Alarm
Alarm type
Phenomenon
Alarm Indication
Buzzer alarm
High
temperature
alarm
temperature alarm
Internal temperature > set
tempe0.5 °C or the
set limit
Display of parameter in
red
Pulse sound alarm
Low
temperature
alarm
Internal temperature < set
temperature -0.5 °C or the set
limit
Display of parameter in
red
Pulse sound alarm
Over
temperature
alarm
Internal temperature > over
temperature protection limit
Display of parameter in
red
Pulse sound alarm
High
concentration
alarm
Internal CO
2
concentration >
set concent0.5% or
the set limit
Display of parameter in
red
Pulse sound alarm
Low
concentration
alarm
Internal CO
2
concentration<
set concentration -0.5% or the
set limit
Display of parameter in
red
Pulse sound alarm
Door opening
alarm
The door is opened for more
than 30s
Alarm fl ashes during
the set alarm time
Giving pulse sound
alarm within the set
alarm time
Sensor fault
Fault of the ambient
temperature sensor
Flicker alarm
Pulse sound alarm
Fault of the temperature
sensor
Flicker alarm
Pulse sound alarm
Fault of the CO
2
concentration
sensor
Flicker alarm
Pulse sound alarm
This series of incubators has automatic alarm restore function:
•In the case of an alarm, press the mute button
on the main interface to stop the buzzer alarm;
•If the alarm condition still exists, the buzzer alarm will automatically restore after the buzzer alarm
delay time is over.
Alarm restore