21
Subject to change without notice
S p e c i f i c a t i o n s
1)
rdg. = reading;
3)
= Common mode rejection ratio;
2)
fs. = full scale;
SPECIFICATIONS
valid at 23 deg. C after a 30 minute warm-up period
DC voltages
Measurement ranges:
500 mV, 5 V, 50 V, 500 V, 600 V
Resolution:
10 μV, 100 μV, 1 mV, 10 mV, 100 mV
Accuracy:
5 V, 500 V, 600 V:
±(0.05% of rdg.
1)
+0.002% of fs.
2)
)
500 mV, 50 V:
± (0.05% of rdg.
1)
+ 0.004% of fs.
2)
)
Overload protection:
V/
/T°/dB/
against COM and to chassis:
850
V
p
at max.60 Hz or 600 V
DC
Low against chassis:
250
V
rms
at max. 60 Hz or 250 V
DC
Input resistance:
50 V, 500 V, 600 V:
10 M II 90 pF
500 mV, 5 V:
>
1 G II 90 pF
Input current:
10 A
CMRR
3)
:
≥
100 dB (50/60 Hz ± 0.5%)
SMRR
4)
≥
60 dB (50/60 Hz ± 0.5%)
dB Mode
Accuracy:
±(0,02 dB + 2 digits)
(display
>
–38,7 dBm)
Resolution:
0.01 dB above 18% of fs.
2)
DC current
Measurement ranges:
500 μA, 5 mA, 50 mA, 500 mA, 10 A
Resolution:
10 nA, 100 nA, 1 μA, 10 μA, 1 mA
Accuracy:
0,5 – 500mA:
±(0.2% of rdg.
1)
+ 0.004% of fs.
2)
),
10
A:
±(0.3%
of rdg.
1)
+ 0.004% of fs.
2)
)
Voltage drop:
10 A range:
0.2 V max.
500 mA range:
2.5 V max.
other ranges:
0.7 V max.
AC voltage
Measurement ranges:
500 mV, 5 V, 50 V, 500 V, 600 V
Resolution:
10μV, 100 μV, 1 mV, 10 mV, 100 mV
Accuracy 0,5-50 V:
40 Hz - 5 kHz:
± (0.4% of rdg.
1)
+ 0.07% of fs.
2)
)
20 Hz - 20 kHz:
± (1% of rdg.
1)
+ 0.07% of fs.
2)
)
500 V and 600 V:
40 Hz - 1 kHz:
± (0.4% of rdg.
1)
+ 0.07% of fs.
2)
)
20 Hz - 1 kHz:
± (1% of rdg.
1)
+ 0.07% of fs.
.2)
)
Overload protection:
V/
/T°/dB/
against COM and to chassis:
850
V
s
at max.60 Hz or 600 V
DC
Low against chassis:
250 V
eff
at max. 60 Hz or 250 V
DC
Input impedance
AC mode:
1
M II 90pF
AC + DC mode:
10 M II 90pF
Bandwidth at -3 dB:
80
kHz
typical
dB Mode:
20 Hz - 20 kHz
Accuracy (–23.8 - 59.8 dBm):
±0.2 dBm
Resolution:
0.01
dB
above
9
mV
CMRR
3)
:
≥
60 dB (50/60 Hz ± 0.5%)
Crest factor:
7
max.
AC current
Measurement ranges:
500 μA, 5 mA, 50 mA, 500 mA, 10 A
Resolution:
10 nA, 100 nA, 1 μA, 10 μA, 1 mA
Accuracy:
0.5-500 mA: ±(0.7% of rdg.
1)
+0.07% of fs.
2)
) for 40Hz-5kHz
10A:
± (1% of rdg.
1)
+ 0.07% of fs.
2)
)
AC + DC measurements
Same as AC + 25 Digits
Resistance
Ranges:
500
, 5 k , 50 k , 500 k , 5 M , 50 M
Resolution:
10
m , 100 m , 1 , 10 , 100 , 1 k
Accuracy:
500 to 500 k :
±(0.05% of rdg.
1)
+0.004% of fs.
2)
+50m )
5
M and 50 M :
±(0.3% of rdg.
1)
+0.004% of fs.
2)
)
Input protected to max. 300 V
eff
Measurement current
500 – 5 k -range: 1 mA
50
k -range: 100
μA
500
k -range:
10
μA
5 - 50 M -range:
100 nA
Measurement voltage
10 V typical for open inputs, depending on the value of
resistance to be measured. Negative polarity of measurement
voltage is across common terminal.
Temperature
2-wire resistance measurement with linearization for
PT100 sensors as per standard EN60751
Range:
–200 °C to +500 °C
Resolution:
0.1°C
Measurement current:
ca. 1 mA
Display:
in °C, °F
Accuracy:
–200 °C to +200 °C
± 0.4 °C + 0,0005 x T
+200 °C to +500 °C
± 0.5 °C + 0,0005 x T
(T in °C, sensor tolerance not included)
Temperature coeffi cient: (Reference 23°C)
V =
500 mV, 50 V
30 ppm/°C
600 V range
80 ppm/°C
other
ranges
20
ppm/°C
V ~
600 V range
80 ppm/°C
other
ranges
50
ppm/°C
mA all
ranges 200
ppm/°C
mA-
all ranges
300 ppm/°C
5 M , 50 M -ranges 200 ppm/°C
other
ranges
50
ppm/°C
Miscellaneous
Operating conditions:
+ 10°C to + 40°C
max. rel. humidity:
80% (without condensation)
Power supply (from HM8001-2):
+ 5 V
300 mA
~26 V
140 mA
Size (W x H x D):
135 x 68 x 228 mm
(without
fl at 22-pole connector)
Weight:
ca.
500g
Included in delivery:
HM8012 Multimeter, Manu-
al, HZ14 Interface cable, HZ15 PVC Test Leads,
Software-CD.
Accessories:
HZ10 Silicone Test
Leads, HZ812 PT100 Temperature Sensor