MFI 4050
V1.3
27
3.
Place the test piston in the tool rack on the side of the instrument.
BE CAREFUL NOT TO TOUCH THE PISTON IT IS HOT
4.
Fit the cleaning probe onto the MFI instrument. If the test mass alone is
not enough to force the cleaning probe through the barrel the test weight
can be locked to the weight lift platform using the Lock pin/ R clip.
5.
Place a cleaning patch over the upper barrel opening. Press clean, the lift
mechanism will force the cleaning patch through the barrel with the
cleaning tool. Remove the patch when it passes out the bottom of the
barrel.
Repeat this with a new patch three times or until the patch is
uncontaminated after being passed through the barrel.
TEMPRATURE VERIFICATION