4. Operation
4.2 Opening and closing the Door
•The door can only be opened if the centrifuge is switched on.
•Do not reach with your fingers between the housing and door.
•Close the door completely before operation.
•
Opening the door
Press the Door ( ) button.
•
Closing the door
Press the edge of the lid with a light force.
4.3 Loading tubes
• Always use the same type of tube.
• Tubes should be loaded symmetrically.
• Do not exceed the maximum rated speed of the tube.
• Same volume of sample should be put on tubes.
• Check symmetric loading by balncing tubes with scales.
①
Check the maximum load for each tube.
②
Put tubes into rotor holes.
③
Tubes located opposite each other must be the same type and contain the same quantity.