ADJUSTMENT PROCEDURE
Locate and identify R1059 (1) on PS-Board (screened section).
Locate connector J4400 ( 8pole Molex) on MB-Board and identify pin 1.
Adjust R1059 (1) for e115 Volts (± 0.1 Volt) at J4400 pin 1 with respect to chassis.
2) R1066: +12 Volt supply.
Locate and identify R1066 (2) on PS-Board.
Locate connector J4400 (8pole Molex) on MB-Board and identify pin 6.
Adjust R1066 (2) for e12 Volts (± 10mV) at J4400 pin 6 with respect to chassis.
At J4400 the other vo65V ( pin 2), -6V (pin 8), +5V (pin 4) and -1950V on the cathode
of the CRT depend on the c12 Volt adjustment. All these voltages with higher
tolerances must be checked and verified.
3) VR8003: Trace Rotation Check.
Locate and identify VR8003 (3) „TR“ on the front panel.
Press channel I GD pushbutton to ground the input (readout: „CH1:ground symbol“).
Rotate VR8003 (3) and check that the range of inclination of the baseline is at least 1mm
at both horizontal limits of the graticule.
Readjust baseline exactly parallel to the horizontal center line of the graticule.
Press channel I GD pushbutton to switch back to the basic setting.
4) R6013: CRT minimum intensity.
Locate and identify R6013 (4) on CRT-Board.
Set INTENS. control to fully left position (beep).
Press ADD-XY pushbutton briefly for XY mode (readout: both channel deflection coefficients and XY.)
Adjust R6013 (4) so that the dot just disappears.
Press ADD-XY pushbutton briefly for time base mode.
5) R6024: Astigmatism correction.
Locate and identify R6024 (5) on CRT-Board.
Press channel I GD pushbutton to ground the input (readout: „CH1:ground symbol“).
Press ADD-XY pushbutton briefly for XY mode (readout: both channel deflection coefficients and XY.)
Adjust INTENS. (front panel) for low intensity.
Turn the FOCUS knob continuously to defocus the spot in both directions from the focus point
until the next procedure is finished.
Adjust R6024 (5) so that the spot shape does not change when defocussed.
Press ADD-XY pushbutton briefly for time base mode.
6) VR9009: Y-Final Amplifier Balance.
Locate and identify VR9009 (6) on YF-Board.
Switch the oscilloscope OFF.
Locate and identify the Delay-Line with the connections 6A and 6B.
Connect 6A and 6B (short) galvanically ( wire with crocodile clips).
Switch the oscilloscope ON.
Adjust VR9009 (6) to set the trace to the horizontal center line of the graticule.
Switch the oscilloscope OFF.
Remove the short-circuiting wire.
Switch the oscilloscope ON.
Содержание HM1004
Страница 1: ... Instruments SERVICE MANUAL HM1004 Oscilloskop HM1004 ENGLISH 4S 1004 00E0 01 NOV 1996 ...
Страница 5: ...Subject to change without notice 4 HM1004 Block diagram ...
Страница 6: ...5 Subject to change without notice HM1004 Block diagram ...
Страница 7: ...Subject to change without notice 6 att14 sch 1 Tue Oct 29 12 14 03 1996 AT Board ...
Страница 8: ...7 Subject to change without notice AT Board ...
Страница 9: ...Subject to change without notice 8 AT Board ...
Страница 10: ...9 Subject to change without notice AT Board Topside ...
Страница 11: ...Subject to change without notice 10 AT Board Bottomside ...
Страница 12: ...11 Subject to change without notice MB Board ...
Страница 13: ...Subject to change without notice 12 MB Board ...
Страница 14: ...13 Subject to change without notice MB Board ...
Страница 15: ...Subject to change without notice 14 MB Board ...
Страница 16: ...15 Subject to change without notice MB Board ...
Страница 17: ...Subject to change without notice 16 MB Board ...
Страница 18: ...17 Subject to change without notice MB Board ...
Страница 19: ...Subject to change without notice 18 MB Board Topside ...
Страница 20: ...19 Subject to change without notice MB Board Bottomside ...
Страница 21: ...Subject to change without notice 20 YF Board ...
Страница 22: ...21 Subject to change without notice YF Board Top and Bottomside ...
Страница 23: ...Subject to change without notice 22 Readout Board ...
Страница 24: ...23 Subject to change without notice Readout Board rd1004 sch 2 Wed Sep 25 11 50 19 1996 ...
Страница 25: ...Subject to change without notice 24 J7012 Readout Board Bottomside ...
Страница 26: ...25 Subject to change without notice Readout Board Topside ...
Страница 27: ...Subject to change without notice 26 key1007 sch 1 Wed Sep 25 11 51 53 1996 FC Board FC Board 1004 ...
Страница 28: ...27 Subject to change without notice key1007 sch 2 Wed Sep 25 11 52 15 1996 FC Board FC Board 1004 ...
Страница 29: ...Subject to change without notice 28 FC Board Topside ...
Страница 30: ...29 Subject to change without notice Processor Board Proc Adapt Z86E63 ...
Страница 31: ...Subject to change without notice 30 FC Board Bottomside and Processor Board ...
Страница 32: ...31 Subject to change without notice IF Board Note IF304 IF1004 1505 ...
Страница 33: ...Subject to change without notice 32 IF Board Top and Bottomside ...
Страница 34: ...33 Subject to change without notice CRT Board ...
Страница 35: ...Subject to change without notice 34 CRT Board Bottom and Topside ...
Страница 36: ...35 Subject to change without notice PS Board ...
Страница 37: ...Subject to change without notice 36 PS Board Topside ...
Страница 38: ...37 Subject to change without notice PS Board Bottomside ...
Страница 42: ...41 Subject to change without notice HM1004 1505 Adjustment Procedure ...
Страница 51: ...Subject to change without notice 50 High Light Display select 22 23 21 MB Board UP DOWN ENTER ...
Страница 53: ...Subject to change without notice 52 MB Board 26 ...