Handling Instructions 

Always  deposit  the  handpiece 
with the top downwards. There-
fore  you  can  avoid  liquid  from 

entering the handpiece. 

 

Never  put  the  handpiece  into 
any kind of liquid, and never oil 
or grease the handpiece. Clean 

its surface with a damped cloth. Pay 
attention,  that  no  moisture  or  liquid 
enters  the  handpiece.  Never  use 
spray dis-infection. 

 

Never  put  any  wet  burs  or  in-
struments  into  the  handpiece. 
Its  moisture  could  damage  the 

interior parts of the handpiece. 

 

Any  materials  used  by  hadewe 
prevend  a  formation  of  rust.  In 
cases in which a rust infestation 

was  found  within  the  handpiece, 
thorough  inspections  and  examina-
tions  always  have  proved  that  the 
rust had a foreign origin (which was 
usually  caused  by  the  tools  used  in 
these circumstances). 

 

Even  in  high-grade  steel  in-
struments  of  well  known  manu-
facturers,  formation  of  rust  can 

appear,  e.g.,  through  wrong  treat-
ment  with  chemical  cleaning  and 
disinfection  liquids.  This  extraneous 
rost  affects  internal  parts  of  the 
handpiece in form of rust bloom and 
causes  pitting.  Therefore,  when  you 
not  working  with  the  handpiece,  al-
ways remove the bur/ tool from it. 

 

Always  make  sure  that  you  do 
not let the handpiece fall down 
or  exposes  it  to  any  shocks. 

The  ballbearings  and  the  motor 
could be damaged by this.  

 
 

Only  work  with  tools,  which  are 
flawless. 
 

Do not work with tools which 

are bent or out of balance 

whose shaft is worn out or 

are rusty. 

 
 

Working  with  faulty  tools  will 
cause  a  stronger  vibration  of 
the handpiece. A damage of the 

ball bearings, clamping technique or 
motor could be the consequence. 

 
Working  with  high  pressure  will 
not  improve  the  outcome  of 
your  work,  since  the  motor  will 

even  be  slowed  down.  A  second 
disadvantage  is  the  stronger  impact 
on  the  ball  bearings  and  the  motor, 
which  will  shorten  their  life  span  in 
the long run. 

 

Always  working  with  maximum 
speed  might  have  a  similar 
effect.  Please  make  sure  that 

the speed setting you use is accord-
ing  to  the  work  you  are  doing  and 
the  instruments  you  are  working 
with. 

Содержание Argos 2

Страница 1: ...Art No 0572 FB 04 278 Rev 0 Date 10 05 2016 Manual Argos 0572...

Страница 2: ...tion 12 Operation 12 Work positioning 13 Infilling the spray liquid 15 Mounting Changing the filter 15 Insert the bur 15 Adjusting the speed 16 Spray 16 Drying changing tool 16 Footswitch 16 Maintance...

Страница 3: ...duct to someone else please include this document as it includes important information on how to correctly use the product Short description of functions The Argos is a device for the treatment of cal...

Страница 4: ...lways be easily accessible Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand Prevent the permeation of liquids into the control unit...

Страница 5: ...should be exchanged at least after 50 operating hours Never use the suction to vacuum any liquids or air water fog which contains particles of liquids If water leaks from the unit then a disturbance...

Страница 6: ...ent injury work with extreme caution on the foot of the patient and use only light pressure To ensure that the burr tool sits firmly in the chuck do not work with tools which have an oily shaft are wo...

Страница 7: ...orkshops and trained technicians upon re quest Never remove the housing while the equipment is plugged in as some components are live A spare parts list may be sup plied to qualified persons repair wo...

Страница 8: ...tion please adhere to the cleaning and disinfection instructions provided be low After every treatment please carry out these steps Remove the rotating tool e g bur first To do so pay attention to the...

Страница 9: ...the water tank 6 Water tank 7 ON OFF button line voltage is not disconnected unit in standby 8 handpiece speed down 9 handpiece speed up 10 Switch for modus change spray or suction 11 Switch for turni...

Страница 10: ...carry case It is suitable for the following scopes Foot care Milling trimming and polishing Attention The suction or spray must be in operation dur ing foot care Manicure Grinding milling and polishin...

Страница 11: ...from the patient s foot The unit should be 30 90cm from the floor Image Example of positioning of foot care unit in treatment room Image Bird s eye view of work area If you have a foot switch insert...

Страница 12: ...t for diamond cutters and high grade corundum sanders in terms of effi ciency Working anti clockwise can be beneficial if for example the dust obscures your field of vision when working clockwise Now...

Страница 13: ...Application Work positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without str...

Страница 14: ...14 Application...

Страница 15: ...item no 3479 should be fully under pres sure otherwise no water will be sucked in The water filter has to be fully under water otherwise no water will be sucked in Mounting Changing the filter Remove...

Страница 16: ...e heard in the tank Drying Changing Tool Press key 11 and the handpiece and spray will be turned off Now only air will come out of the top of the handpiece which you can use to dry the spot of the rec...

Страница 17: ...find some remove such eventually Please pay attention to the way it was installed Push the new washer in the top Water Filter Art no 3497 Even smallest parts of dirt could block the nozzle which sits...

Страница 18: ...rts of dirt are rinsed into the tank Do not use this water again but clean the tank immediately after realiz ing a blockage in the nozzle Otherwise the dirt parts might stick together Clean the water...

Страница 19: ...a small drop of oil on your thumb and pointer Rub both fingers until you find very little of oil on your fingers Now take the chuck and put it between these two fingers and turn it around There shoul...

Страница 20: ...al cleaning and disinfection liquids This extraneous rost affects internal parts of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece al ways rem...

Страница 21: ...in the manual Rubber ring under the knob twists while using it Twist it of 180 It does not spray Pressure cannot be set up in the tank because the lid is not tightened well There are parts of dirt in...

Страница 22: ...ters There is air in the spray system Af ter 5 min running this effect will dis appear Acceleration of the spray stream Note To reduce the starting time of the spray turn the spray regulator on Max Af...

Страница 23: ...for extended usage Ambient conditions The device is intended for use in dry closed rooms Operation Temp 10 C to 26 C Rel humidity 25 to 75 not condensed Air pressure 700 hPa to 1060 hPa Storage Temp 5...

Страница 24: ...ent Direct current Safety fuse Type B applied part Socket foot switch Date of manufacture Disposal The device and filter may contain infectious material such as residual sanding dust or skin particles...

Страница 25: ...tion spray clamping system Inspection before use During the production control the following inspections as per DIN EN 62353 VDE 0751 1 are conducted among others Optical control unit discharge cur re...

Страница 26: ......

Страница 27: ......

Страница 28: ...hadewe GmbH Grambartstra e 10 30165 Hannover Germany info hadewe de www hadewe de...

Отзывы: