STORAGE MODE ONLY

In

 XY 

storage mode the readout indicates 

“XY” 

and

the

 RFR LED (9) 

is lit. No other STOR. MODE can be

chosen. Instead of the time deflection coefficient, the
readout displays the sampling rate (e.g. “100MS/s”)
which can be selected by the

 TIME/DIV. 

knob

 (29)

.

In contrast to analog mode, the

 INV. (38) 

pushbutton is

active and the X signal can be inverted. Also the

 Y-POS.

II (15) 

knob must be used for X position shift instead of

the

 X-POS. (19) 

control. If

 HOLD 

is active, the current

XY

 signal can be moved in

 X (Y-POS. II) 

and

 Y (Y-POS.

I)

 direction.

(24) TRIG.

 ALT

- Pushbutton with double function and associated

LEDs.

The pushbutton and the LEDs are deactivated in XY
mode and if line (mains) triggering is selected.

TRIG.

With the aid of this pushbutton, the trigger source can
be chosen. There are three trigger sources available:

channel I, channel II (both designated as internal trigger
sources) and the

 TRIG. EXT. (39) 

input for external

triggering. The availability of the internal sources depends
on the actual channel mode. The actual setting is
indicated by the associated LED(s) and the readout.

Briefly pressing the pushbutton switches over in the
following sequence:

I - II - EXT - I 

in DUAL mode,

I - EXT - I 

if mono channel I is present,

II - EXT - II 

under mono channel II conditions.

Each condition is indicated by the associated LED and
displayed by the readout (

“TR:Y1...”

“TR:Y2...”

 and

“TR:EXT...”

). The trigger point symbol is switched off

in external trigger condition.

ALT

Pressing and holding the pushbutton selects alternate
triggering in DUAL mode. Under these conditions both 

I

and 

II

 LEDs lit and the readout displays

 “TR:ALT...”

. As

alternate triggering requires alternate channel operation
in analog mode, alternate channel switching is set
automatically. A change of the time base then has no
affect regarding the channel switching mode. In addition
to the deflection coefficients display,

 “ALT” 

is displayed

by the readout instead of 

“CHP” 

if analog mode is present.

In alternate trigger mode the trigger point symbol is
switched off.

 TVL

,

 TVF 

trigger coupling and line (

~

)

triggering can not be selected.

Alternate triggering is not available or automatically
switched off under the following conditions:

SINGLE 

(SGL) mode

ADD 

(addition) mode,

alternate 

(A & B) time base mode,

time base mode and

line 

(mains) trigger coupling.

STORAGE MODE ONLY

In

 ROLL 

mode the signal capture is untriggered. Con-

sequently alternate triggering is not available.

(25) VOLTS/DIV.

 – Control knob

This control knob for

 channel II 

has a double function.

The following description relates to the input attenuator
function (

VAR

 LED dark).

Turning the control knob clockwise increases the
sensitivity in a 1-2-5 sequence and decreases it if turned
in the opposite direction (ccw.). The available range is
from 1mV/div up to 20V/div. The knob is automatically
switched inactive if the channel related to it is switched
off, or if the input coupling is set to GD (ground).

The deflection coefficients and additional information
regarding the active channels are displayed in the
readout, i.e. 

“Y2: deflection coefficient, input coup-

ling”

. The

 “ : “

 symbolizes calibrated measuring condi-

tions and is replaced by the “

 > 

“ symbol in uncalibrated

conditions.

In

 DUAL

,

 ADD 

(addition) and

 XY 

mode both deflection

coefficients etc. are displayed. Instead of

  “Y1:...” 

and

“Y2:...” 

the readout displays

  “Y:...” 

and

  “X:...” 

in XY

mode.

(26) CH II

 

VAR.

 - Pushbutton with several functions.

Channel mode

Briefly pressing the pushbutton sets the instrument to

channel II (Mono CH II) 

mode. The deflection coefficient

displayed in the readout indicates the current conditions
(

“Y2...“

). If neither external nor line (mains) triggering

was active, the internal trigger source automatically
switches over to channel II (

“TR:Y2...“

). The last function

setting of the 

VOLTS/DIV (25) 

knob remains unchanged.

