5  Start-up 

5.1  Electrical connections 

 
 

Use only IEC 6 LR 61 batteries. 

 

 

Open the battery compartment on the rear of the instrument by sliding off the 
compartment cover. Plug the batteries in to the battery hook-up, making sure that the 
battery terminals are properly aligned. Place the battery in the compartment and 
replace the cover. 
 

5.2  Pneumatic connections 

Connect the EMA 200 to the pressure source at the overpressure port (+) when 
measuring overpressure, at the vacuum port (-) for measuring vacuum pressure, or at 
both ports (making certain to connect the + and - ports appropriately) for measuring 
differential pressure. 
 

6  Operating the instrument 

The following sections explain how to operate the instrument correctly.  
 

6.1  Switching the instrument on/off 

 

Turn the instrument on or off by briefly depressing the  

       key. The 

instrument displays its measurement range in mbar for roughly 1 s 
immediately after it is switched on.

 

 

6.2  Display modes: measurement, hold, max., min. 

 
 

The          key allows the user to toggle between the various display modes 
Pressing this key once will call up a different display mode. In hold, max and 
min modes, users can begin a new measurement by briefly depressing the 
zeroing               key. In the display modes min and max, this will reset the  
 
display value, because the actual value will be the new min or max value.  
 

Содержание EMA 200

Страница 1: ...ssure Gauge halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup walcher com Internet www halstrup walcher com Doc...

Страница 2: ...ng the instrument on off 7 6 2 Display modes measurement hold max min 7 6 3 Changing pressure units 8 6 4 Zeroing key 8 6 5 Analogue output 9 7 Configuration 9 7 1 Setting units of pressure 9 7 2 Temp...

Страница 3: ...out the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials provided b...

Страница 4: ...the rating plate and in the Technical data section of this manual The instrument may only be handled as indicated in this manual Modifications to the instrument are prohibited The manufacturer is not...

Страница 5: ...warns you of a potential hazard that could lead to bodily injury up to and including death if the corresponding instructions are not followed WARNING This warns you of a potential hazard that could le...

Страница 6: ...ng units At its heart is a pressure measurement capsule with a beryllium bronze membrane spring which is displaced by the pressure difference between the two chambers of the measurement capsule Measur...

Страница 7: ...onnect the and ports appropriately for measuring differential pressure 6 Operating the instrument The following sections explain how to operate the instrument correctly 6 1 Switching the instrument on...

Страница 8: ...l use the new units until it is switched on off or until the next time the units are changed The units selected during configuration see section 7 are always set when the instrument is switched on 6 4...

Страница 9: ...point key until ConF appears in the top line of the display After the operator releases this key the instrument launches the first item setting units of pressure To exit the configuration menu simply...

Страница 10: ...pressing the and keys Briefly depressing the key will exit this mode If the flow rate display has been activated the menu for setting the density of the medium to be measured appears if not the next...

Страница 11: ...ween 20 and 100 of the measurement range While operators may enter final pressures 20 doing so will decrease the DA resolution The bottom line of the display will show the code AnH Selects which digit...

Страница 12: ...s an auto on off function that can be used to avoid consuming battery power unnecessarily This feature which can be toggled on and off will power down the instrument after 20 minutes if the pressure d...

Страница 13: ...s parameter is used to reset the instrument to its default settings if desired The bottom line of the display will show the code rES Turns this function on yes or off no Turns this function on yes or...

Страница 14: ...nstall new battery use a 9V IEC 6 LR 61 battery instrument returns false values zero point has shifted battery is low adjust zero point see section 4 4 Batt appears in the display field i e replace ba...

Страница 15: ...in the meas range 3 place LCD for flow rate temp measurement resolution 1 C accuracy 1 C measurement range 0 60 C analogue output max output voltage 3 3V permissible load 2k temperature range storage...

Страница 16: ...pressure changes by more than 2 of the final value or the operator presses a button Each time this happens the wait period is reset to 20 minutes Appendix A Parts in contact with measurement medium B...

Страница 17: ...EMA 200 Instruction Manual 17 10 Drawings analogue output 3 5 mm jack...

Отзывы: