Hameg HM1008 Скачать руководство пользователя страница 28

28

Subject to change without notice

7.2.5

  Edit 

>

 Constant Edit

In the CW position of the INTENS knob 

 an additional item 

”Edit“ and an arrow symbol is displayed. Pressing the associated 
function button opens the sub submenu ”Constant Edit“. This 
enables to choose a Number, its ”Dec.Point“ (decimal point) and 
its ”Prefi x“ of the ”Unit“, all with the INTENS knob 

. The unit 

need not be selected it is only shown to assist the memory. 

7.3

 

Display

STOP

 Attention!

 

The display of mathematic signal(s) automatically 
switches previously displayed reference signals off 
and vice versa. 

The function ”Display“ is available twice in the menu and may 
be switched on or off in any combination. This allows to display: 
no result, one result, the result of two equations as signals. The 
displays will come forward upon leaving the ”Mathematics“ 
menu. Also the designation of the equation (e.g. MA2) will be 
shown. Select the equations to be displayed with the INTENS 
knob 

.

The mathematics signal is automatically scaled, this is inde-
pendent of the graticule, of Y and time base parameters, the 
scale will not be shown. Hence the measurement of the signal 
amplitudes must be performed using the CURSOR (V to GND) 
after the ”reference“ (e.g. MA2) of the CURSOR to the ”mathe-
matics signal“ and its scale was established (AUTO/CURSOR 
MEASURE 

 pushbutton 

>

 Cursors 

>

 reference 

>

 e.g. MA2). 

The readout may then display e.g.: ”V(MA2): 900 mV“.

Divisions by zero will be ignored and an error message dis-
played.

7.4 Units

Each function ”Display“ will be associated with a function ”Unit“  
which can be selected with INTENS 

 and will be attached to 

the result. 

  ACQUIRE

 

Available only in DSO mode.

This pushbutton opens the menu ACQUIRE which offers these 
modes:

8.1 Refresh capture/display.

In this mode repetitive signals may be captured and displayed 
much like in analog mode. The readout will show ”rfr“. Signal 
acquisition may be stopped or started with the RUN/STOP push-
button. The STOP pushbutton will be illuminated in stop.

A trigger will start a new acquisition which will overwrite the 
display of the former. The display will remain on screen until 
the next acquisition. This mode is available over the full time 
base range (50 s/cm to 5 ns/cm).

STOP

 Please note:

 

After changing the time base acquisition and display 
will always start fi rst at the trigger position which is 
undelayed (Readout: ”Tt:0s“) on the screen centre. 
With the 2

nd

 acquisition the display will start at the 

screen left. In most cases this is meaningless, but 
the scope may seem not to react at slow time base 
settings combined with long Posttrigger times. 

 

In the utmost left position of the trigger point selec-
ted with the HORIZONTAL control 

 the readout 

will indicate ”Tt:1.85ks“ for a time base setting of 
50 s/cm. This means that 1,600 seconds must elap-
se until the trace will become visible at the screen 
left, after another 250 s it will have reached the 
screen centre (1,600 s + 250 s = 1.85 ks).

8.2 Envelope capture/display

Envelope is a special mode in refresh mode, the readout will 
show ”env“. Also in this mode there must be suffi cient signal 
for triggering.

In contrast to the refresh mode the results of several captures 
will be examined and the maxima and minima stored, the enve-
lope of the signal will then be displayed if it changes in amplitude 
or/and frequency. Also any jitter will be shown. 

Also in this mode pressing RUN/STOP 

 will stop the acquisi-

tion, indicated by STOP illuminated. After pressing the pushbut-
ton again the formerly stored signals will be erased and the en-
velope calculation starts anew. In order to prevent an accidental 
turning on of this mode operating any control which infl uences 
the signal display will automatically switch envelope off.

Because this mode requires many signal repetitions and acqui-
sitions it is not compatible with single sweep/acquisition. AUTO 
or normal trigger modes must be selected.

8.3 Average mode capture/display

Also this is a special mode within the refresh mode. Also here 
signal repetitions are needed.

The weighting of each acquisition can be selected with ”Ave-
rage“ in the menu, any number between 2 and 512 may be 
chosen using the INTENS knob 

. The readout will show e.g. 

”avg#512“.

The higher the number of acquisitions averaged the lower the 
contribution of a single acquisition will be and the longer the 
averaging will take. Averaging is a means to increase the ac-
curacy inspite of the 8 bit converters, it is an exchange of time 
against accuracy. Noise  will be reduced by averaging. 

The same holds as for envelope: the acquisition may be stopped 
by pressing the RUN/STOP pushbutton, STOP will be illumina-
ted. Pressing RUN/STOP again will restart. In order to prevent 

C o n t r o l s  a n d  R e a d o u t          

EXIT MENU

INTENS

!

HM1008

ANALOG

DIGITAL

OSCILLOSCOPE

1 GSa 

·

 1 MB

100 MHz

DIGITAL

ANALOG

FOCUS
TRACE

MENU

REMOTE OFF

MATH

SAVE/

RECALL

AUTOSET

RUN

ACQUIRE

SETTINGS

HELP

STOP

REM

POWER

POWER

1

2

3

4

5

6

7

8

9

10

11

12

Содержание HM1008

Страница 1: ...100 MHz CombiScope HM1008 Manual English...

Страница 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Страница 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Страница 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Страница 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Страница 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Страница 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Страница 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Страница 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Страница 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Страница 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Страница 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Страница 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Страница 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Страница 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Страница 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Страница 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Страница 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Страница 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Страница 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Страница 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Страница 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Страница 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Страница 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Страница 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Страница 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Страница 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Страница 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Страница 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Страница 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Страница 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Страница 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Страница 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Страница 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Страница 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Страница 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Страница 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Страница 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Страница 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Страница 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Страница 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Страница 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Страница 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Страница 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...

Отзывы: