C11440-42U/C11440-42U01 Instruction manual_Ver.1.1
■
Calculation formula (Rapid Rolling Mode)
Vn = Number of vertical line
(The center of the set area is the middle of the sensor.)
Exp1 = 1 ms to 10 s
Exp2 = 1.05 ms to 10 s
1H
= 10 × 10
-6
int () = The decimal point is rounded down.
•
The Exp1 and Exp2 values must be input to the calculation formula below in units of
seconds.
Binning
Horizontal
width
Vertical
width
Calculation formula
1×1
1024
1536
2048
-
1/(int(Vn/2048/30/1H)×1H)
1×1 512
-
Free running mode
2×2
4×4
-
-
1/(Vn/2×1H)
1×1 112
≦
Vn
≦
2048
1/(Vn/2×1H+Exp1+1H×9)
1×1
1024
1536
2048
8
≦
Vn
≦
104
1×1 512
-
External trigger mode
(Edge trigger)
2×2
4×4
-
-
1/(Vn/2×1H+1H×9+Exp1)
1×1
1024
1536
2048
112
≦
Vn
≦
2048
1/(int(Vn/2048/30/1H)×1H)
1×1
1024
1536
2048
8
≦
Vn
≦
104
1×1 512
-
External trigger mode
(Level trigger)
2×2
4×4
-
-
1/(Vn/2×1H+1H×4+Exp2)
1×1
1024
1536
2048
32
≦
Vn
≦
2048
1/(int(Vn/2048/30/1H)×1H)
1×1
64
1024
1536
2048
8
≦
Vn
≦
24
1×1 512
-
External trigger mode
(Synchronous readout)
2×2
4×4
-
-
1/(Vn/2×1H+1H×17)
20