Halma Alicat Scientific IVCD Скачать руководство пользователя страница 21

2021.05.26 

• 

REV. 1 

• DUAL-VALVE PRESSURE CONTROLLER OPERATING MANUAL

 21

Troubleshooting

 

If you run into trouble with your device’s installation or operation, please get in touch with us by phone, chat, or email 

(

page 2

). You’ll also find help on our website, 

alicat.com/support

, and in the pages that follow.

General Use

 Issue: 

My device does not turn on or has trouble staying on.

 Action: 

Check power and ground connections. Please reference the technical specifications to assure you have the proper 

power for your model.

 Issue: 

The buttons do not work, and the screen shows 

LCK

.

 Action: 

The device buttons were locked out via a serial command. Press and hold all four outer buttons to unlock the interface.

 Issue: 

I can’t read the display easily.

 Action: 

During the day, you can increase the visibility of the display by increasing the contrast or brightness (

page 17

)

For monochrome displays under low-light conditions, push the logo button located below the display to turn on the 

backlight (

page 6

).

 Issue: 

The analog output signal indicates values lower than what appears on my instrument’s display.

 Action: 

Analog signal voltage degrades over long distances. You can minimize this effect by using wires with a heavier gauge, 

especially in the ground wire.

 Issue: 

How often do I need to calibrate my device?

 Action: 

Annual recalibrations are recommended. Check your device’s last calibration date by selecting 

MENU 

 ABOUT 

 

About Device

. If it is time to recalibrate, request a recalibration at 

alicat.com/service

 or get in touch with support 

(

page 2

).

 Issue: 

I dropped my device. Is it OK? Do I need to recalibrate?

 Action: 

If it turns on and appears to respond normally, then it is probably OK. It may or may not need a recalibration. Give it a 

tare, and compare it against a pressure standard you know is accurate. If it checks out, keep using it, but tell us about 

the drop at your next annual recalibration so we can check it out for you.

 Issue: 

How can I see pressure in different units?

 Action: 

From the main menu, select 

SETUP

 

 

Sensor

 

 

Engineering Units

. For more information, see (

page 16

).

Pressure Readings

 Issue: 

My controller won’t reach its setpoint.

 Action: 

You must ensure that there is enough supply pressure to reach the setpoint. If increasing the pressure doesn’t help, 

check to see if there is a clog. Teflon tape can get stuck in the flow channel and block flow. Make sure to clean out 

any loose Teflon tape and never tape the first two threads entering the device to help avoid this issue 

 Issue: 

My pressure readings are negative.

 Action: 

If a negative reading is not expected, your device may need to be tared (

page 11

). Ensure pressure is at zero, and 

tare it.

 Issue: 

My pressure readings jump to zero when pressures are low.

 Action: 

Your device is equipped with a programmable zero band that is preset at the factory. Reduce your deadband threshold 

by selecting 

SETUP → Sensor → Zero Band

 Issue: 

Does the controller work if it is laying down? Will it be accurate?

 Action: 

Yes to both for small valve controllers! The pressure controller is internally compensated for many situations, so you 

can use it sideways, on its back, or upside-down. For upside-down installations, you may want to rotate the display 

180° (

page 17

). However, they should be tared again after changing their orientation. Large-valve controllers 

(PCRD- and PCRD3-Series) should be operated with the valve cylinder vertical and upright.

 Issue: 

Can I attach the controller to a vibrating device? Will it be accurate?

 Action: 

Yes, and yes! The pressure controller is internally compensated for any changes in orientation, including rapid vibrations. 

Noise will increase if the pressure controller is vibrating. Large-valve controllers are not recommended for use on 

vibrating surfaces.

Содержание Alicat Scientific IVCD

Страница 1: ...OPERATING MANUAL FOR DUAL VALVE PRESSURE CONTROLLERS Models PCD PCDS PCD3 IVCD...

Страница 2: ...ler every year Annual calibration is necessary to ensure the continued accuracy of readings and to extend the Limited Lifetime Warranty Fill out the Service Request Form at alicat com service or conta...

Страница 3: ...sure controllers for closed volumes PCDS Series anti corrosive absolute gauge and differential pressure controllers for closed volumes PCD3 Series remote sensing pressure controllers for closed volume...

Страница 4: ...PD I Control Algorithms 13 Tuning the PD PDF Control Algorithm 13 Tuning the PD I Control Algorithm 13 Troubleshooting Valve Performance with PID Tuning 14 Using a Control Deadband for Pressure Contr...

Страница 5: ...ndard dual valve pressure controller Your pressure controller s appearance and connections may differ See page 2 for more examples Backlight The monochrome display comes equipped with a backlight To t...

Страница 6: ...ode active HLD Valve hold active LCK Front display is locked POV Pressure over range of device Mounting Pressure controllers do not require straight runs of pipe upstream or downstream Most pressure c...

Страница 7: ...need Teflon tape applied to the threads Warning It is not recommended to use pipe dopes or sealants on the process connections as these compounds can cause permanent damage to the controller should th...

Страница 8: ...tween two points in the system These devices have two ports located on the front face of the unit for connection to two points in the system where the differential pressure is measured from The upstre...

Страница 9: ...nstruments with Liquids Many pressure devices may be used with chemically compatible liquids but a few warnings must be taken into account Water is about 50 times more viscous than air This is importa...

Страница 10: ...ndard 0 5 Vdc analog output signal Pin 2 is normally a constant 5 12 Vdc More pinout configuration options begin on page 24 Analog Signals Primary Analog Output Signal Most devices include a primary a...

Страница 11: ...to Tare Before every new pressure measurement cycle After significant changes in temperature or pressure After dropping or bumping the pressure controller After installing the controller in a differen...

Страница 12: ...set to Serial Front Panel the controller will accept input from either the front panel or an RS 232 RS 485 connection Neither source is a slave of the other so the controller will accept the most rec...

Страница 13: ...ow to actuate its valve s in order to achieve the commanded setpoint These settings are tuned for your specific operating conditions but changes to your process sometimes require on site adjustments t...

Страница 14: ...nd of 0 25 PSIA allows absolute pressure to vary between 29 75 and 30 25 PSIA The device will increase pressure until it reaches the setpoint of 30 PSIA With a setting of Close valve it will close bot...

Страница 15: ...Only in the serial data frame Screen and Digital Both the main display and serial data frame Device Information The ABOUT menu MENU ABOUT contains useful information for setup configuration and troub...

Страница 16: ...g and logging of all data Before connecting the pressure controller to a computer ensure that it is ready to communicate with your computer by checking the options in this menu For more on how to issu...

Страница 17: ...isplay Screen Lighting The options and wording in the screen lighting menu will vary for monochrome versus color displays On monochrome displays press LESS CONTRAST or MORE CONTRAST to adjust the cont...

Страница 18: ...others can be used if the computer its software and the controller are all set to the same rate Data bits 8 Parity none Stop bits 1 Flow control none Alicat s Serial Terminal Application Alicat s Ser...

Страница 19: ...eads zero Taring an absolute pressure sensor requires a barometer and aligns the absolute pressure sensor with the external barometric pressure For an accurate tare a gauge or absolute pressure sensor...

Страница 20: ...us sign New setpoint unit ID s floating point number setpoint Example as 15 00 setpoint of 15 00 PSIG Quick Command Guide Note Serial commands are not case sensitive In these examples the unit ID of t...

Страница 21: ...sure standard you know is accurate If it checks out keep using it but tell us about the drop at your next annual recalibration so we can check it out for you Issue How can I see pressure in different...

Страница 22: ...Unit ID to make sure you are addressing it properly with your serial commands 3 Check the pinout common pinouts are listed starting on page 24 4 Make sure the COM number matches the one your software...

Страница 23: ...rrD Torr mmHgA mmHgG mmHgD Millimeter of mercury at 0 C inHgA inHgG inHgD Inch of mercury at 0 C mmH OA mmH OG mmH OD Millimeter of water at 4 C NIST conventional mmH OA mmH OG mmH OD Millimeter of wa...

Страница 24: ...ptional 1 5 Vdc or 0 10 Vdc output signal 7 Power In 8 Ground common for power digital communications analog signals and alarms Caution Do not connect power to pins 1 through 6 as permanent damage can...

Страница 25: ...og Out Analog Out Analog Out RX or A Analog Out NC Analog Out 3 Ground Analog In NC NC NC NC NC 4 NC Ground NC NC NC Analog Out NC 5 Power In Ground Power In Ground Ground Power In Ground 6 NC Ground...

Страница 26: ...nalog Setpoint Input Analog Setpoint Input 5 Serial RS 232 TX signal Optional RS 485 B Serial RS 232 TX Signal Optional RS 485 B 6 Static 5 12 Vdc Optional Secondary analog output 4 20 mA 0 5 Vdc 1 5...

Страница 27: ...2021 05 26 REV 1 DUAL VALVE PRESSURE CONTROLLER OPERATING MANUAL 27...

Страница 28: ...d Plot No A 147 Road No 24 Next to Spraytech Circle opp Metropolitan Company Wagle Industrial Estate Thane West Mah r shtra 400 604 91 022 41248010 China SE Asia info cn alicat com alicat com cn 2nd F...

Отзывы: