Connect the device to voltage
sources, which correspond to the
electrical requirements on the label
attached to the device.
Check the wall outlet is earth-
grounded, surge-protected and
can supply enough power.
Connect the power cable to the
distribution box, or install the plug
and socket that are packed with
the device.
Switch off the device after use.
Unplug the power plug before
cleaning or unused for a long
period of time.
Check the Line Voltage
Grounded Outlet
Power Connection
Switch Off the Device
Quick Guide
Component R12 / R6
Cat. No. : CP-R12 / CP-R6
Allow a safety clearance of
30 cm around the centrifuge
during operation.
Rotors
Spread the grease lightly
on the bucket where it
contacts the rotor and the
pivot.
Remove any dust regularly
from the ventilation slits on
the left side of the device.
There is a drain hose at the
bottom of the device.
The condensation water is
automatically drained through
the hose. Check and remove it
regularly.
Component R6
Component R12
30 cm
30 cm
30 cm
Emergency Lid Open
If the centrifuge lid cannot be opened,
insert the provided T-wrench into the
emergency release hole and turn it
clockwise. (Before activating emergency
open, switch off the device and wait for
the rotor to stop.)
Wipe up the moisture in the chamber with a
dry cloth after use. Keep the lid open when
not in use.
Clean the chamber and rotor by wiping
with a cloth soaked in mild detergent and
dry completely.
Keep the Chamber Dry
Cleaning
Check the rotor, buckets and tubes are free of
scratchand crack before operation.
Check the Accessories for Abnormalities
Load tubes symmetrically into the rotor
with identical amount.
Load Tubes Symmetrically
Place the lid on the wind shield rotor before
centrifugation.
Check the Lid
After relocating the device or loading the rotor,
balance them by using a leveler. If necessary,
level the device by adjusting the ‘Leveling foot’
on the bottom of the device.
Balancing the Centrifuge and Rotor
3.0KVA
(Component R12)
5.0KVA
(Component R6)
220~240V 50/60Hz
Maintenance
Electrical Safety
Cautions for Use
Rev.1
Proudly made by Jiyoung Shim, Hyewon Jeong
Order & Inquiry l +82-2-3472-0727 l [email protected] l www.ihanil.com
Service l +82-2-3452-8966 l [email protected]
Before Operation
After Operation
Maintenance of
Swing-Out Rotor
Remove Dust
Empty the Condensation
Water
Allow a Safety Clearance
of 30 cm
WS1000-6
WS2400-6
6 x 1,000 mL / Max. 5k
6 x 2,400 mL / Max. 4.8k