Lift type:
Contract no.:
s/n:
Name:
Version:
Address:
Prod. year:
Type of inspection:
First-time nspection
Periodic inspection
Verification inspection
11.1 Inspection check list RiseAtlas
In accordance with ISO 10535:2006 Annex A-Periodic inspection
Date
of
inspection:
Test-loaded
with:
Sign installer:
Sign Customer:
A. Ocular inspection of
D. Load and speed testing of (performance)
0
Wear and damage.
0
Test load
1
Strap
1
Speed up/down
2
Cables and connections
3
Reported errors
E. Overhead systems
4
Correct labelling
1
Wear or damage
5
Documentation
2
Screws, bolts
6
Environment
3
Functionality
4
End stops
B. Functionality testing of
5
Trolley
0
Start and stop
6
Labelling
1
Charge
7
Installation Manual
2
Hand Control
8
Operation and charging routines
3
Up/down function
9
Photograph
4
Movement
10
Cleaning
5
Limit switch
11
Accessories
12
Weight testing procedure
C. Safety testing of
0
Emergency stop
1
Mechanical and electrical.
2
Connection points
33
S E R V I C E M A N U A L
System
RoMedic
TM
© Handicare AB • www.handicare.com