Battery information on the display 

The battery discharging will be shown in four stages:

Battery state 1

: The battery is ok, no need for charging (100 - 50 %).

Battery state 2

: Battery needs charging. (50 - 25 %)

Battery state 3:

 Battery needs charging. (Less than 25 %) A tone sounds when a button is pressed in 

this battery state.

Battery state 4

: The battery needs charging. (17V or lower) At this stage some of the functionality of the 

lift is lost.At this battery stage it is only possible to drive the lift arm down. Furthermore an audio signal 
will sound when a control button is activated. The symbol will switch between the two pictures for 10 
seconds.

The battery symbol is shown when the control box is active until power down (2 minutes after use). 
It is not possible to use other batttery types than BAJ1/BAJ2.

The battery level is measured via voltage. This means that it is possible to experience e.g. that the 
battery switches from state 1 to state 2 and back to state 1.

        Charging batteries

A tone sounding when using the lift indicates that the battery need recharging.
Charge the lift after use to ensure that the battery is always fully charged.
Lock the castor wheels when charging the battery. 

1. Connect the charging cable to a power outlet.
2. Check to ensure that the lamps on the control box light up. The green LED lamp indicates that the 
charger is receiving power and the yellow LED lamp indicates that the battery is charging. 
3. Charging stops automatically when the battery is fully charged.

Wall-mounted charger

1. Remove the battery pack from the lift and place it in the wall-mounted charger. 
2. Check to ensure that the LED lamp on the front of the charger lights up.

NOTE!

Before the lift is used for the first time, it must be charged for at least 4 hours.
For maximum battery life, charge batteries regularly. We recommend daily charging when the lift is 
used daily.
The emergency stop must be deactivated during charging.

8

M A N U A L System

RoMedic

TM

Содержание SystemRoMedic Carina350EE

Страница 1: ...e s t a t e d g e n e r a l t o l e r a n c e s a c c o r d i n g t o I S O 2 7 6 8 m h c l a r e k l T h i s d r a w i n g a n d a n y i n f o r m a t i o n o r d e s c r i p t i v e m a t t e r s e...

Страница 2: ......

Страница 3: ...Table of contents English 4 Svenska 15 Norsk 26 Dansk 37 Suomi 48 Deutsch 59 Nederlands 70 Fran ais 81 Italiano 92 Espa ol 103 Dimensions 114 116 Serial number 117 3 M A N U A L SystemRoMedic TM...

Страница 4: ...trol box 6 Rear castors with brakes 7 Sling bar with safety latches 8 Front castors 9 Emergency lowering manual 10 Motor actuator for lift arm 11 Locking handle 12 Pedal for manual width adjustment EM...

Страница 5: ...e lift arm image 2 7 Check all mounting fixtures and locking handles images 1 and 2 8 Deactivate the emergency stop image 3 and perform a final inspection see Final inspection To collapse the lift fol...

Страница 6: ...on the respective pedals for widening and narrowing the base For EE use the buttons on the hand control Test lift function by lifting a person not a user using an approved sling At the same time chec...

Страница 7: ...g a transfer situation Under no circumstances may the safe working load be exceeded See section on Safe working load Never move the lift by pulling on the actuator The lift must not be used or immerse...

Страница 8: ...ntil power down 2 minutes after use It is not possible to use other batttery types than BAJ1 BAJ2 The battery level is measured via voltage This means that it is possible to experience e g that the ba...

Страница 9: ...the actuator Total work done by the actuator ampere usage times seconds in use Total number of overloads Days since last service Days between services Hand control Raising lowering the lift arm Symbol...

Страница 10: ...Take the lift out of operation and contact your dealer Emergency stop To activate the emergency stop Press the red emergency stop button on the control box Resetting Turn the button in the direction...

Страница 11: ...minium sling bar available in three variants with different width All variants of SlingBar have safety latches which prevent the sling straps from creeping out of the sling bar and all variants are fo...

Страница 12: ...ining phenol or chlorine as this could damage the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should be used Storage and transportation If the li...

Страница 13: ...sures such as reorienting or relocating the equipment system or shielding the location EMC Read user manual May not be discarded in domestic waste The product complies with the requirements of the Med...

Страница 14: ...inutes maximum 5 switching cycles per minute Push 6000N Motor base DC 24V 5 A IP X4 Operationtime 10 at maximum continous operation of 2 min utes maximum 5 cycles per minute Push 3000N Sound level Wit...

Страница 15: ...ina350EE EM 60600009 60600011 All measurement are in mm inch 285 11 2 580 22 8 580 22 8 860 33 8 1121 44 1 125 4 9 530 20 8 1680 66 1 490 19 3 1840 72 4 1020 1120 40 1 44 1290 50 7 114 M A N U A L Sys...

Страница 16: ...h Dimensions Carina350EM EE with low legs 60600012 60600014 285 11 2 520 20 4 580 22 8 860 33 8 1120 44 1 125 4 9 580 22 8 1680 66 1 490 19 3 1840 72 4 1035 1135 40 7 44 6 1290 50 7 13 0 5 61 2 4 115...

Страница 17: ...3 540 21 2 27 5 37 4 41 3 Dimensions Turning radius is 615 mm 24 2 inch The total weight of lift is 33 kg 73 lbs f The weight of the heaviest component is 17 5 kg 38 58 lbs D is movement in forward d...

Страница 18: ...Serial number 117 M A N U A L SystemRoMedic TM...

Страница 19: ...119 M A N U A L SystemRoMedic TM...

Страница 20: ...vironment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the manual The most recent version of all manuals are availab...

Отзывы: