Checking Piston Alignment
It is very important to ensure the instrument barrel and test piston is aligned
before use. This should also be checked again after changing to a different test
mass.
1. First use the included bulls’ eye spirit level as described previously to ensure
the instrument is level.
2. Fit the test mass onto the weight lift platform. The 2.06kg or optional 1.1kg
test mass fits on the instrument first and has three holes that line up with the
cone shaped locating posts on the weight lift platform.
3. Fit the test Piston to the first weight by pulling on the ring-pull to retract the
securing clip. Push the piston into the weight and release the ring-pull to lock
it into place.
4. Select System from the main screen.
5. Press the – button next to Lift to manually lower the piston towards the
opening of the barrel.