background image

9

A/S Inquiry : +82-2-3452-8966 / [email protected]

USER MANUAL

MINI CENTRIFUGE

  Do not open the lid until operation is completely stopped as 

it will decelerate immediately when the lid is opened during 
operation.

- Press the lid open button.

3. Opening the lid

2. Starting/Stopping the centrifugation

Short Mode (S)

- When the lid is closed, centrifutaion starts.

Continuous Mode (C)

- Operataion stops after 15 min. of operation.

5-2. Operation

1. Installing the rotor

- Load the rotor vertically onto the motor shaft.
- Put the cap nut on the center of the rotor and
   turn the wrench clockwise untill it secured.

Содержание ARA Smart 6

Страница 1: ...ARA Smart 6...

Страница 2: ...ARA Smart 6 Research Use Only Copyright 2019 Hanil Scientific Inc All rights reserved Contact Us If you have any questions contact Hanil Scientific Inc techsupport ihanil com info ihanil com...

Страница 3: ...2 Intended Use 5 3 Product overview 6 4 Technical specifications 7 5 Installaion Operation 8 5 1 Power connection 8 5 2 Operation 9 6 Troubleshooting 10 CONTENTS The appearance or specifications of t...

Страница 4: ...2 According to IEC61010 2 020 maintain a 30cm clearance envelope around the device while the rotor is spinning 3 Do not operate the device in areas where work is completed with explosice substances 4...

Страница 5: ...pill liquid shut off the power supply and unplug the powercord from the outlet and contact Hanil Technical Support Team 8 Do not block vents 9 Do not drop or subject the device to severe impacts 10 Ca...

Страница 6: ...erview Lid open button DC adaptor Connector Continuous Normal Switch Operation mode switch Lid Closing the lid Start the centrifugation Opening the lid Stop the centrifugation Short Mode S Operate aft...

Страница 7: ...Tube capacity 6 x 1 5 2 0 mL 2 x 8 tube PCR Strips Accel 4 sec Decel 2 sec Weight 0 65 kg Power supply DC 12V 2A AC100 240V 50 60Hz Dimension 150W x 150D x 103H mm Cat No AC Smart 6 Delivery package A...

Страница 8: ...he delivery for completeness The device should be installed on flat surface to maintain level Do not bend or pull the AC DC adaptor Keep this manual near the device at all times 1 Connect the power ca...

Страница 9: ...during operation Press the lid open button 3 Opening the lid 2 Starting Stopping the centrifugation Short Mode S When the lid is closed centrifutaion starts Continuous Mode C Operataion stops after 1...

Страница 10: ...ed 1 Power failure Check the mains connection 2 AC adaptor defect Check the output DC volt of the adapter using a multimeter 3 Lid sensor failure Contact Hanil Technical Support Team 4 If the beep doe...

Страница 11: ......

Страница 12: ...im ara com OMM7PDEN1901 C...