41

Subject to change without notice

Frequency and period

12

FREQ./PERIOD

Switching between frequency and period measurement by
repeatedly pressing this button. At measurement of DC voltage
the display shows “0 Hz“ for frequency and “INF“ for period
measurement (INF = infinity). As the period is calculated from
the measured frequency it is a division by zero.

There is no auto range function for frequency and
period measurements. That means the range of
the V

AC 

measurement is taken. Is necessary to

measure the alternating voltage in V

AC

 first and

afterwards call up the FREQ./PERIOD function.

Temperature measurement

Switching between 2-wire and 4-wire measurement by
repeatedly pressing 

ϑ

PT-button 

13

. This is indicated in the

display by „2w“ for 2-wire and by “4w“ for 4-wire measurement.
Additionally the terminals to be used are illuminated. For
compensation of the wiring resistance at 2-wire measure-
ments 100 m

 is stored by default. This value can be changed

via interface.

For exact measurements it is necessary to cali-
brate the measurement section with ZERO 

. This

calibration is done for PT sensors by resistance
measurement or for thermocouples by voltage
measurement but not by temperature measure-
ment (see ZERO

).

13

  

ϑ

PT with 4-wire-temperature measurement

Measuring method:

4 wire resistance measurement with linearisation
according to EN60751 for PT100 and PT1000.

Temperature sensor:

PT100, PT1000 resistance sensors

Display range:

Display scale

Resolution

Celsius:

–200 °C to +800 °C

0.01 °C

Fahrenheit:

–328 °F to +1472 °F

0.01 °F

Test current:

PT100

1 mA

PT1000

100 µA

C o n t r o l   e l e m e n t s   a n d   d i s p l a y

V

AC

Alternating voltage measurement up to 600 V, true RMS
without the DC component. In AC a capacitor is inserted. The
input impedance of the HM8112-3 is R

i

 = 10 M

.

V

AC+DC

Alternating voltage measurement up to 600 V, true RMS with
DC component.  Direct coupling of the circuit to the instrument
and using of the same high precision input divider like V

DC

.

The input impedance of the HM8112-3 is 10 G

 in 100 mV

range, 10 M

 in the other ranges.

Current
measurement

A

DC

Direct current
measurement.
Auto range func-
tion up to and
including
the range 1 A

A

AC+DC

Alternating current measurement, true RMS with DC com-
ponent. Auto range function over the entire range of 1 A.

Resistance measurement

Switching between 2 wire and 4 wire measurement by repeatly
pressing 

-button 

11

. This is shown in the display by „2w“ for

2 wire and by „4w“ for 4 wire measurement. Additionally the
terminals to be used are illuminated. For exact measurements
it is necessary to null any offsets by pressing ZERO 

.

11

 2-wire

resistance
measurement

For compensation
of the wiring resis-
tance with 2-wire
measurements
100 m

 is stored by

default. This value
can be changed via
the interface.

11

 4-wire

resistance
measurement

V

A

SENSE

SOURCE

LO

HI

max.

250Vrms

max.

850

Vpk

max.

850

Vpk

,

ϑ

FUSE
1A 
F250V

CAT 

II

max. INPUT

600Vrms / 1Arms

+

Power
input

(Source)

+

Voltage

measurement

(Sense)

Resistance measurement

4-wire

V

A

SENSE

SOURCE

LO

HI

max.

250Vrms

max.

850

Vpk

max.

850

Vpk

,

ϑ

FUSE
1A 
F250V

CAT 

II

max. INPUT

600Vrms / 1Arms

+

resistance
measurement

2-wire

STOP

STOP

V

A

SENSE

SOURCE

LO

HI

max.

250Vrms

max.

850

Vpk

max.

850

Vpk

,

ϑ

FUSE
1A 
F250V

CAT 

II

max. INPUT

600Vrms / 1Arms

+

Direct  or

alternating voltage

V

A

SENSE

SOURCE

LO

HI

max.

250Vrms

max.

850

Vpk

max.

850

Vpk

,

ϑ

FUSE
1A 
F250V

CAT 

II

max. INPUT

600Vrms / 1Arms

    +

Direct or alter

Direct or alter

Direct or alter

Direct or alter

Direct or alter-----
nating current

nating current

nating current

nating current

nating current

    –

V

A

SENSE

SOURCE

LO

HI

max.

250Vrms

max.

850

Vpk

max.

