20
ELECTROMAGNETIC IMMUNITY GUIDANCE
ELECTROMAGNETIC IMMUNITY GUIDANCE
Guidance and manufacturer’s declaration-electromagnetic immunity
The
HairMax Laser Device is intended for use in the electromagnetic environment specified below. The customer or the user of the HairMax Laser Device
should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV for power supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5
± 1kV differential mode
± 2kV common mode
± 1kV differential mode
± 2kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
0% UT; 0,5 cycle g) At 0˚, 45˚, 90˚,
135˚,180˚, 225˚, 270˚, and 315˚
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0˚
0% UT; 250/300 cycle
0% UT; 0,5 cycle g) At 0˚, 45˚, 90˚, 135˚,180˚,
225˚, 270˚, and 315˚
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0˚
0% UT; 250/300 cycle
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the HairMax Laser Device requires continued
operation during power mains interruptions, it is
recommended that the HairMax Laser Device be
powered from an interruptible power supply or a
battery.
Power frequency
(50, 60 Hz) magnetic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.