Tuning POP-UPs
Output Tuning is Recommended (>10% variance)
If the Output tuning is greater than >10% variance, but
less than < 20%, an
“Output tuning is recommended” caution will pop –up
on the screen.
User has the option to select IGNORE to continue
without tuning, or TUNE to be taken to the Device
Output Tuning screen.
When IGNORE is selected, user is brought back to previous screen, and allowed to continue treatment.
If TUNE is selected, user is taken to Device Output
Tuning screen. Press the START button below.
Gradually turn the Tuning Knob on the applicator head
to display the most amount of green and the least
amount of yellow on the bar below. Press DONE when
complete.
Upon completion of tuning, user is prompted to select
DONE. When DONE is pressed, User is returned to the
previous screen viewed before Warning was generated.
User is brought back to previous screen they were in before Caution pop-up warning was prompted.
Output Tuning is Required (>20% variance)
If the Output tuning is greater than >20%. An “Output
tuning is required” will pop-up on the screen, It is
required to tune the device to continue treatment.
Please select TUNE to be taken to the Device Output
Tuning screen.
Press the Start button below. Gradually turn the Tuning
Knob on the applicator head to display the most amount
of green and the least amount of yellow on the bar
below. Press DONE when complete.
Upon completion of tuning, user is prompted to select DONE. When DONE is pressed, User is returned to the
previous screen viewed before Caution pop –up warning was generated.