43

Subject to change without notice

STOP

M e n u   s t r u c t u r e

The autorange function is to be used with care. For
measurements on high impedance source and
measurement voltages in the range (90%) of full
scale 1 V, changing to a higher range is possible
with activated AUTO function. The HM8112-3 has
an input impedance of 10M

 in the 10 V range

instead of 1 G

 in the 1 V range. By loading a high

impedance source of several 100 M

 with by input

impedance of 10 M

 the measurement result will

be errouneous.

Menu structure

Menu prompting

From every measurement function the menu can be entered
by pressing MENU 

18

. Within the menu, every button which

can be used is illuminated. The menu can be always left by
pressing ESC 

19

 without acceptance of entered values.

Call of the menu by MENU 

18

.

Choice of menu item with 

 

20

 and 

 

23

. The menu item is

opened with MENU 

18

 or branch to the next menu level.

Selection of parameters shown with 

 

20

 and 

 

23

.

Acceptance of parameters changed with MENU 

18

. If the menu

is left, the instrument will return to the last measurement
function.

19

  ESC

Leaving the menu. Return to the last measuring function
without acceptance of the value entered.

20

  

Rotating menu prompting. Jump to the next menu item with
every key operation. On reaching the last menu item the display
continues with the first menu item.

23

  

Rotating menu prompting. Jump to the previous menu item
with every key operation. On reaching the first menu item the
display will roll over and continue with the last menu item.

22

  ENTER

Use this button only in the logger menu „6:Logger.“ Switching
to the next buffered value by every key operation or acceptance
of an input.

Menu structure and function

The menu will be accessed by pressing MENU 

18

. It branches

to the submenues described below.

0:Time

The time intervals between the measurements are adjustable
from 0.01 s to 60 s. That means, a reading is taken every 0.01 s
or only every 60 s.
The sampling rate can assume the following values:

10 ms

(only via interface)

50 ms

(only via interface)

100 ms

(default setting after switch-on)

500 ms
1 s
10 s
60 s

That means, for example, that every 500 ms a measurement
is taken and the value is updated in the display. After switch-
on a value of 100 ms will be preset. Removal of the line voltage
will not save a selected value.

If the measurement function is changed, the HM8112-3
assumes the default sampling rate, unless a sampling rate
between 10 s and 60 s is chosen. Then changing the measuring
function sets the sampling rate automatically to 1 s.

Example: 

The sampling rate for V

DC 

is set to 60 s. Then the

function A

DC 

is selected. The instrument will reduce the

sampling rate automatically to 1 s. The new sampling rate
applies to all functions. If a sampling rate greater than 1 s is
needed, it has to be selected after every change of function.

A sampling rate of 60 s means:
The HM8112-3 integrates the input voltage and the
the reference voltage over a period of 60 s. After
expiry of this time the value calculated will be
displayed.

1:Filter

Selection of the number of values taken for averaging. In case
of selection of a number greater than 1, the selected number
will be taken for averaging. By calculating a new averaged
value, the first measured value will be discarded and the mean
value will be computed.

OFF

(default setting after switch-on)

2
4
8
16

2:Temp

In this menu item the dimension for the temperature measure-
ment is selected.

° C

Degrees Celsius

° F

Degrees Fahrenheit

The dimensions selected last will be saved even if the mains
will be turned off.

3: Sensor

Here the temperature sensor used is selected.
After switch on of the HM8112-3 and selection of the menu item
“3:Sensor“, if a measurement function other than temperature
measurement was set, PT 100 as temperature sensor is selected
by default. If the thermocouple is chosen the HM8112-3 is in
measurement function 

15

 

ϑ

TH.

Also the instrument will return to the measurement function

13

 

ϑ

PT after selection of the PT-sensor. The sensor type

selected last will be stored in the instrument even if the main
voltage is turned off.

K–type: thermocouples NiCr-Ni (default setting after
switch-on)

J–type: thermocouples Fe-CuNi

PT1000: platinum resistance sensor with R

= 1000

PT100: platinum resistance sensor with R

= 100

 (default

setting after switch-on)

Comp

For measurements with thermocouples a reference with a
known temperature must be defined. This reference tempera-
ture is provided to the HM8112-3. Therfore 3 methods are
possible:

HINT

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

Содержание HM8112-3

Страница 1: ...6 D I G I T P R E C I S I O N M U L T I M E T E R H M 8 1 1 2 3 Handbuch Manual Deutsch English 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com...

Страница 2: ...e sufficient to guard against the intrusion of undesired signals This will not cause destruction or malfunction of HAMEG instruments however small deviations from the guranteed specifications may occu...

Страница 3: ...ective Switch Off 37 Power supply 37 Line fuse 37 Power switch 37 Control Elements 38 Concept of the HM8112 3 39 Reference 39 Integrated AD converters 39 Moving average 39 Measurement of alternating v...

Страница 4: ...ments of AC DC or AC 2 wire 4 wire measurements Resolutions 100 nV 100 pA 100 0 01 C F Offset correction Range selection automatically and manual 10 measurements per second Measuring time 0 1 sec to 6...

Страница 5: ...sec 1 to 60 sec Display ranges 120 000 digit 1 200 000 digit 1 A range 100 000 digit 1 000 000 digit Resolution 1 nA 100 pA Accuracy DC 45 Hz 1 kHz 1 kHz 5 kHz 1 year 23 2 C 0 02 0 002 0 1 0 08 0 2 0...

Страница 6: ...1010 1 Please observe all warnings in this manual in order to preserve safety and guarantee operation without any danger to the operator According to safety class 1 requirements all parts of the housi...

Страница 7: ...on our web site http www hameg com Even after expiry of the warranty period please do not hesitate to contact our HAMEG customer service for repairs and spare parts Return Material Authorization RMA...

Страница 8: ...t series 17 MIN min value during a test series 18 MENU Call of the menu acceptance of values entered 19 ESC Leaving the menu without acceptance of the values entered 20 down Switching to a higher rang...

Страница 9: ...e frequency of the signal and the period will be calculated from the frequency This combined measurement of the number of zero points and of the period of a signal allows the measurement of very small...

Страница 10: ...must be chosen PT100 2 wire 4 wire 1 k range PT1000 2 wire 4 wire 10 k range Thermocouple VDC 100 mV range Whether 2 wire or 4 wire measurement has to be selected depends on the PT temperature sensor...

Страница 11: ...up to 600 V true RMS with DC component Direct coupling of the circuit to the instrument and using of the same high precision input divider like VDC The input impedance of the HM8112 3 is 10 G in 100 m...

Страница 12: ...d 100 ms to 60 s Delay 100 ms after change of function Display Dimension C or F Linearisation according to EN60584 V A SENSE SOURCE LO HI max 250Vrms max 850 Vpk max 850 Vpk FUSE 1A F250V CAT II max I...

Страница 13: ...fter switch on a value of 100 ms will be preset Removal of the line voltage will not save a selected value If the measurement function is changed the HM8112 3 assumes the default sampling rate unless...

Страница 14: ...MENU MENU 2 Temp 16 8 4 2 Off default Choice temperature sensor PT100 PT100 Fe CuNi NiCr Ni MENU MENU 3 Sensor K J PT1000 PT100 Comp Fixing the reference for the thermocouple MENU MENU Comp Comp Ext...

Страница 15: ...es If several thermocouples will be attached to the HM8112 3 via a scanner the use of the ice bath would be necessary for each thermocouple To overcome this the ambient temperature or even a source wi...

Страница 16: ...is displayed 7 COM In this menu the baud rate can be chosen Either 9600 baud or 19200 baud are available The remaining interface parameters cannot be changed Interface parameters adjustable Rs Off The...

Страница 17: ...ring connectors Therefore all terminals V SENSE 24 ground 25 and A SOURCE 26 should be disconnected The cover of the fuse holder has to be turned ccw with a screw driver having a suitable blade As the...

Страница 18: ...y of at least 35 ms must be observed then the next command can be sent A transmission of invalid commands is answered with 2D0LF 2D0LF 2D0LF 2D0LF 2D0LF in case of wrong length of the command or void...

Страница 19: ...ty 10 Ohm D Sensor RTD 2WIRE Pt100 Pt1000 E Sensor RTD 4WIRE Pt100 Pt1000 F Sensor TH J K 0 AUTO RANGE OFF ON UP DOWN 1 MEAS Time 10ms 50ms 100ms 500ms 1s 10s 60s UP DOWN 3 Filter CONT 2 4 8 16 0 4 Ma...