2-18
Describing the PCS2 Device Components
P/N 85266-30, Manual revision: A
Procedure acronym
The upper left portion of the display screen will contain the acronym selected to
describe the type of procedure in progress.
Display screen icons
These symbols, located on the upper left side of the display screen, provide a
pictorial representation of the operating state, or mode, in progress.
DPM bar graph
This is a visual representation of the pressure reading in the donor-line tubing, as
it measured by the donor pressure monitor. The DPM bar graph (when visible) is
located on the right side of the screen. The contrast between filled and non-filled
area in the column will vary to depict the fluctuations in the DPM readings.
Table 2-1: PCS2 display screen icons
Display
screen icon
Explanation
Operating
state/mode
Displayed during AC solution priming sequence.
PRIME
Displayed when the device in a non-active state,
or ready for an operator command.
READY
Displayed as donor blood is being drawn into the
centrifuge bowl.
DRAW
Displayed as blood components/fluids are being
returned to the donor.
RETURN
Displayed briefly at the end of the procedure,
prior to the completed procedure statistics
displayed.
PROCEDURE
COMPLETE
Displayed when the centrifuge is stopping.
AC