EMA 200 Instruction Manual 

 

10 

 

7.2  Temperature display 

The symbol '

°C

' is shown in the bottom line of the display and the current 

status of this switch (on or off) is shown in the top line. Users may change the  

status by depressing the              and               keys, and may finish entering  

their settings by pressing the              key. The temperature display will 

automatically switch off if, during the subsequent step, the flow rate display is 

activated. (The instrument can only display either temperature or flow rate, but 

not both.)  

 

7.3  Flow-rate display(not available with 200kPa Range) 

'SPd' (

SP

ee

d

) will now appear in the bottom line of the display to indicate the 

function that controls the flow-rate display. The current status of this switch (on  

or off), shown in the top line, may be altered by pressing the            and             

keys.  Briefly depressing the               key will exit this mode. If the flow-rate 

display has been activated, the menu for setting the density of the medium to 

be measured appears; if not, the next step is to set the scale for the analogue 

output (section 7.6). 

7.4  Media density (not available with 200kPa Range) 

This configuration parameter is represented in the bottom line of the display by 
the symbol 'ro'. The top line displays the density as it is currently set and the 
instrument waits for the user to enter the density of the medium to be 
measured. The keypad buttons perform the following functions: 
 
 

Selects which digit of the display is to be changed (digit will flash).  

 

Increases the selected digit in increments of +1 (0...9

0). If the set next 

value exceeds the limits of the setting range, it is not added onto the 
highest digit (no overflow); the digit in question is instead set back to 0. 

 
Exits the density entry mode and continues on to the pitot factor. 

Содержание EMA 200

Страница 1: ...ssure Gauge halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup walcher com Internet www halstrup walcher com Doc...

Страница 2: ...ng the instrument on off 7 6 2 Display modes measurement hold max min 7 6 3 Changing pressure units 8 6 4 Zeroing key 8 6 5 Analogue output 9 7 Configuration 9 7 1 Setting units of pressure 9 7 2 Temp...

Страница 3: ...out the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials provided b...

Страница 4: ...the rating plate and in the Technical data section of this manual The instrument may only be handled as indicated in this manual Modifications to the instrument are prohibited The manufacturer is not...

Страница 5: ...warns you of a potential hazard that could lead to bodily injury up to and including death if the corresponding instructions are not followed WARNING This warns you of a potential hazard that could le...

Страница 6: ...ng units At its heart is a pressure measurement capsule with a beryllium bronze membrane spring which is displaced by the pressure difference between the two chambers of the measurement capsule Measur...

Страница 7: ...onnect the and ports appropriately for measuring differential pressure 6 Operating the instrument The following sections explain how to operate the instrument correctly 6 1 Switching the instrument on...

Страница 8: ...l use the new units until it is switched on off or until the next time the units are changed The units selected during configuration see section 7 are always set when the instrument is switched on 6 4...

Страница 9: ...point key until ConF appears in the top line of the display After the operator releases this key the instrument launches the first item setting units of pressure To exit the configuration menu simply...

Страница 10: ...pressing the and keys Briefly depressing the key will exit this mode If the flow rate display has been activated the menu for setting the density of the medium to be measured appears if not the next...

Страница 11: ...ween 20 and 100 of the measurement range While operators may enter final pressures 20 doing so will decrease the DA resolution The bottom line of the display will show the code AnH Selects which digit...

Страница 12: ...s an auto on off function that can be used to avoid consuming battery power unnecessarily This feature which can be toggled on and off will power down the instrument after 20 minutes if the pressure d...

Страница 13: ...s parameter is used to reset the instrument to its default settings if desired The bottom line of the display will show the code rES Turns this function on yes or off no Turns this function on yes or...

Страница 14: ...nstall new battery use a 9V IEC 6 LR 61 battery instrument returns false values zero point has shifted battery is low adjust zero point see section 4 4 Batt appears in the display field i e replace ba...

Страница 15: ...in the meas range 3 place LCD for flow rate temp measurement resolution 1 C accuracy 1 C measurement range 0 60 C analogue output max output voltage 3 3V permissible load 2k temperature range storage...

Страница 16: ...pressure changes by more than 2 of the final value or the operator presses a button Each time this happens the wait period is reset to 20 minutes Appendix A Parts in contact with measurement medium B...

Страница 17: ...EMA 200 Instruction Manual 17 10 Drawings analogue output 3 5 mm jack...

Отзывы: