Technical specifications
Version 2008v2
22
2.3
Electrical wiring diagram of the XP
2.4
Error codes of the XP fault table
Check the malfunction/signal with the aid of the following list.
Flash
Description
1
User Fault
Possible stall timeout or user error
2
Battery Fault
Check the batteries & cabling
3
Left Motor Fault
Check the left motor connections & cabling
4
Right Motor Fault
Check the right motor connections & cabling
5
Left Parking Brake Fault
Check the left parking brake connections & cabling
6
Right Parking Brake Fault
Check the right parking brake connections & cabling
7
SHARK Control Unit Fault
Check the SHARK Communications Bus
8
SHARK Power Module Fault
Check the SHARK connections & wiring
9
SHARK Communications Fault
Check the SHARK connections & wiring
10
Unknown Fault
Check all connections & wiring
11
Incompatible Control Unit
Wrong type of Control Unit Connected
2.5
Fully discharged PowerSupport
Normally the PowerSupport starts charging when the charger is connected. If this does not happen, the
PowerSupport may be fully discharged. To activate the PowerSupport external voltage is needed. The motor can
serve as a source for this. Connect the charger and turn the wheel calmly round. Increase speed slowly until the
led lights flashes on the control unit. Because the charger is already connected, the PowerSupport will start
charging. Please note: the PowerSupport must be fully charged.
Please feel free to contact one of our service engineers if you have any further questions.