18
Operation
4.4.2
SENSOR setup
SELECT SENSOR (if there is more than one sensor)
WIPE
Triggers a wiping operation
CALIBRATE (turbidity)
SET OUTMODE
Behavior of the outputs during calibration and zero point adjustment
HOLD
ACTIVE
SET TRANSFER
SELECTION
SENSOR MEASURE
Current, uncorrected measurement value
FACTOR
Can be set from 0,10 to 10,00; a detailed description is provided in section
4.5
CALIBRATE
OFFSET
Can be set from -100 to +100, a detailed description is provided in section
4.5
CALIBRATE
CALIBRATE
MEMORY
POINT 1
Calibration point 1 is recorded
POINT 2
Calibration point 2 is recorded
POINT 3
Calibration point 3 is recorded
CLEAR MEMORY
Clears the recorded values for all points.
POINT 1
Current calibration for point 1
POINT 2
Current calibration for point 2
POINT 3
Current calibration for point 3
SET CAL DEFLT
Security prompt, reset to default calibration
CALIBRATE (TS content)
SET OUTMODE
Behavior of the outputs during calibration and zero point adjustment
HOLD
ACTIVE
SET TRANSFER
SELECTION
SENSOR MEASURE
Current, uncorrected measurement value
FACTOR
Can be set from 0,10 to 10,00; detailed description in section
4.5 CALIBRATE
CALIBRATE
MEMORY
POINT 1
Calibration point 1 is recorded
POINT 2
Calibration point 2 is recorded
POINT 3
Calibration point 3 is recorded
CLEAR MEMORY
Clears the recorded value for all points.
POINT 1
Current calibration for point 1
POINT 2
Current calibration for point 2
POINT 3
Current calibration for point 3
SET CAL DEFLT
Security prompt, all calibration points are cleared