Leitlinien und Herstellererklärung
Test
Specification
Comment
Störabstrahlung 10 m
Entfernung
IEC/CISPR 16-2-3:2006
Edition 2
EN 55016-2-3:2006
30 dBμ V/m, 30 - 230 MHz
37 dBμ V/m, 230 - 1000 MHz
RF-Elektromagnetische
Felder
EN ISO 10535:2006
EN 60601-1-2:2007 clause 6
3 V/m, 80% AM, 1kHz 80 MHz - 800 MHz
10 V/m, 80% AM, 1 kHz 800 MHz - 2.0 GHz
3 V/m, 80% AM, 1kHz 2.0 GHz - 2.5 GHz.
ESD
EN 60601-1-2:2007 clause 6
IEC 61000-4-2:2008 Edition 2
EN 61000-4-2:2009
± 6 kV Kontaktentladung
± 8 kV Luftentladung
Electrical fast Transienten
(Bursts)
EN 60601-1-2:2007 clause 6
IEC 61000-4-4:2004 +
A1:2010
EN 61000-4-4:2004 +
A1:2010
Signal and control lines: ± 1 kV
DC mains inputs and outputs: ± 2 kV
AC mains inputs and outputs: ± 2 kV
Induced conducted
disturbances
EN 60601-1-2:2007 clause 6
IEC 61000-4-6:2008 Edition 3
EN 61000-4-6:2009
3 Vrms, 80% AM, 1kHz 0.15 - 80 MHz
EMC
Deutsch
24
Содержание Height adjustable tilting commode/shower chair
Страница 47: ...47...