MP5A Operation and Maintenance Manual
________________________________________________________________________________________________________________
Copyright
©
2023
‐
2024
Halogen
Systems,
Inc.
All
rights
reserved
worldwide.
Rev.:
4.02
39
Slave
Address
Change
Address
Bus
Search
Modbus
Enabled
Maintenance
Stop
Replace
pH
sensor
reset
hours
Long
Menu
Replace
wear
parts
reset
hours
Diagnostics
Set
pH/Cond/CHL
Set
CHL
zero
and
Slope
CHL
Zero
CHL
Slope
Conductivity
Cond
Zero
Cond
Slope
pH
Zero
&
Slope
Edit
Slope
Edit
Offset
Logs
View
last
log
Erase
logs
Clear
Errors
13.
Other Installation Methods
13.1.
Pipe Installation
Locate
the
sensor
in
a
straight
length
of
pipe
at
least
1.5
times
the
pipe
diameter
both
in
before
and
after
the
sensor.
The
sensor
should
not
be
exposed
to
large
debris.
Ideally,
the
sensor
should
be
mounted
in
a
horizontal
position.
The
Sensor
Adapter
(SD
‐
40
‐
010)
should
be
installed
in
a
2”
pipe
tee
fitting
with
CPVC
solvent
cement.
Route
the
cable
away
from
any
high
voltage
conductors.
Orient
sensor
(SD
‐
30
‐
000)
with
outlet
facing
perpendicular
to
flow.
Insert
the
sensor
into
the
Adapter
until
it
seats
in
the
Adapter
and
tighten
the
Retaining
nut
(SD
‐
10
‐
040)
to
seal
the
sensor
in
the
plumbing.