10
Delete all data:
Click
Delete all data
to delete all waveform data on the screen.
Dump all cuffs:
Click
Dump all cuffs
to dump all cuff.
(4)
Sound adjustment
Click Sound button and select
R
or
L
side from where PV waveform sounds
come out from the speaker and move the slider to adjust the sound volume.
Check the Mute check box to mute the sound volume.
Note: No sounds during oscillometry testing.
(5)
Removing waveform window(s)
To remove waveform window(s) you do not use, click each trash can next to
the window, clicked can becomes fat , and then click
Remove
button to
proceed.
Click
&
buttons to get the removed window(s) back.
4-2-2. Starting Automatic Arterial Testing
(1) Wrap cuffs snugly
on each side of brachium, ankle, and great toe.
Note: Use appropriately sized cuffs for each patient.
(2) Hook up each color coded tubing between cuff and the inflation port with the
same color as tubing
’s for each site.
(3) Connect 2 PPG probes to the unit and affix
each PPG transducer with clear side
against the pad of each great toe with PPG
probe clip.
Note: PPG probe is used for taking great toe pressures.
(4) Click
Auto Start
button and auto-testing will start on both sides simultaneously
in the order selected in Auto-testing order.
Select
R
or
L
.
Slider
Trash can
Remove button
Clip PPG probe
Toe cuff