35
• Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
• Remove the cuvet.
• Add the content of one packet
of HI 93702-0 Copper
Reagent. Replace the cap and
shake gently for 15 seconds.
• Add the content of one packet
of HI 93702T-0 Copper Total
Reagent. Replace the cap and
shake vigorously for 15
seconds.
• Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press TIMER and the display will show the countdown prior to
measurement. Alternatively, wait for 45 seconds and press READ
DIRECT. In both cases “SIP” will blink during measurement.
• The instrument directly displays concentration in
mg/L of total
copper
on the Liquid Crystal Display.
Total Copper