All channel related controls are active if 

INPUT CHII

 

(36)

is not set to

 GD (38)

.

VAR.

Pressing and holding this pushbutton selects the

 VOLTS/

DIV. (25) 

control knob function between attenuator and

vernier (variable). The current setting is displayed by the

VAR-

LED located above the knob.

After switching the

 VAR-LED (25) 

on, the deflection

coefficient is still calibrated. Turning the 

VOLTS/DIV. (25)

control knob counter clockwise reduces the signal height
and the deflection coefficient becomes uncalibrated. The
readout then displays i.e.

  “Y2>...” 

indicating the

uncalibrated condition instead of

 “Y2:...”

. Pressing and

holding the

 CHII 

pushbutton again switches the LED

off, sets the deflection coefficient into calibrated
condition and activates the attenuator function. The
previous vernier setting will not be stored.

The

 CHII 

pushbutton can also be pressed simultaneously

with the

 DUAL (23) pushbutton

. Please note item

 (23)

.

Controls and Readout

Содержание HM1507-3

Страница 1: ...ls 32 Line Mains triggering 32 Alternate triggering 32 External triggering 33 HOLD OFF time adjustment 33 B time base 2nd time base 33 Triggering after Delay 33 AUTO SET 34 Component Tester analog mod...

Страница 2: ...operation the following hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer...

Страница 3: ...Variable 2 5 1 up to 1 25s div X MAG x10 5 max 5ns div Holdoff time variable to approx 10 1 Bandwidth X amplifier 0 3MHz 3dB X Y phase shift 3 below 220kHz Digital Time Base Accuracy 3 1 2 5 sequence...

Страница 4: ...in a socket outlet provided with a protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor The mains line plug must be insert...

Страница 5: ...on ments The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be use...

Страница 6: ...wnwards at each change Pure direct voltages can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbo...

Страница 7: ...In this case an appropriate high voltage capacitor approx 22 68nF must be connected in series with the input tip of the probe With Y POS control input coupling to GD it is possible to use a horizontal...

Страница 8: ...its full length at not too great steepness and that the horizontal distance at 10 and 90 of the amplitude is measured If the edge shows rounding or over shooting the 100 should not be related to the p...

Страница 9: ...as short and thick as possible When the attenuator probe is connected to a BNC socket a BNC adapter should be used In this way ground and matching problems are eliminated Hum or interference appearing...

Страница 10: ...and no other LED on the front panel is lit except A All INTENS settings are stored after the instrument is switched off The AUTO SET function switches the readout on and selects A time base mode A LED...

Страница 11: ...cally again it will be required to stop rotating the knob for at least about 2 seconds Attention The dynamic range limits of the A D converter may become visible if a Y position shift is performed af...

Страница 12: ...STOR MODE Pushbuttons with associated LEDs These functions are not available in analog mode If digital SINGLE SGL mode has not been chosen one of the associated LEDs is lit The signal capture and disp...

Страница 13: ...g mode on or off SINGLE mode is indicated by the associated SGL LED SINGLE mode is available in digital as well as in analog mode and remains unchanged when switching over from analog to digital mode...

Страница 14: ...tton until a beep confirms reference memory acquisition Before starting the procedure the current signal can but must not be protected by the HOLD function As the reference signal is then equal to the...

Страница 15: ...the DC content of a signal by comparing the actual Y position with the previously determined 0 Volt Y position Y POS I Symbol Provided that the readout is displayed and DC REFE RENCE ON is selected i...

Страница 16: ...control knob causes a different trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal voltage crosses the trigger point In most Yt modes the trigger po...

Страница 17: ...o GD 34 38 Whether alternated or chopped channel switching is present in analog mode depends on the actual time base setting and is displayed in the readout ALT analog mode only displayed in the reado...

Страница 18: ...e A B time base mode B time base mode and line mains trigger coupling STORAGE MODE ONLY In ROLL mode the signal capture is untriggered Con sequently alternate triggering is not available 25 VOLTS DIV...

Страница 19: ...start of the intensified sector In the free run condition delay trigger not active of the B time base an approximate value is displayed in the readout Dt This is an aid to find the position of the int...

Страница 20: ...information is also displayed in the readout as an approximate value e g Dt 2 5ms related to the calibrated A time coefficient uncalibrated i e Dt 2 5ms The width of the segment decreases when the B t...

Страница 21: ...on coefficient setting is not possible in storage mode Underneath the front panel sector described before the BNC sockets and four pushbuttons are located 32 INPUT CH I BNC socket This BNC socket is t...

Страница 22: ...he CRT there are the controls for the readout the component tester and the square wave calibrator with their outputs Please note The following description of the cursor related controls assumes that t...

Страница 23: ...nt must be activated Briefly pressing the CHI II 1 t ON OFF pushbutton switches from f to t and the following measure ment modes until the readout displays f again In all conditions vertical cursor li...

Страница 24: ...ayed in the readout must be multiplied by 100 if e g a x100 100 1 probe is used In case of x10 10 1 probes the probe factor can be automatically included see AC DC 33 and AC DC 37 t In time or frequen...

Страница 25: ...connect the test lead s to the test point s Then switch on the instrument and after wards the device under test The oscilloscope is switched on by depressing the red POWER pushbutton After a few seco...

Страница 26: ...tton is depressed Connect the probe to CH I input Depress the CAL pushbutton for 1MHz Operate the oscilloscope as described under 1kHz but select for 0 2 s div time deflection coefficient setting Inse...

Страница 27: ...are the same for both channels because the X x10 magnifier is inactive in XY mode The bandwidth of the X amplifier is lower than the Y amplifier and the phase angle which increases with higher freque...

Страница 28: ...l cycles of the modulation frequency are visible Strictly speaking triggering should be external with modulation frequency from the AF generator or a demo dulator However internal triggering is freque...

Страница 29: ...ger coupling setting In the absence of an adequate trigger signal or when the trigger controls particularly the trigger LEVEL control are misadjusted no trace is visible i e the screen completely blan...

Страница 30: ...ernal trigger voltage itself is displayed first with internal triggering In most cases the composite video signal has a high DC content With constant video information e g test pattern or color bar ge...

Страница 31: ...iting the trigger pulse This causes an untriggered display and a flashing trigger LED HOLD OFF time adjustment For instrument specific information please note DEL POS HO 28 in section Controls and Rea...

Страница 32: ...ically for a display of approx 2 signal periods The time base setting occurs randomly if complex signals consisting several frequencies e g video signals are present AUTO SET sets the instrument autom...

Страница 33: ...mination of actual values will come with experience or by direct comparison with a component of a known value Testing Capacitors and Inductors Capacitors and inductors cause a phase difference between...

Страница 34: ...ared with single components In case of doubt one component terminal may be unsoldered This terminal should then not be connected to the ground socket avoiding hum distortion of the test pattern Anothe...

Страница 35: ...d Controls Vertical resolution The dot density in each operation mode is 8 bits 28 256 dots displayed over a height of roughly 10 divisions The instrument is adjusted for 25 dots per division This eas...

Страница 36: ...k must on no account be visible Visible trace fault without input signal bright dot on the left side or decreasing brightness from left to right or shortening of the baseline Cause Incorrect Unblankin...

Страница 37: ...mitted This is achieved by adjusting the trimmer located in the 2 1 attenuator The shape of the square wave should then be the same in each input attenuator position Operating Modes CH I II DUAL ADD C...

Страница 38: ...Then both readout information regarding the time deflection coefficient and the time difference show exactly the same value After switching over from time to frequency measurement the readout display...

Страница 39: ...ng on of the oscilloscope and the first command SPACE CR 20hex 0Dhex sent from the PC the baud rate is recognized and set automatically between 110 baud and 19200 baud The oscilloscope is then switche...

Отзывы: