500
74100
3180
10
50
9330
1020
67
59600
83200
3615
15
59
10480
1147
68
65400
92500
4063
16
60.8
10720
1173
70
67200
94400
4155
17
62.6
10950
1199
71
68500
96300
4245
18
64.4
11190
1225
73
69800
98200
4337
19
66.2
11430
1251
74
71300
100200
4429
20
68
11670
1278
76
72400
102100
4523
21
69.8
11910
1305
78
74000
104000
4617
22
71.6
12150
1332
79
75200
105900
4711
23
73.4
12390
1359
81
76500
107900
4805
24
75.2
12640
1386
82
78300
109800
4902
25
77
12880
1413
84
80000
111800
5000
26
78.8
13130
1440
86
81300
113800
5096
27
80.6
13370
1467
87
83000
115700
5190
28
82.4
13620
1494
89
84900
117700
5286
29
84.2
13870
1521
90
86300
119700
5383
30
86
14120
1548
92
88200
121800
5479
31
87.8
14370
1575
94
90000
123900
5575
EC/TDS
CALIBRATION
Содержание Edge
Страница 1: ...Instruction Manual Multiparameter Benchtop Meter ...
Страница 92: ...Warranty 92 Notes ...
Страница 93: ...Warranty 93 Notes ...
Страница 94: ...Warranty 94 Notes ...
Страница 95: ...Warranty 95 Notes ...