10-4
Shroud and Base Assemblies
P/N 108755-US, Manual revision: AA
M
OTOR
R
EPLACEMENT
Note: Motor Assemblies P/N 10402300 and P/N 10402405-0001 are no longer
in use. When replacing, use Motor Assembly P/N 10401300-00 (Motor P/N
10401301-00 and 10402407). See Caution below.
Caution:
The new motor is incompatible with the old power supply PCBAs, P/N
10403200. Power supply P/N 10403600 must be in place when replacing an old
motor. See Chapter 14, Power Supply Replacement on page 14-2.
1.
Motor failure is indicated when the shroud will not rock, yet the LED illu-
minates and comes to the "READY" mode.
Figure 10-1, Location of Motor Screws (Fourth Screw Obscured by Link Y)
2.
To replace the motor subassembly, remove base plate (Chapter 4, Base
Plate Removal on page 4-2) and shroud (Shroud Removal on page 4-3).
Remove Body Board and Power Supply Assemblies (Chapter 5, Strain
Gauge Inspection and Replacement on page 5-2 and Body Board Strain
Gauge Amp Adjustment Procedure on page 5-7). Remove Link Assembly
(Chapter 6, Link Assembly Removal on page 6-3) Remove Crank M from
the motor shaft by removing the hex bolt screw. Remove the four screws
that secure the motor assembly to the base assembly.
3.
Reverse disassembly procedures to install new motor assembly.
Link Y
Motor Screw (x4)
Crank M
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...