850

Vpk

,

ϑ

FUSE
1A 
F250V

CAT 

II

max. INPUT

600Vrms / 1Arms

+

Power
input

(Source)

+

4-wire-temperature measurement with PT100

Voltage

measurement

(Sense)

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

Содержание HM8112-3

Страница 1: ...6 D I G I T P R E C I S I O N M U L T I M E T E R H M 8 1 1 2 3 Handbuch Manual Deutsch English 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com...

Страница 2: ...e sufficient to guard against the intrusion of undesired signals This will not cause destruction or malfunction of HAMEG instruments however small deviations from the guranteed specifications may occu...

Страница 3: ...ective Switch Off 37 Power supply 37 Line fuse 37 Power switch 37 Control Elements 38 Concept of the HM8112 3 39 Reference 39 Integrated AD converters 39 Moving average 39 Measurement of alternating v...

Страница 4: ...ments of AC DC or AC 2 wire 4 wire measurements Resolutions 100 nV 100 pA 100 0 01 C F Offset correction Range selection automatically and manual 10 measurements per second Measuring time 0 1 sec to 6...

Страница 5: ...sec 1 to 60 sec Display ranges 120 000 digit 1 200 000 digit 1 A range 100 000 digit 1 000 000 digit Resolution 1 nA 100 pA Accuracy DC 45 Hz 1 kHz 1 kHz 5 kHz 1 year 23 2 C 0 02 0 002 0 1 0 08 0 2 0...

Страница 6: ...1010 1 Please observe all warnings in this manual in order to preserve safety and guarantee operation without any danger to the operator According to safety class 1 requirements all parts of the housi...

Страница 7: ...on our web site http www hameg com Even after expiry of the warranty period please do not hesitate to contact our HAMEG customer service for repairs and spare parts Return Material Authorization RMA...

Страница 8: ...t series 17 MIN min value during a test series 18 MENU Call of the menu acceptance of values entered 19 ESC Leaving the menu without acceptance of the values entered 20 down Switching to a higher rang...

Страница 9: ...e frequency of the signal and the period will be calculated from the frequency This combined measurement of the number of zero points and of the period of a signal allows the measurement of very small...

Страница 10: ...must be chosen PT100 2 wire 4 wire 1 k range PT1000 2 wire 4 wire 10 k range Thermocouple VDC 100 mV range Whether 2 wire or 4 wire measurement has to be selected depends on the PT temperature sensor...

Страница 11: ...up to 600 V true RMS with DC component Direct coupling of the circuit to the instrument and using of the same high precision input divider like VDC The input impedance of the HM8112 3 is 10 G in 100 m...

Страница 12: ...d 100 ms to 60 s Delay 100 ms after change of function Display Dimension C or F Linearisation according to EN60584 V A SENSE SOURCE LO HI max 250Vrms max 850 Vpk max 850 Vpk FUSE 1A F250V CAT II max I...

Страница 13: ...fter switch on a value of 100 ms will be preset Removal of the line voltage will not save a selected value If the measurement function is changed the HM8112 3 assumes the default sampling rate unless...

Страница 14: ...MENU MENU 2 Temp 16 8 4 2 Off default Choice temperature sensor PT100 PT100 Fe CuNi NiCr Ni MENU MENU 3 Sensor K J PT1000 PT100 Comp Fixing the reference for the thermocouple MENU MENU Comp Comp Ext...

Страница 15: ...es If several thermocouples will be attached to the HM8112 3 via a scanner the use of the ice bath would be necessary for each thermocouple To overcome this the ambient temperature or even a source wi...

Страница 16: ...is displayed 7 COM In this menu the baud rate can be chosen Either 9600 baud or 19200 baud are available The remaining interface parameters cannot be changed Interface parameters adjustable Rs Off The...

Страница 17: ...ring connectors Therefore all terminals V SENSE 24 ground 25 and A SOURCE 26 should be disconnected The cover of the fuse holder has to be turned ccw with a screw driver having a suitable blade As the...

Страница 18: ...y of at least 35 ms must be observed then the next command can be sent A transmission of invalid commands is answered with 2D0LF 2D0LF 2D0LF 2D0LF 2D0LF in case of wrong length of the command or void...

Страница 19: ...ty 10 Ohm D Sensor RTD 2WIRE Pt100 Pt1000 E Sensor RTD 4WIRE Pt100 Pt1000 F Sensor TH J K 0 AUTO RANGE OFF ON UP DOWN 1 MEAS Time 10ms 50ms 100ms 500ms 1s 10s 60s UP DOWN 3 Filter CONT 2 4 8 16 0 4 Ma...

Отзывы: