background image

Hampshire Controls Corp.

 

 

1 Grove St. 

Dover, NH 03820 

(603) 749-9424 or Toll Free US/Canada (866) 496-9424   

www.hampshirecontrols.com

 

 
 

LD-215 Liquid Nitrogen Level Sensor

 

User Instructions

 

 

 

 

Overview of Operation

 

 
The 

LD-215

  uses  a  cryogenic  RTD  to  sense  the  surface  of 

liquid  nitrogen  by  monitoring  the  self-heating  of  the  probe.  The 
alarm will sound and the output relay will change state when the 
liquid  level  drops  below  the  probe  tip.  In  reverse  mode,  the 
alarm  will  sound  and  the  output  relay  will  change  state  if  the 
liquid level makes contact with the probe tip. 
  
The  alarm  can  be  silenced  for  a  selectable  period  of  time  by 
pressing  the 

“MUTE”  button.    During  mute  the  “ALARM”  LED 

will continue to flash and the 

output relay will stay in the “alarm” 

condition. A return to normal liquid level will clear the alarm. See 
page 2 for mute duration adjustment. 
 
An alarm delay can also be specified.  After the unit recognizes 
an alarm condition, the green status

 

“TIMING” LED will switch to 

yellow  and  the  unit  will  delay  for  the  selected  period  before 
alarming. See page 2 for alarm delay adjustment. 
 
Any alarm condition, muted or not, will cause the relay to switch 
to alarm state. The user can wire to the “Normally Open” (NO, 
closed  on  alarm)  or  “Normally  Closed”  (NC,  open  on  alarm) 
contacts for external dialers, or other functionality. 

 
 
 

Setting the Trip Point (Calibration)

 

 
Make sure the probe tip is fully submerged in liquid Nitrogen, wait 2 minutes for the probe temperature to 
fully stabilize, press the 

RED

 push-button twice (pictured below).  Observe the 

“TIMING” LED flash green 

7 times to confirm successful setting of the new setpoint. 

 

Отзывы: