background image

 

 

- 3 - 

 

Battery

 

 
Insert the included batteries in the compartment at the back of the unit, please note the polarity. 
 
The LD-215 uses two rechargeable 1.2V NiMH AA cells for backup, and can run more than 24 hours on 
battery power. The batteries are continuously charged while the device is powered by the AC adapter. If 
AC power is lost for an extended period and the batteries state-of-charge becomes too low,  the LD-215 
will  alarm.    The  low-battery  alarm  is  indicated  by  a  repeating  pattern  of  two  fast  beeps  followed  by  a 
pause. 
  
NiMH rechargeable batteries 

must

 be used.  If replacement is necessary please contact the factory. 

 

First Use

 

 
Plug the power supply into an AC outlet and insert the rechargeable batteries. 
 

 

Push the RESET button to clear the program. 

 

Put the tip of the probe into the liquid Nitrogen to be monitored, height adjustment is available by 
using the supplied Allen-wrench to adjust the shaft collar, up or down, on the probe.  

 

After an initial beep the 

“TIMING” LED should blink green (if it blinks yellow go to page 1 

“Setting 

the Trip P

oint”).

 Test the installation by sliding the probe up until it is no longer submerged.  

 

The 

“TIMING”LED  will  start  to  blink  yellow  within  10  seconds.  This  indicates  that  the  probe  is 

registering an alarm level. 

 

After the alarm-delay period, the red 

“ALARM” LED will blink, and the beeper will sound and the 

output relay will change state. 

 

Care and Maintenance 

 

The LD-215 case can be cleaned with a soft dry cloth.  Avoid harsh chemicals including alcohols as they 
may permanently mar the case surface. 
 
Periodically inspect the probe and power supply cables and connections for damage.  Although the probe 
cable is jacked in PTFE (Teflon

®

), a highly resilient plastic, the temperature of liquid nitrogen does cause 

the material to stiffen.  AVOID quickly flexing any cable that has been directly exposed to LN2 liquid until 
it has time to warm up. 
 
The  LD-215  should  be  recalibrated  periodically  as  electronic  device  parameters  may  drift  over  time.  
Simply follow the instructions found on the first page  in the section title

d “

Setting the Trip Point

”.  After 

recalibrating,  pull the probe tip up  out  of the liquid (one inch or less above  the  surface) and wait for the 
device to alarm.  

Verify that the “ALARM” LED flashes, the beeper sounds, and the relay changes state.  

To avoid a lengthy wait, you may want to set the alarm delay to zero before the test and then return it to 
the desired setting when complete. 
 
The batteries may be tested periodically by unplugging the power supply and allowing the unit to run on 
battery  power  for  5-10  minutes.    If  the  LD-215  fails  to  operate,  or  if  the  low  battery  alarm  activates,  the 
batteries should be replaced. The batteries should be replaced every 3-5 years. 
 
 
 

Отзывы: