Hallowell EMC Tafonius Junior Скачать руководство пользователя страница 15

Tafonius 

 

Junior

 – A User Guide 

Pressure Gauges

Junior w/gas also has 2 pressure guages on the front panel.  One for Breathing System 
Pressure the range of which is -10 – +60 cmH

2

O and one for Oxygen Supply pressure with a

full scale of 100psi (700kPa).

Junior

 and Spontaneous Breathing

Thinking of the Piston as a “Virtual bag

When the ventilator is idle and a patient is connected, the system behaves like a ‘perfect’ 
bag. By a perfect bag we mean that when the patient breathes out there is no expiratory 
resistance and when the patient breathes in there is no inspiratory resistance.  This action is
possible because the airway pressure is sensed at the centre of the Y-piece.  As soon as a 
pressure deviation in excess of 0.5 cmH

2

O from the ambient pressure is detected the piston 

is moved so as to negate this pressure change. If the patient breathes out then the piston 
withdraws, effectively filling.  If the patient inspires then the piston moves down, effectively 
emptying.  Because a motor drives the piston it is the motor that overcomes the resistances 
of any piping, valving or soda lime.  To the patient it feels like breathing to room air with no 
resistance.

Dictating the size of the “Virtual Bag”

The size bag you use on your anaesthesia machine depends on the size of the patient.  The
breathing bag must be large enough to hold the largest size breath the patient may take.  
But it should not be overly large because the excess volume increases the time constant of 
the breathing system making it take longer for changes in vaporizer settings or the FiO

2

 

setting equilibrate and take effect.

The tidal volume and the concept of a buffer volume are used to control the size of the 
“Virtual Bag”, the volume under the piston, in 

Junior

 .  The piston is not allowed to rise 

above the combined volumes of TV + BV.  Whenever the piston volume reaches this upper 
limit, gas is allowed to flow out to the scavenging system.   

In this manor 

Junior

 is changing the Virtual bag size every time you change the TV.

Add-a-Litre feature

In addition to continually managing the maximum volume is the system 

Junior 

also works 

to keep a the minimum volume in the system at all times.   

Junior 

considers the piston to 

be in the low volume zone whenever it dips below the 1.5 litre point.  When this happens 
gas flows into the system during current and subsequent the inspiratory phase(s) until a liter 
has been added and the 

LOW SYSTEM RESERVE VOLUME

 alarm will sound.

Example 1:

A 300kg patient breathing spontaneously.  Rule of thumb would suggest setting the TV 
control to 3.0L in preparation for IPPV.  The BV is set to its default of 5.0L.  The upper limit 
of the piston volume is then  3.0 + 5.0 L = 8.0L, the patient is effectively breathing from an 
8L bag.  

Say the patient has just been induced, initial gas flow rates may be of the order of 5-10L per 
minute. This will tend to fill the system rather quickly. 

When the piston volume reaches the 8.0L limit the piston stops rising and the excess gas 
will be discharged from the system through the electronically controlled Dump Valve.  You 
will hear the dump valve opening and closing as a clunk... clunk.  After induction you would 
turn the fresh-gas-flow down but lets say you turn it down a little to far.  The piston volume 
falls a little breath by breath, the clunking sound will stop, until finally the piston dips below 
the 1.5L level towards the end of a cycle.  At this point an extra 1.0L of gas will be added to 
Printed: 26/Oct/18

Page 15 of 20

DOCA4962B

TOC

Содержание Tafonius Junior

Страница 1: ...Tafonius JunioR J2 A USERS GUIDE v01Jun2017 Printed 26 Oct 18 DOCA4962B...

Страница 2: ...Tafonius Junior A User Guide Notes Printed 26 Oct 18 Page 2 of 20 DOCA4962B TOC...

Страница 3: ...EP 13 Dump Valve Button 13 Buffer Volume Setting 14 Controls on the Anesthesia Machine Side 14 Oxygen Flowmeter 14 Air Flowmeter 14 N2O Flowmeter 14 Oxygen Flush Button 14 Scavenging Flowmeter 14 Pres...

Страница 4: ...SSURE LOW 19 OXYGEN SUPPLY PRESSURE LOW 19 LOW BREATHING SYSTEM PRESSURE 19 BATTERY 1 CRITICALLY LOW and or 19 BATTERY 2 CRITICALLY LOW 19 NO RESPONSE TO INPUT KNOBS 19 RD VACUUM FAILURE 19 Technical...

Страница 5: ...nly Junior sans gas for use with an existing gas machine Junior sans gas Shown here with a Matrx VML by Midmark gas machine head Printed 26 Oct 18 Page 5 of 20 DOCA4962B TOC 6 152mm Locking Front Cast...

Страница 6: ...thing Circuit connection port qVGA Display Soft Power Switch Future RM button Optional two position Select a TEC backbar Scavenging flowmeter RCD GFCI mains breaker Mains fuses Power cord Gas hose han...

Страница 7: ...nd air Only features additional to those in the ventilator only version above are labeled below Printed 26 Oct 18 Page 7 of 20 DOCA4962B TOC Vaporizer mounting locations CO2 Absorber pan Patient Wye 2...

Страница 8: ...ersion 6021 with Boot Loader ver 106 followed buy the Runtime in the format hours on hours ventilating Next the battery voltages are checked The result of this and each subsequent test is displayed on...

Страница 9: ...ed Follow the instructions on the screen Fit a stopper to the patient wye piece turn on the gas flow to bring the breathing system pressure up to 20 cmH2O Adjust the flow rate to hold the pressure ste...

Страница 10: ...d in this area as seen above The display of warning and error messages will replace the machine name when they occur 2 Piston Graphic and Settings Information Area 2 shows a representation of the pist...

Страница 11: ...w other parameters are affected by the change being made One can therefore set a dependent variable For example you can set the MV by changing the TV and or the RR and watch the resulting affect on th...

Страница 12: ...nes maximum capacity of 900 lpm inspiratory flow When a limit is reached the setting will not advance as the knob is turned THE CONTROL KNOBS Tidal Volume The range is 0 1L to 20 0L in 0 1L increments...

Страница 13: ...el the ASSIST function This is a new feature and we currently have no guidelines on sensitivity values to set The software offers a range of between 1 LPM and 200 LPM as trigger values We would welcom...

Страница 14: ...uld on any other anaesthesia machine Air Flowmeter An optional air flowmeter with either a yellow knob for the US or a black and white knob for most other countries may be installed on your machine Th...

Страница 15: ...volume are used to control the size of the Virtual Bag the volume under the piston in Junior The piston is not allowed to rise above the combined volumes of TV BV Whenever the piston volume reaches th...

Страница 16: ...pendant parameters such as minute volume inspiratory flow etc are not affected by these measurements and reflect only their calculated values based on the settings for TV RR and Itime Junior and IPPV...

Страница 17: ...t or ceiling drop via a properly color coded hose to the DISS or NIST Vacuum or EVAC connector below the scavenging flowmeter at the right rear side of the machine In addition a 5 litre reservoir bag...

Страница 18: ...E VOLUME The piston has entered the low volume zone at the 1 5L point If this is during an active inspiratory phase then a litre of gas will automatically be added to the breathing system through the...

Страница 19: ...when the patient takes a breath at the same time as the ventilator delivers a breath The Alarm will continue until either a pressure in excess of 7cm H2O is registered during an inspiratory phase OR t...

Страница 20: ...racy better that 25ml Buffer Volume BV 2 0 L to 20 L Respiratory Rate 1 0 to 30 0 bpm RR settings resolution 1 0 to 2 0 bpm 0 2 bpm 2 0 to 5 0 bpm 0 5 bpm 5 0 to 30 0 bpm 1 bpm Inspiratory Time 0 5 to...

Страница 21: ...01A4313 120 3048mm BREATHING BAG 200A4980 VACUUM RESERVOIR CO2 ABSORBENT 202A2906R 60 TUBE SIDE STREAM GAS SAMPLING TUBE ASSY 200A4453 C VACUUM PUMP AMBIENT 2X 180A3241 VALVE BLOCK O RING D 1 8 27 FNP...

Страница 22: ...This Page Left Intentionally Blank Section End...

Страница 23: ...d turn it OFF 2 Remove the expandable drain clamp from either tube under the bottom pan 3 Fit the manual breathing tube over the pipe 4 Attach the bag to the connector on the side of the machine The O...

Страница 24: ...ing by undoing the 4 Velcro retainers Remove the Y circuit from the ventilator Flush clean and hang to dry Undo and remove the 6 thumbscrews holding the valve block cover Remove and clean the valve ho...

Страница 25: ...A3239 ABSORBER PAN 000A4621 BOTTOM PAN DIVIDER SEAL 180A4639 LINEAR ACTUATOR PISTON COVER UPPER CYLINDER HALF PISTON LOWER CYLINDER HALF BOTTOM PAN INSPIRATORY FLOW EXPIRATORY FLOW VACUUM HOLDS SILICO...

Страница 26: ......

Страница 27: ...damaged sampling pressure lines seals and gaskets rolling diaphragms etc The service strip down and re assembly will include but is not limited to Visual assessment Check the battery integrity power c...

Страница 28: ...stomer s responsibility to schedule service dates within the Extended Warranty or Annual Service timeframe Manufacturers Warranty Please see our Tafonius Limited Warranty Statement for Full Details Ha...

Страница 29: ...y with required Annual Service Call for an appointment Annual Service 4 500 No Charge 20 Discount TBD 12 months on entire machine4 Eligibility for Extended Warranty for 12 months 1 Time Travel time bi...

Страница 30: ...wivel valve block permits the breathing tubes to exit to the right or the left n Stainless hoop handle and bumpers n GFI protected with 3 auxiliary AC power outlets n Panel mount e cylinder pressure g...

Страница 31: ...ius junior Sans gas For use with your current gas machine Pictured with Matrx VML by Midmark Small Animal Machines 1500 3000 ml TV 300 1600 ml TV 20 350 ml TV 5 200 ml TV Model 2002 PRO One Machine th...

Отзывы: