Configuring the R-1307 for a Local Target
Note:
Shut off power to the readout before connecting or disconnecting a target from the local port.
1.
Connect the cabled target to the local port of the readout
2.
Press and hold the MENU button for approximately 2 seconds to enter configuration mode.
3.
Set the Measurement Units
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow keys
to select either
for inches or
for millimeters.
4.
Set the Dampening Level
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow keys to
set the number of averages. Adjust this value as required to suit the application. The default for this
application should be changed to at least 8. For long distance shots, use 16 or 32.
5.
Set the Readout Function to Local Target
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow keys
to select
.
6.
Select the PSD descriptor applicable to your target
Press the MENU button until the upper display shows
, where
designates the target
number. Each R-1307 has three target descriptors:
TGT=0 (for HLI use only. Do not use )
TGT = nn, P.10.10 (pulsed beam mode)
TGT = nn, F.10.10 (fixed beam mode)
nn= R-1037 Readout number and Matching Target number
Press the UP or DOWN arrow to select the correct target number and to change the second window.
For example,
.
.
or
.
.
for R-1307 #2
WARNING:
Targets are matched to specific R-1307 Readouts. For example, Target #1 must be
connected to Readout #1 or the calibration is void.
7.
To exit configuration mode, press and hold the MENU button for approximately three seconds
until
the display returns to normal mode.
The R-1307 will also return to normal mode automatically after approximately four seconds of
inactivity.