6. Troubleshooting
6.2 Error messages
Error
code
Problem
Cause
Recommended Action
E13
Low voltage
Error
•Motor applied voltage decreases
abnormally during operation.
• Switch power off.
E15
RPM
Sensing
error
• Rpm sensor failure
• Restart the device.
E17
communica
tion error
• Occurs when
communication between Main and
io board or between display and IO
board is broken
• Restart the device.
E20
Lid_1_sens
or error
• Lid_1_sensor failure
• Restart the device.
E21
Motor_C_s
ensor error
• Motor Close sensor failure
• Restart the device.
E22
Motor-
O_sensor
error
• Motor Open sensor failure
• Restart the device.
17 l www.ihanil.com
Please contact us if any other error codes occur or if you are not satisfied with the
above method.