          F u n c t i o n s

input A. The bandwidth for low frequencies 
is as low as 10 Hz (3 dB) when the input is AC 
coupled. (Input C has a fi xed AC coupling).

  1:20  (pushbutton)

 

Selection of input signal attenuation. Pressing 
this button attenuates the input signal by 26 dB 
before it is applied to the input amplifi er.

  Auto Trigger (AC)  (pushbutton)

 

With Auto Trigger active the counter triggers 
in the middle of the input signal. Auto Trigger 
always uses AC-coupling. (AC = pushbutton 
depressed).

  INPUT A  (BNC connector)

 

Signal input with a sensitivity of 20 mV

rms

 up to 

80 MHz and 60 mV

rms

 up to 150 MHz. The input 

is protected against overvoltage up to 400 V 
(DC + AC

peak

).

 

Input impedance: 1MΩ II 40pF.

  TRIGGER LEVEL  (adjusting knob)

 

Continuously adjustment of trigger level.

 

TRIGGER (LED)

 

 

3 State trigger indicator. The LED fl ashes when 
triggering is correct. The LED lights when the 
trigger level is above the input signal level, it 
is not activated when the trigger level is below 
the input signal level.

  8 digit display  

(7 segment LEDs, 7.65 mm high)

 

for the measuring result (8 digit max. + expo-
nent).

 

Hz (LED):

 Indicates the measurement of a 

frequency.

 

Sec (LED):

 Indicates the measurement of 

time.

HM 8001-2:

 

External Gate Input  (BNC connector)

 

Allows the measurement of the input signal, 
controlled by an external source.

Triggering, measuring 

and counting

Power-on test

A practical test of the correct operation of the 
HM8021-4 is run automatically at power on. As 
soon as power is applied, the display shows type 
and version of the actual instrument and the GATE 
indicator appears, LEDs are lit and the Eprom and 
all functions of the counter are tested. The test 
runs for about 2 seconds. If an error is detected 
it is indicated by an ”I” at the leading digit and 
followed by the number of the test that failed.

 

microprocessor RAM error

 

program ROM error

 

counting chain error

When the tests are completed satisfactorily, the 
counter sets the display to zero and selects the 
default measuring function A.

Measuring functions

FA/FC: 

Sets the instrument to measure the 
frequency of the signal connected to 
input A/C.

TOT.: 

The counter will totalize events 
(pulses or cycles) on input A. 
Measurement stops and display 
freezes as soon as the input signal 
is removed or DISPLAY HOLD is 
depressed. Depressing RESETcluring 
totalizing clearsthe display and starts 
a new measurement when releasing 
the button.  Reset is active as long as 
the switch 

 is depressed. Totalizing 

A during External Gate is performed 
applying a TTL signal to the Ext. Gate 
input.

PA:  

Sets the counter to measure the peri-
od duration of the signal connected to 
input A.

TI 

:  Sets the counter to measure the

(AVG)  

average pulsewidth between positive 
slope and the negative slope of an 
input signal at input A.

Содержание HM8021-4

Страница 1: ...r s a l C o u n t e r H M 8 0 2 1 4 Handbuch Manual Manuel Manual Deutsch English Fran ais Espa ol 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www Tes...

Страница 2: ...cable HZ72 from HAMEG is suitable 2 Signal cables Basically test leads for signal interconnection between test point and instrument should be as short as possible Without instruction in the manual fo...

Страница 3: ...rmation regarding CE marking 14 Universal Counter HM8021 4 16 Speci cations 17 Important hints 18 Safety 18 Used Symbols 18 Operating conditions 18 Warranty and repair 18 Maintenance 19 Operation of t...

Страница 4: ...put A Input impedance 1 M maximum sensitivity 20mVrms Input C Input impedance 50 maximum sensitivity 30mVrms Time interval resolution up to 10 ps Offset mode over the entire measurement range Gate inp...

Страница 5: ...ty normal triggering DC 80 MHz 20 mVrms sine wave 80 mV pulse 80 MHz 150 MHz 60 mVrms sine wave 20 Hz 80MHz auto trig 50 mVrms sine wave Minimum pulse width 5ns Input noise 100 V typ Coupling AC or DC...

Страница 6: ...t or discon nection of the protective earth terminal is likely to render the instrument dangerous Intentional interruption is prohibited The istrument must be disconnected and secured against unintent...

Страница 7: ...d rear frames After removal of the two screws at the rear of the module both chassis covers can be lifted When reclosing the module care should be taken that the guides engage correctly with the front...

Страница 8: ...ng will restart when Display Hold is switched off Display Hold starts and stops counting in the TOTALIZE function mode Function indicators LEDs Refer to Measuring functions Function pushbuttons The le...

Страница 9: ...led by an external source Triggering measuring and counting Power on test A practical test of the correct operation of the HM8021 4 is run automatically at power on As soon as power is applied the dis...

Страница 10: ...r level setting The input signal is triggered from 50mV up to the maximum input voltage of 5V The input frequency for the C input must always be in the range from 100MHz to 1000MHz For frequencies low...

Страница 11: ...ate time of 10sec Period measurement For measurement of the period duration the re ciprocal value of the frequency T 1 f is measured in seconds The signal is applied as for frequency measurement Time...

Страница 12: ...and DISPLAY HOLD should be in OFF position 2 Apply a frequency standard of 1 5 or 10MHz to input A and adjust channel A trigger setting for a stable reading 3 Depress RESET and then GATE TIME for appr...

Страница 13: ...the counter has recognized a 10MHz standard During the following 45 seconds the signal is measured and the HM8021 4 is recalibrated Finally the new calibrated unit automatically switches back to norm...

Отзывы: