background image

25 

 

Avg. Mode 

 

 

Normal Running Average of 

n

 Samples (Default) 

 

Reset Average if Position Change Exceeds STDEV (Standard Deviation) 

 

Reset Average if Position Change Exceeds 2*STDEV (2X Standard Deviation) 

 

Reset Average if Position Change Exceeds 4*STDEV (4X Standard Deviation) 

 

Reset Average if Position Change Exceeds 8*STDEV (8X Standard Deviation) 

 

Reset Average if Position Change Exceeds 4 

μm (4 M

icrons) 

 

Reset Average if Position Change Exceeds 8 μm (8 Microns)

 

 

Reset Average if Position Change Exceeds 16 μm (16 Microns)

 

 

Reset Average if Position Change Exceeds 32 μm (32 Microns)

 

 

Reset Average if Position Change Exceeds 64 μm (64 Microns)

 

Note: 

Reset Position Modes should be used only for very stable environments. 

 

Pos. Sensing 

 

The Unitarget utility is 

not

  compatible with the Type I (IR or radio) Unitargets.  The utility is only 

intended for use with Type II A-1519 and A-1520 Unitargets. The Type II Unitargets have a USB and an 

AUX port on the left side of the target and a circular connector in the back. The Type I Unitarget is 

smaller and only has a single circular connector in the back. 

 

There are two types of Type II Unitargets in the field. The A-1519 targets can be identified by the sensor 

type, visible through the front window as shown 

in Figure 44. 

 

Type II-Legacy, Single Axis Sensor 

 

Compatible with radio version of the 

Type I Unitarget. 

 

Supports all scanning lasers. 

 

Does 

not

  support fixed (continuous) or 

pulsed line lasers. 

 

Type II-Current Version, Dual Axis Sensor 

 

Supports all scanning lasers. 

 

Supports fixed (continuous) and pulsed 

line lasers. 

 

Configure the target using the utility, based on 

the laser type and Type II Unitarget as follows: 

 

Scanning, Single Axis Legacy Mode: 

Select 

this mode for Type II Legacy Targets 

only

. Select 

for use with scanning lasers only.   

Scanning, Dual Axis  Mode: 

Select this mode 

for Type II, current version. Select for use with 

scanning lasers only. 

Fixed Beam, Dual Axis: 

Select this mode for all continuous line lasers (for example the L-700 and L-

705 in continuous mode). 

Pulsed Beam, Dual Axis: 

Select this mode for use with pulsed line lasers only (for example, the L-705 

in pulsed beam mode).  

 

Figure 44 – 

A-1519 Single and Dual-Axis Sensors 

Содержание A-1519 Series

Страница 1: ......

Страница 2: ...ticles of equipment manufactured by other companies and limited by a warranty extending for less than one year In such an event the more limited warranty applies to said instrument or article of equip...

Страница 3: ...he A 1519 1520 Type II Configuration Utility 14 Installation 14 Users Database 14 Initial Setup 14 Editing the User Database 15 Logging In 16 Startup 17 Troubleshooting 18 Switching Between Normal App...

Страница 4: ...nal temperature sensor Improved Background light noise suppression Internal dampening by averaging of 2 to 64 laser position scans Identifying Target Features 1 Right Side Auxiliary Ports 2 Front View...

Страница 5: ...battery life by approximately 10 percent Note The shelf life of lithium polymer batteries is 2 3 years under optimal conditions Expect the battery capacity to drop to 80 percent after 500 charge disc...

Страница 6: ...307W 2 4ZB is designed to suit a user s specific needs with the R 1307 2 4ZB and the R 1307W 2 4 ZB models featuring ZigBee radio technology The readout is available with a radio frequency of either 9...

Страница 7: ...adequate power 5V 400 mA through the USB port Note When using the USB Extender cable extension kit plug the A 910 2 4 into an A C power supply 5 Antenna See Appendix C on Page 30 for radio specificat...

Страница 8: ...s with the R 1307 2 4ZB models and plugs directly into any unused computer USB port to communicate directly with the targets Utility software is provided for driver installation setup and configuratio...

Страница 9: ...e Readout Function to Local Target Press the MENU button until the upper display shows Use the UP and DOWN arrow keys to select 6 Select the PSD descriptor applicable to your target Press the MENU but...

Страница 10: ...display shows with the current System ID blinking Use the UP and DOWN arrow keys to set the System ID Note that nn must be set to the same number as the channel switch setting of the A 910 radio trans...

Страница 11: ...gets and one Radio Transceiver as a System Group 1 Set all targets to the same System ID ranging from 1 to 9 2 Set each target in the System Group to a different Target Network ID ranging from 01 to 9...

Страница 12: ...er Host computer control the radio transceiver transmits a message called a polling request that contains the Target Network ID of one specific target All targets receive all polling requests but only...

Страница 13: ...or easy removal and placement 1 Mount the Charging Station to the wall or any other suitable vertical support structure using two screws of the appropriate type for the mounting structure refer to Fig...

Страница 14: ...ly the included Hamar Laser power supply Use of another power supply may void the warranty of the Charging Station and targets Input 120VAC 60Hz US Output 7 5VDC 2A Charging Rate 400mAH Cradle Target...

Страница 15: ...the CD ROM icon and open it Click the USB Driver folder see Figure 13 3 Locate the Preinstaller exe icon and click to initiate the installation process The Install Driver dialog box displays see Figur...

Страница 16: ...Transceiver and then to any available USB port on the computer The Windows Found New Hardware Wizard see Figure 16 and Figure 17 displays a number of times Each time the Wizard displays select Install...

Страница 17: ...iles always keep the newest files 6 After Setup completes click Start Program Files Hamar Laser Utilities Click A 1519 Type II Configuration Utility to start the program Users Database The Configurati...

Страница 18: ...cannot make changes to the Primary Administrator account Power User This user can access to all program features but does not have permission to access the Users Database Operator This user has acces...

Страница 19: ...Database and a valid password for that user the OK button is enabled to indicate that the password is valid Note that the Users button is also enabled because user Admin2 is a Backup Administrator wi...

Страница 20: ...roximately once every three seconds Figure 27 Target detected in Configuration Mode Target Connected in Application Mode Figure 28 The target must be in Application Mode to be used with standard appli...

Страница 21: ...et Ensure that the target is ON The center LEDs see Figure 1 Front View 2 1 should be lit continuously if the target is in Application Mode or blinking once every three seconds if the target is in Con...

Страница 22: ...The Periodicity Sync indicator _ The Incident Light Level bar _ The Battery Charge indicator Center Packet data Green fields indicate normal values Yellow fields indicate caution values are borderline...

Страница 23: ...ion Sensor readings Power Supply Voltages Beam on Sensor Time and Incident Noise Characteristics 1 Set the target to DEBUG B Packet Mode see Figure 35 The Diagnostic Readout window displays see Figure...

Страница 24: ...with an xls extension you will be able to import the data directly into Microsoft Excel 5 Click Stop Polling Target to stop saving data Dynamic Plot Check Underlay Position Plot The position readings...

Страница 25: ...tility to finish processing the file for uploading to the target see Figure 40 When the file processing completes verify that you want to update the target firmware with the specified file version see...

Страница 26: ...e the firmware upload is in progress see Figure 41 When the upload completes an EOF End of File verification message displays see Figure 42 Click OK Figure 41 Target firmware upload in progress Figure...

Страница 27: ...ght sources such as fluorescent lighting fluctuates at the AC line frequency or twice the AC line frequency Set the Light Periodicity to the AC line frequency at the location where the target will be...

Страница 28: ...in the back The Type I Unitarget is smaller and only has a single circular connector in the back There are two types of Type II Unitargets in the field The A 1519 targets can be identified by the sens...

Страница 29: ...er most assistance is limited to information regarding the structure of the calibration file Hamar Laser Calibration Method The A 1519 A 1520 targets are calibrated using a proprietary automated stati...

Страница 30: ...Number Sensor Resolution Linearized Accuracy Range A 1519 0 00002 in 0 0005 mm 0 00015 in 0 0035 mm 1 0 in 24 mm A 1520 0 00001 in 0 00025 mm 0 00006 in 0 0015 mm 0 2 in 5 mm Range may be limited by o...

Страница 31: ...d This makes is very likely that the error in measurement at the point where the target is zeroed is nearly identical to the error in each subsequent measurement because the difference in sensor area...

Страница 32: ...ea is 1 micron or less Approx 4 mm of measuring area each vertical line on the graph equals 1 mm Total Error 3 3 microns 2 microns 2 microns 0 000 Each green vertical line equals 1 mm of measuring are...

Страница 33: ...t recommended IC Industry Canada Certification Contains Model 24XStream Radio 2 4 GHz IC 4214A 12008 Complies with IC ICES 003 Complies with ETSI France France imposes restrictions on the 2 4 GHz band...

Страница 34: ...ustry Canada Certification Contains Model 9XCite Radio 900 MHz IC 4214A 9XCITE Agency Certifications for the XBee 802 15 4 Series 1 FCC United States of America Certification Contains FCC ID OUR XBEE...

Страница 35: ...32 Appendix D Dimensional Drawings A 1519 Type II Universal Wireless Target All dimensions in inches...

Страница 36: ...33 A 1519 RS 485 Cable Diagram...

Страница 37: ...34 Appendix E Using the A 151945 4 Port Target Communication Wired Hub...

Страница 38: ...ter than or equal to the scan rate For most Hamar Laser scanning lasers the recommended polling rate is 4 times second 250 milliseconds interval Polling the same target at a rate faster than the laser...

Страница 39: ...Information Page 38 VP VERTICAL POSITION COUNTS 32768 TO 32767 1 m 2 COUNTS A 1519 1 m 4 COUNTS A 1520 VCO SENSOR CENTER OFFSET 4000 TO 4000 1 m 2 COUNTS A 1519 1 m 4 COUNTS A 1520 BAT BATTERY VOLTAGE...

Страница 40: ...767 1 m 2 COUNTS A 1519 1 m 4 COUNTS A 1520 VCO VERTICAL CENTER OFFSET 4000 TO 4000 1 m 2 COUNTS A 1519 1 m 4 COUNTS A 1520 BAT BATTERY VOLTAGE 0 TO 5000 MILLIVOLTS TEMP INTERNAL TEMPERATURE COUNTS 16...

Страница 41: ...SER DETECTION STATUS BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0 0 1 1 1 0 1 0 0 BIT 7 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 BIT 6 0 0 0 0 1 1 1 1 0 0 0 0 1 1 1 1 BIT 5 0 0 1 1 0 0 1 1 0 0 1 1 0 0 1 1 B...

Страница 42: ...a small screwdriver Check A C power source Use second A C power supply if available 5 Noisy target A 1519 1520 readings on PDA NOTE On average the user can expect 00002 in ft to 00005 in ft 0 0025 mm...

Страница 43: ...ion is connected to USB port Ensure the correct COM port is selected for the USB bridge controller see Windows Device Manager must be lower than COM10 Check Device Manager in Windows Control Panel Set...

Страница 44: ...s connected to the laptop s USB port If problem persists contact HLI 13 PDA locks up Hit RESET switch Dell PDA RESET is on right hand side of the back of PDA near the lower right corner of the radio m...

Страница 45: ...the type intended for nail polish removal NOTE When cleaning precision optics even with the best quality optical cleaning tissue use gentle pressure to avoid scratching the surface or damaging the op...

Страница 46: ...ows 2000 Windows XP etc and open the folder 5 Locate the CP210x_VCP icon and click to initiate the installation process The Install Driver dialog box displays 6 Click Browse to select an installation...

Страница 47: ...utility does not automatically detect the COM port it must be manually selected 3 The Target System ID or R 1307 CH channel is the number associated with the A 1519 1520 targets or R 1307 Readout If...

Страница 48: ...button and select Control Panel 2 Click the System icon 3 In the System window click on the Device Manager link located under the System heading 4 In Device Manager scroll down to Ports Expand the li...

Страница 49: ...e background light causes a warning to display on the PDA an optional light shield may be installed on the A 1519 Target as follows 1 Roll each set screw in Loctite 2 Install set screws in the holes o...

Отзывы: