ADJUSTMENT PROCEDURE
Analog Scope HM1004 / HM1505
WARNING
The Instrument must be disconnected from the mains power supply whenever you open
the case, repair or exchange parts.
HIGH VOLTAGE WARNING!
Hazardous High Voltage of up to 12,000 Volts is present inside this Instrument. The areas
particularly affected by High Voltage are the high voltage circuit on the PS-board, the CRT
anode contact and the CRT-board.
SERVICE AND ADJUSTMENT
•
of this instrument should only be performed in accordance and in conjunction with the
operating manual and the WARNINGS contained therein, particularly „Service Instructions“
and „Operating Instructions“.
•
should only be performed by suitable qualified and experienced service personnel, or should be
referred to one of the HAMEG companies listed on the rear cover of the manual.
Test Instruments required:
1)
Scope Tester HZ 60-2.
2)
Constant amplitude sinewave generator, 20Hz - 250MHz, output 5mV - 5V into 50
Ω
,
preferably with 20dB attenuation (e.g. HM 8133, TEK SG502 + TEK SG503).
3)
Amplitude Calibrator with 1kHz square wave output and 600
Ω
impedance, risetime faster
than 150ns. Output voltage 2mV - 20Volts in 1-2-5 sequence for 4 divisions display
amplitude (e.g. HZ62, TEK PG506).
4)
Time mark generator from 5ns/div to 5s/div.
Output min. 10mV into 50
Ω
(e.g. HZ62, TEK TG501).
5)
Pre-attenuator 2:1 (1 M
Ω
parallel with 12-48pF), e.g. HZ20.
6)
50
Ω
BNC through termination, e.g. HZ22.
7)
2 BNC-cables, 50
Ω
, e.g. HZ34.
8)
BNC-T-connector.
9)
Oscilloscope probe 10:1, with exactly 9M
Ω
series resistance
and compensated for test oscilloscope mentioned under 10).
10) Oscilloscope 100MHz, 5mV/div to 5V/div, e.g. HM1005.
11) Trimming/adjusting tool.
12) Variable output safety insulation transformer.
13) Video signal generator with positive and negative signal output.
This procedure covers all adjustments and the most important - but not all - performance checks.
The correct sequence of all adjustment steps must be strictly followed.
Exact adjustment is only possible when any influence of the earths‘ magnetic field has been
compensated with the trimmer marked TR (trace rotation).
All adjustments should only be performed by qualified and experienced personnel. This is particularly
important for adjustments in the high voltage section of the instrument.
NOTE
The adjustment procedures assume that the instrument had once been properly adjusted in
the factory and adjustments are required due to temperature drift or the replacement of
defective components.
Содержание HM1004
Страница 1: ... Instruments SERVICE MANUAL HM1004 Oscilloskop HM1004 ENGLISH 4S 1004 00E0 01 NOV 1996 ...
Страница 5: ...Subject to change without notice 4 HM1004 Block diagram ...
Страница 6: ...5 Subject to change without notice HM1004 Block diagram ...
Страница 7: ...Subject to change without notice 6 att14 sch 1 Tue Oct 29 12 14 03 1996 AT Board ...
Страница 8: ...7 Subject to change without notice AT Board ...
Страница 9: ...Subject to change without notice 8 AT Board ...
Страница 10: ...9 Subject to change without notice AT Board Topside ...
Страница 11: ...Subject to change without notice 10 AT Board Bottomside ...
Страница 12: ...11 Subject to change without notice MB Board ...
Страница 13: ...Subject to change without notice 12 MB Board ...
Страница 14: ...13 Subject to change without notice MB Board ...
Страница 15: ...Subject to change without notice 14 MB Board ...
Страница 16: ...15 Subject to change without notice MB Board ...
Страница 17: ...Subject to change without notice 16 MB Board ...
Страница 18: ...17 Subject to change without notice MB Board ...
Страница 19: ...Subject to change without notice 18 MB Board Topside ...
Страница 20: ...19 Subject to change without notice MB Board Bottomside ...
Страница 21: ...Subject to change without notice 20 YF Board ...
Страница 22: ...21 Subject to change without notice YF Board Top and Bottomside ...
Страница 23: ...Subject to change without notice 22 Readout Board ...
Страница 24: ...23 Subject to change without notice Readout Board rd1004 sch 2 Wed Sep 25 11 50 19 1996 ...
Страница 25: ...Subject to change without notice 24 J7012 Readout Board Bottomside ...
Страница 26: ...25 Subject to change without notice Readout Board Topside ...
Страница 27: ...Subject to change without notice 26 key1007 sch 1 Wed Sep 25 11 51 53 1996 FC Board FC Board 1004 ...
Страница 28: ...27 Subject to change without notice key1007 sch 2 Wed Sep 25 11 52 15 1996 FC Board FC Board 1004 ...
Страница 29: ...Subject to change without notice 28 FC Board Topside ...
Страница 30: ...29 Subject to change without notice Processor Board Proc Adapt Z86E63 ...
Страница 31: ...Subject to change without notice 30 FC Board Bottomside and Processor Board ...
Страница 32: ...31 Subject to change without notice IF Board Note IF304 IF1004 1505 ...
Страница 33: ...Subject to change without notice 32 IF Board Top and Bottomside ...
Страница 34: ...33 Subject to change without notice CRT Board ...
Страница 35: ...Subject to change without notice 34 CRT Board Bottom and Topside ...
Страница 36: ...35 Subject to change without notice PS Board ...
Страница 37: ...Subject to change without notice 36 PS Board Topside ...
Страница 38: ...37 Subject to change without notice PS Board Bottomside ...
Страница 42: ...41 Subject to change without notice HM1004 1505 Adjustment Procedure ...
Страница 51: ...Subject to change without notice 50 High Light Display select 22 23 21 MB Board UP DOWN ENTER ...
Страница 53: ...Subject to change without notice 52 MB Board 26 ...