3. Installation

3.1 Packing Inspection

3.2 Installation

▶ You can get contact details on the packing box or manual.

• Check packing conditions carefully, before unpacking.
• Contact Hanil Scientific Inc. immediately if damages found.
• Check the delivery for completeness.

Installation on hard and flat ground.

• Centrifuge should be installed on a hard and flat place.

• If the centrifuge is installed in an inclined place, the shaft may be bent

due to the weight of the rotor.

Good ventailation.

• For air circulation and safety, maintain a 30cm“clearance envelope”

around the centrifuge while the rotor is spinning.

Constant temperature/humidity

• Centrifuge equipped with the sensitive electronic software which is

fragile with humidity and temperature.

• Must avoid direct ray or heater and be put in the ambience of controlled

temp.and air.

Avoid the corrosive gas

• Install the centrifuge in a place where corrosive gas is not generated.

• Sulfur dioxide gas and chlorine gas may cause corrosion.

Leveling

• The shaft should be put exactly vertical on the hotizontaliy flat ground by

the leveling tool.

Содержание M15R

Страница 1: ...Operation Manual M15R Micro Centifuge Where Purchased Date of Purchase...

Страница 2: ...fic Inc If you have any questions contact our Technical Support Center 82 2 3452 8966 techsupport ihanil com www ihanil com The appearance or specifications of the device is subject to partial change...

Страница 3: ...ng Closing the Door 9 3 5 Installing Removing the Rotor 10 3 6 Rotor ID 11 3 7 Loading Tubes 12 4 Operation 13 4 1 Control Panel 13 4 2 Setting RPM RCF 14 4 3 Setting Run Time 14 4 4 Setting Temperatu...

Страница 4: ...be deviced 7 The balancing work of samples should done in advance before operation 8 To ensure safe use of the device do not expose the device to strong acids strong bases cesium salt or alkaline dete...

Страница 5: ...lug the powercord from outlet and contact your place of purchase Unusual noises or smell from the equipment Damage or wear of a power cord Breakdown of circuit breaker fuse or safety device If you spi...

Страница 6: ...22 2 Product Description 2 1 Structure Door RPM measuring window Chamber Ventilation slots Display control panel Power Socket Drain cap Power Swithch Device Power cable Operating manual T wrench 2 2...

Страница 7: ...ity 10 x 5 0 mL ACC DEC steps 9 10 steps Acc time to max speed 15 sec Dec time from max speed 15 sec Temperature range 10 to 40 Fast cool function Yes Program memory 10 Noise level 60 dB Imbalance cut...

Страница 8: ...t due to the weight of the rotor Good ventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidity Cen...

Страница 9: ...vice Use sockets with a protective earth conductor and provided power cord If the door is not closed the device does not operate 1 Connect the power cable to the power socket on the right side of the...

Страница 10: ...otor When installing the rotor the temperature of the rotor and motor shaft should be at room temperature 10 30 2 Insert the rotor you want to fit into the motor shaft 3 Hold the rotor with one hand a...

Страница 11: ...be centrifuged within the maximum speed 1 Press the Setup Enter button 2 Press the left or right button to change to the rotor ID setting mode Rotor Id Setting mode Rotor flashes on the display 3 Pres...

Страница 12: ...loaded symmetrically Do not exceed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales 1 Check the maximum load for e...

Страница 13: ...e cooling the rotor and chamber Left Left button is used to select parameter Right Right button is used to select parameter Setup Enter Setup Enter button can be pressed to go into setup mode or compl...

Страница 14: ...hes RCF setting mode RCF xg display flashes 3 Press Up or Down button until the desired RPM RCF is displayed 4 Press the Setup Enter button 1 Press the Setup Enter button 2 Press the Left or Right but...

Страница 15: ...e display 00 flashes 3 Press Up or Down button until the desired temperature is displayed 1 Set the desired Temperature Refer to 4 4 Setting Temperature 2 Press the Down button more than 2sec A rotor...

Страница 16: ...splay flashes 0 Decel rate setting mode Decel display flashes 0 3 Press the Setup Enter button 4 6 Setting Acceleration Deceleration Start 1 Check all parameters are correct 2 Press the Start Stop Pul...

Страница 17: ...F Run time Accel Decel rate 2 Press the Left of Right button until Prog display flashes 0 You can save 10 0 9 programs 3 Press Up or Down button until the desired program number is displayed 4 Press t...

Страница 18: ...ive chemicals on the device such as alchol benzene acetone or phenol 3 Make sure do not scratch the surface of deivice when cleaning it Do not use a metal sponge If the device is rusted remove it with...

Страница 19: ...d Check the lid is closed completely Door cannot be opened Press the Door button Check the power connection Door cannot be closed Remove the dirt at the door latch and close the lid Check the lid latc...

Страница 20: ...System error Contact the Technical support team E10 Inner Temperature Over OR Inner temperature sensor error Occurs when the set temperature is not reached Inner temperature sensor error Make sure th...

Страница 21: ...ce door cannot be opened you can activte the emergency open manually 1 Wait for rotor to stop before activating the emergency open 2 Insert the T wrench into the emergency open hole and turn it clockw...

Страница 22: ...0 2mL strips 6 5 x 17 55 5 12 500 9 852 AE 2 0 24 Hole angle 45 Max Capacity 24 x 1 5 2 0 mL Size x H 172 x 38 5 mm Max height for tube fit 56 mm Incl a lid A2 0MP 24 incl plastic lid 1 5 2 0 mL 11 x...

Страница 23: ......

Страница 24: ...RECYCLABLE 16 Arayukro Gimpo 10136 Rep of Korea Service Inquiries 82 2 3452 8966 techsupport ihanil com www ihanil com Doc No OMM9PDEN2002...

Отзывы: