background image

 

 

16 

Application 

 

For  security  reasons,  only 
change  the  tool  if  the 
handpiece is turned off. 

 

Adjusting the speed 

 

Put  the  handpiece  back  in  the  hand-
piece  holder  or  keep  it  in  your  hand. 
Switch ON the unit with switch (7). Now 
you  can  adjust  the  requested  speed 
with  the  switches  (8)  &  (9).  Choose  the 
direction with switch (12).

 

 

Spray 

Turn  on  the  spray  function  by  pushing 
switch  (10).  Now  the  spray  system  will 
be  filled  automatically  and  a  strong 
spray  stream  will  be  blowed  out  of  the 
handpiece for the first 15 sec. Then you 
can  adjust  the  requested  spray  stream 
while turning knob (3).

 

 

When  turning  off  the  unit  or  the  spray 
function,  the  spray  liquid  is  pumped 

back  completely  into  the  tank  from  the 
hoses.  With  this  process,  a  bubbling 
can be heard in the tank

 

Drying/ Changing Tool 

Press  key  (11)  and  the  handpiece  and 
spray will be turned off. Now only air will 
come  out  of  the  top  of  the  handpiece, 
which you can use to dry the spot of the 
recent treatment.  
When  you  push  button  (11)  again,  the 
spray and handpiece are working again. 

 

Hint:  This  function  can  also  be  used 

to  change  your  tools  while  the 
handpiece is not running. 

 

Foot switch (Art. 0970) 

 
Commencement of foot switch  

Only  use  the  original  hadewe  foot 
switch  to  avoid  damages.  Insert  the 
plug  of  the  foot  switch  into  the  socket 

with this symbol: 

 

 

Function of the foot switch 

While the spray is on, and you push the 
switch once, the spray stream will stop. 
If  you  push  it  a  second  time,  the  hand-
piece  will  stop.  Pushing  the  switch  a 
third  time,  both  functions  will  start  to 
work again. 

 
When  working  with  a  foot  switch, 
take  care  that  you  only  use  instru-
ments  that  are  approved  for  use  at 
the  maximum  speed  of  the  device

.

Содержание Argos 2

Страница 1: ...Art No 0572 FB 04 278 Rev 0 Date 10 05 2016 Manual Argos 0572...

Страница 2: ...tion 12 Operation 12 Work positioning 13 Infilling the spray liquid 15 Mounting Changing the filter 15 Insert the bur 15 Adjusting the speed 16 Spray 16 Drying changing tool 16 Footswitch 16 Maintance...

Страница 3: ...duct to someone else please include this document as it includes important information on how to correctly use the product Short description of functions The Argos is a device for the treatment of cal...

Страница 4: ...lways be easily accessible Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand Prevent the permeation of liquids into the control unit...

Страница 5: ...should be exchanged at least after 50 operating hours Never use the suction to vacuum any liquids or air water fog which contains particles of liquids If water leaks from the unit then a disturbance...

Страница 6: ...ent injury work with extreme caution on the foot of the patient and use only light pressure To ensure that the burr tool sits firmly in the chuck do not work with tools which have an oily shaft are wo...

Страница 7: ...orkshops and trained technicians upon re quest Never remove the housing while the equipment is plugged in as some components are live A spare parts list may be sup plied to qualified persons repair wo...

Страница 8: ...tion please adhere to the cleaning and disinfection instructions provided be low After every treatment please carry out these steps Remove the rotating tool e g bur first To do so pay attention to the...

Страница 9: ...the water tank 6 Water tank 7 ON OFF button line voltage is not disconnected unit in standby 8 handpiece speed down 9 handpiece speed up 10 Switch for modus change spray or suction 11 Switch for turni...

Страница 10: ...carry case It is suitable for the following scopes Foot care Milling trimming and polishing Attention The suction or spray must be in operation dur ing foot care Manicure Grinding milling and polishin...

Страница 11: ...from the patient s foot The unit should be 30 90cm from the floor Image Example of positioning of foot care unit in treatment room Image Bird s eye view of work area If you have a foot switch insert...

Страница 12: ...t for diamond cutters and high grade corundum sanders in terms of effi ciency Working anti clockwise can be beneficial if for example the dust obscures your field of vision when working clockwise Now...

Страница 13: ...Application Work positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without str...

Страница 14: ...14 Application...

Страница 15: ...item no 3479 should be fully under pres sure otherwise no water will be sucked in The water filter has to be fully under water otherwise no water will be sucked in Mounting Changing the filter Remove...

Страница 16: ...e heard in the tank Drying Changing Tool Press key 11 and the handpiece and spray will be turned off Now only air will come out of the top of the handpiece which you can use to dry the spot of the rec...

Страница 17: ...find some remove such eventually Please pay attention to the way it was installed Push the new washer in the top Water Filter Art no 3497 Even smallest parts of dirt could block the nozzle which sits...

Страница 18: ...rts of dirt are rinsed into the tank Do not use this water again but clean the tank immediately after realiz ing a blockage in the nozzle Otherwise the dirt parts might stick together Clean the water...

Страница 19: ...a small drop of oil on your thumb and pointer Rub both fingers until you find very little of oil on your fingers Now take the chuck and put it between these two fingers and turn it around There shoul...

Страница 20: ...al cleaning and disinfection liquids This extraneous rost affects internal parts of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece al ways rem...

Страница 21: ...in the manual Rubber ring under the knob twists while using it Twist it of 180 It does not spray Pressure cannot be set up in the tank because the lid is not tightened well There are parts of dirt in...

Страница 22: ...ters There is air in the spray system Af ter 5 min running this effect will dis appear Acceleration of the spray stream Note To reduce the starting time of the spray turn the spray regulator on Max Af...

Страница 23: ...for extended usage Ambient conditions The device is intended for use in dry closed rooms Operation Temp 10 C to 26 C Rel humidity 25 to 75 not condensed Air pressure 700 hPa to 1060 hPa Storage Temp 5...

Страница 24: ...ent Direct current Safety fuse Type B applied part Socket foot switch Date of manufacture Disposal The device and filter may contain infectious material such as residual sanding dust or skin particles...

Страница 25: ...tion spray clamping system Inspection before use During the production control the following inspections as per DIN EN 62353 VDE 0751 1 are conducted among others Optical control unit discharge cur re...

Страница 26: ......

Страница 27: ......

Страница 28: ...hadewe GmbH Grambartstra e 10 30165 Hannover Germany info hadewe de www hadewe de...

Отзывы:

Похожие инструкции для Argos 2