14-2
Power Supply
P/N 108755-US, Manual revision: AA
P
OWER
S
UPPLY
R
EPLACEMENT
Caution:
Always wear an ESD strap when working on the Shaker.
Warning: The Input Power fuse "F1" fuse holder, located on the power supply
PCBA, is exposed when the Body Board and Power Supply PCBA's are removed
from the base. Use CAUTION when handling this assembly.
Note: Schematics are needed for this operation.
Figure 14-1, Location of Fuse Holder
Power supply
inspection
1.
Remove base plate (see Chapter 4, Base Plate Removal on page 4-2).
2.
Check the P1 connector for proper voltages. If no voltage is present, check
F1. If F1 is good, replace the power supply.
Power supply
replacement
Note: Power supply P/N 10403200-xx and P/N 10402405-XXX are no longer in
use. Contact Haemonetics to determine the appropriate replacement.
Power Supply
Assembly
(Intact in Shaker)
F1
Fuse Holder
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...