Voltage measurements

 

The maximum input voltage of HM8012 
is 600 V

DC

 when the COM terminal is 

to  ground  potential,  i.e.  by  connecting 
HM8012 to the object to be measured, 
the sum of the measurement voltage and 
that of the COM terminal with respect 
to ground shall not exceed 600 V

DC

.  In 

this case, the maximum voltage value 
between the COM terminal and ground 
is 250 V

DC

.

For AC voltages, the true rms value of the input 
voltage will be measured, and the DC component 
eliminated in AC mode. If possible, the COM ter-
minal 

 shall be connected directly to ground or 

to the point of the measurement circuit having 
the lowest potential. The 0.5 V and 5 V voltage 
measurement ranges are protected from input 
voltages to 300 V

rms

; all the other ranges are 

protected to 850 V

rms

.

During measurements on circuits using inductive 
components, inadmissible high voltages may 
appear when the circuit is opened. In such cases, 
take steps to prevent the destruction of HM8012 
by inductive voltages.

Input impedance in the DC range

To make the most of the excellent linearity of 
the measurement system, the input impedance 
for voltage measurements is very high (1 G  ) for 
some ranges. For instance, this makes possible 
to perform accurate measurements on ranges of 
up to ±5 V, even when the internal impedance of 
the source to be measured is high. For instance, 
for the 500 mV range, an internal 5 M   source resi-
stance will induce a maximum error of 150 μV.

Indeed, switching to the 50 V range will cause the in-
put voltage to drop because of the input impedance 
of 10 M   which can cause the multimeter to switch 
to the lower range and so on. After some swings, 
the instrument goes into MANUAL mode.

Current measurement 

For current measurements, the connection of the 
object to be measured is made across the mA 

 

terminal or to the A 

 terminal for currents of up 

to 10 A. The AUTO mode is inhibited because there 
is only one range.
 

The HM8012 should be connected to the circuit 
whose potential with respect to ground is lowest. 
For safety reasons, the COM terminal 

 must 

not exceed 250 V

DC

 with respect to ground. 

Current ranges are protected by fuse against over-
loads. If the fuse blows, the cause of the overload 
must be eliminated. In case of an open fuse the 
instrument has to be sent in for repair. A change of 
the fuse by the customer is not permitted. 

AC voltage measurement 

The instrument measures the true rms value of 
the input voltage with or without its DC component. 
To measure low voltages, or in the event of high 
noise, it is necessary to use a shielded cable.
Take into account the input impedance of the mul-
timeter. It is 1 M   in the AC mode and 10 M   in the 
AC + DC mode. In addition, there is a slight measu-
rement difference between these two modes due 
to the input circuits. If AC measurements without 
a DC component are to be made, it is preferable 
to use the straight AC mode.
When the multimeter is used in AUTOMATIC mode, 
there can be continuous swinging between two 
ranges for frequencies above 30 kHz, because 
of the frequency response difference of the two 
ranges. After some swings, the instrument goes 
into MANUAL mode.
Measuring errors can be caused by an overload of 
the measuring amplifi ers and the A/D converter. 
For measurements of mixing tensions (positive or 
negative DC voltage overlaid with an alternating 
voltage) select a measuring range which is not 
exceeded by the peak value of the alternating 
voltage – the largest deviation from the reference 
potential (normally 0 V). That applies also, if only 
the DC voltage portion is to be measured. 

A DC voltage of 350 mV is superimposed with a 
sine signal with an amplitude of 200 mV

p

 (400 

mV

pp

). The highest deviation from 0 V amounts 

550 mV

p

 (350 mV

p

 + 200 mV

p

). For DC and AC+DC 

measurements the 5 V-range must be selected. 
AC measurements can be done in the 500 mV 
range, because then a capacitor in the entrance 
suppresses the DC voltage. It is reasonable, fi rst 
to determine the height of the possibly existing 
alternating voltage with AC measurement. Then 
switch to DC or AC+DC function and consider the 

          F u n c t i o n s

Содержание HM 8012

Страница 1: ...tion 10 V 0 01dBm 10nA 10m 0 1 C F Offset function relative value measurement in basicmeasurement functions Input impedance 1G 0 5V and 5V DC range RS 232 interface PC software for control and data lo...

Страница 2: ...curacy 23 8 59 8dBm 0 2dBm Resolution 0 01dB above 9mV CMRR3 60dB 50 60Hz 0 5 Crest factor 7 max AC current Measurement ranges 500 A 5mA 50mA 500mA 10A Resolution 10nA 100nA 1 A 10 A 1mA Accuracy 0 5...

Страница 3: ...r discon nection of the protective earth terminal is likely to render the instrument dangerous Intentional interruption is prohibited The istrument must be disconnected and secured against unintention...

Страница 4: ...ould be taken to ensure that it properly ts under the edges of the front and rear frames After removal of the two screws at the rear of the module both chassis covers can be lifted When reclosing the...

Страница 5: ...in con junction with the COM input low potential Current in excess of 10 A max 20 A must not be applied for a period exceeding 30 s otherwise the internal measurement resistor thermal device will blo...

Страница 6: ...er range On each press the new range is displayed eetingly on the display in code form L1 for lowest range L2 for second range etc pushbutton Pushbutton for changing to higher range Each time pressed...

Страница 7: ...ower range is obtained when the value drops below 4 900 counts It is possible to know which range has been selected by the unit by temporarily deactivating the AUTO mode so that the range indication c...

Страница 8: ...8012 should be connected to the circuit whosepotentialwithrespecttogroundislowest For safety reasons the COM terminal must not exceed 250VDC with respect to ground Currentrangesareprotectedbyfuseagain...

Страница 9: ...erpreting measurement values and for evaluating its accuracy It is de ned as the ratio between the peak voltage and the rms value of the signal Crest factor CF Vp VRMS It is a measurement of the dynam...

Страница 10: ...temperature of 0 C and using the OFFSET function Note Measurement of the temperature probe line resistance should be made on the same in strument used for temperature measurements Only this procedure...

Страница 11: ...eys on the front panel AY CR switches to automatic range change AN CR switches to manual range change R CR switches to the next higher range R CR switches to the next lower range If it is impossible t...

Страница 12: ...mA Diode 3 50 V 50 k 50 mA 4 500 V 500 k 500 mA 5 1000 V 5 M 6 50 M 10 A P CR This command alone is used for reco vering complete parameter settings of the equipment The instrument returns string_F s...

Страница 13: ...500 k 5M 0 02 for instance models CNS020 by Vishay Test procedure If one of the indicated calibrators is available or if precision calibrators are appropriate all the HM8012 measurements ranges can be...

Страница 14: ...guration This one can be saved and recalled later When the con guration of the instrument is done a set of measurements can be carried out and saved for future use Furthermore the software can show d...

Отзывы: