A: Calibration Basics
000-10000-191-02-201601
19
Standard Uncertainties and Uncertainty Budgets in % at Wavelength
For DH-3P (-BAL) CAL Cosine Corrector Calibration
Uncertainty Budgets in [%] at Wavelength
Wavelength (nm)
UV Calibration
Standard VIS Calibration
Extended Calibration
230
250
300
400 500
600
1000
1400
1600
2000
2400
Standard
deviation
k = 1
7.7
6.9
6.5
6.0
4.1
3.9
3.8
4.3
5.6
10.6
11.7
k = 2
15.4
13.8
13.0
12.0
8.2
7.7
7.7
8.5
11.2
21.2
23.5
For DH-3P (-BAL) CAL Bare Fiber Calibration
Uncertainty Budgets in [%] at Wavelength
Wavelength (nm)
UV Calibration
Standard VIS Calibration
Extended Calibration
230
250
300
400
500
600
1000
1400
1600
2000
2400
Standard
deviation
k = 1
N/A
9.5
8.7
8.1
6.8
6.7
6.7
6.9
7.8
11.9
12.9
k = 2
N/A
19.0
17.4 16.1 13.6
13.4
13.3
13.8
15.6
23.8
25.9
Place the light source horizontally. A value of 6 % of expanded uncertainty must be taken into
account for vertical placement.