If your motorized, traversing

 

lift has a 

return-to-charge feature (RTC)

pressing & holding the blue and yellow colored directional hand control 
buttons simultaneously for 3-5 seconds (to produce beeping noise) will 
automatically retract the lift’s carry bar and drive the lift along its track 
until it docks at the charger.  
 
C-1000’s is equipped with new 

weight sensitive adaptive return-to-

charge feature, 

which

 

offers added protection during RTC movement. 

The salient features of new “Load Sensitive RTC” are as under 
1)  If user/operator may have accidentally enabled RTC while a patient is 

in the lift , the RTC function will stop automatically by sensing ex-
cess load on the lift during RTC. The overload threshold  is between 
40-70 lbs. 

2)  During RTC , if lift carry bar accidentally gets tangled with obsta-

cles , Furniture , drapery etc., the lift will automatically detect the 
change in the load and if any variation of more than 15-30 Lbs. is de-
tected the lift will stop automatically. 

 

Note:- RTC feature can always be interrupted manually by either 
pressing any buttons on the hand control or by holding the carry bar 
firmly. 

 
Note that the RTC feature has several characteristics that may be altered 
to the user’s preference: 
 
RTC Max. Time may be set anywhere from 60 sec to 240 sec in 60 sec 
increments and represents the maximum time allowed for the lift to travel 
to and dock at the charger before aborting the operation. 

 

RTC Drop Time may be set anywhere from 3 sec to 24 sec in 3 sec incre-
ments and represents the time the lift devotes to lowering the carry bar 
upon docking at the charger. 

 

RTC Speed may be set to 2, 4, 6

 

or 8 and represents the relative speed the 

lift travels at to reach the charger. 

Contact your Service Technician to change these settings. 

Return To Charge (If Equipped) 

Figure 8D - Press & Hold the blue and 
yellow color hand control buttons simul-
taneously to activate the RTC feature. 

Caution: Always use extreme care when moving the traversing beam. Watch out for 

and avoid any obstructions that may cause injury to the individual in the sling, or  
damage to the lift/track. 

Caution: Always ensure the carry bar is attached to the lift before activating RTC 
function. The added weight of the carry bar is required to ensure correct working of 

limit switches. Failure to have carry bar attached can result in limit switch error , refer 
to “Troubleshooting” Section at page 25 for recommended solution. 

Содержание C1000

Страница 1: ...lift along the track 9 Moving the H system traversing beam 10 Attaching the carry bar to the lift 11 Return to Charge 12 Basics in transferring an individual 13 C1000 Tandem Bariatric System 16 Chargi...

Страница 2: ...0 lift is one of three major components that make up this technology The other two components are the track and sling The C1000 lift runs on the lift track which is securely mounted to the ceiling str...

Страница 3: ...ft system Figure 1B Alternate sample floor plan showing basic components of a ceiling lift system Figure 2A Basic components of the ceiling lift Figure 3A Underside view of the lift Figure 2B The C100...

Страница 4: ...at all times Lift Motor 24 VDC Traverse Motor 24 VDC Optional at time of Purchase H Frame Traverse Motor 24 VDC Optional at time of Purchase Charger Input 120VAC 1 0 Amps Charger Output 24 VDC 1 5 Am...

Страница 5: ...ngle phase supply circuit when users in the United States connect the equipment to a 240 V supply system Models Table for C1000 Lift Code Description 329100 C1000 Manual Traverse 329101 C1000 Manual T...

Страница 6: ...ral Inspection and Maintenance Any accessories used with the C1000 including track and sling s should be checked to ensure that they are in good working order Check for signs of wear or fraying prior...

Страница 7: ...igure 3C Perform a brief test to ensure proper connectivity Turn the lift ON and OFF Raise and lower the carry bar For motorized trav erse lifts move the lift left and then right If these functions wo...

Страница 8: ...Figure 4C Motorized traverse H Press any button to turn ON NEW LIFTS OR LIFTS THAT HAVE HAD BATTERIES CHANGED SHOULD BE CHARGED A MINIMUM OF 30 MINUTES BEFORE USE Refer to figures 4A 4B and 4C to det...

Страница 9: ...lift a person is standing on 6C Raising lowering the carry bar Operation By pressing the DOWN arrow button or the UP arrow button the carry bar can be lowered raised to the correct height for attachin...

Страница 10: ...hand control buttons are pressed may be different than shown since the track and lift orientation may be different than installed This can be accomplished in one of two ways If the installed H traver...

Страница 11: ...r and release the latch Make sure the latch is closed safely after the Dowel Pin sit ting into the hook of Quick Release Carabiner Refer to figure 8B The Carry Bar is attached to the Lift s Quick Rele...

Страница 12: ...l or by holding the carry bar firmly Note that the RTC feature has several characteristics that may be altered to the user s preference RTC Max Time may be set anywhere from 60 sec to 240 sec in 60 se...

Страница 13: ...rry bar Caution The following steps are intended to generally illustrate the procedure involved in the lifting and transferring of an individual from one location to another using the lift track and s...

Страница 14: ...oop is allowed to rotate and rest on the top of the hook s Please check to ensure that the loops are on the hook not resting in the hook Failure to adhere to these instructions could result in serious...

Страница 15: ...now be moved to a safe location until further use or relocated to its original loca tion The lift should be turned off when not in use It is recommended that the lift be left on charge when not in ope...

Страница 16: ...Mounted on dual ceiling tracks running parallel to each other each track fitted with its own inde pendently operated motor the C 1000 Tandem Bariatric system allows a caregiver to separately lift tor...

Страница 17: ...YSTEM CARRY BAR ACCESSORY Tandem Bariatric System Carry Bar allows users to take full advantage of dual motor positioning giv ing greater control of a client s weight between the two torso and leg lif...

Страница 18: ...udible alarm beeping every 10 seconds The charger contacts with two metal charging strips that run along the inside of the track Whenever the lift is over a section of track with charging strips it wi...

Страница 19: ...Constant Charger The constant charger system has two metal roller bearings Figure 11B C that contact charging strips that run along the inside of the track When the lift is over a section of track wi...

Страница 20: ...Note As the lift is initially switched on the level of battery charge displayed may be incorrect However as soon as the lift is actually operated the charge level will update to the correct level In N...

Страница 21: ...lternate between Maintenance Lifts in the first line if user selected Number of Lifts Mode The display will alternate between Maintenance Battery in the first line if user selected Battery Level Mode...

Страница 22: ...o be reset in order to operate again This can be achieved buy pressing the plastic tab located at the end of the RED CORD back into the lift unit Then simply press any button on the hand control to re...

Страница 23: ...Handi care dealer Figure 14A Locate Allen Key supplied with the lift Step 1 Locate the Allen key in the plas tic bag containing the owners manual clev is pin ring and hand control Step 2 Locate a smal...

Страница 24: ...the lift This lift is not drip proof or water tight Failure to protect the lift from liquids may result in damage to the lift and or may cause personal injury The lift strap may be disinfected using...

Страница 25: ...local authorized dealer so that the lift can be checked to en sure proper continued operation The red indicator light on the under side lift turns RED and or a loud alarm sound is heard when an indi v...

Страница 26: ...s or sharp edges on the carry bar that may damage the straps of the sling The lift has no unusual sounds when the carry bar is moved UP DOWN or the lift is moved LEFT RIGHT Ensure that there are end s...

Страница 27: ...r your local authorized dealer must be consulted as to suitability purchase and installation Slings are the most common after purchase accessory A variety of styles sizes and colors are available Cus...

Страница 28: ..._______ Comments Complete the following section on Purchase and Service Information as soon as this equipment is installed Use the service record history to record to any completed service and repairs...

Страница 29: ...____________________ Printed Name Signature Company _____________________________________________________________ Remarks Action Taken Date _______________________ Time ________________________ Servic...

Страница 30: ...____________________ Printed Name Signature Company _____________________________________________________________ Remarks Action Taken Date _______________________ Time ________________________ Servic...

Страница 31: ...____________________ Printed Name Signature Company _____________________________________________________________ Remarks Action Taken Date _______________________ Time ________________________ Servic...

Страница 32: ...ral forces user neglect or misuse or to deliberate destruction or to batteries more than 90 days after original purchase 4 This guarantee shall be void if the equipment is not serviced by Handicare or...

Отзывы: