Section 4

Operation

4.1

User interface and navigation

The sensor can be operated with all sc controllers. Refer to the controller documentation 
for keypad description and navigation information.

4.2

Sensor setup

When the sensor is connected for the first time, the sensor serial number is displayed as 
the name of the sensor. To change the sensor name:

1.

Open the·MAIN MENU.

2.

Select SENSOR SETUP and confirm.

3.

Select the corresponding sensor and confirm.

4.

Select CONFIGURE and confirm.

5.

Select EDIT and confirm.

6.

Edit the name and confirm to return to the CONFIGURE menu.

Complete sensor configuration in the same manner, with the following menu options 
selected:

MEAS UNITS

PARAMETERS

CLEAN. INTERVAL

RESPONSE TIME

LOGGER INTERVAL

7.

Go back to the MAIN MENU or the measurement mode display. 

4.3

Sensor data logger

There is a data log and event log available for each sensor. The data memory is used to 
store measurement data at preset intervals; the event memory stores events such as 
configuration changes, alarms and warning conditions. Both logs can be exported to CSV 
format (refer to the controller manual).

4.4

Menu structure

Note: 

Refer to 

Section 6 Troubleshooting

 for a list of all possible error and warning messages 

together with a description of all necessary countermeasures to be taken.

4.4.1

SENSOR STATUS

SELECT SENSOR (if there is more than one sensor)

ERROR LIST

Possible error messages: MEAS OVERRANGE, CAL. INSUFF. +/–, ZERO, CAL REQUIRED, EE 
RSRVD ERR, ERROR PROBE, LED FAILURE

WARNING LIST

Possible warning messages: REPLACE PROFILE, TEST/MAINT, GASKET

Содержание TSS HT sc

Страница 1: ...DOC023 52 90154 TSS sc TSS W sc TSS HT sc TSS VARI sc TSS XL sc TSS TITANIUM2 sc TSS TITANIUM7 sc User Manual 09 2018 Edition 3 ...

Страница 2: ......

Страница 3: ...es 11 2 4 Handling 11 2 5 Scope of delivery 11 2 6 Function test 11 Section 3 Installation 13 3 1 Installation overview of basin sensor 13 3 2 Pipe installation options for sensor installation 14 3 3 Connect the sensor cable 15 Section 4 Operation 17 4 1 User interface and navigation 17 4 2 Sensor setup 17 4 3 Sensor data logger 17 4 4 Menu structure 17 4 4 1 SENSOR STATUS 17 4 4 2 SENSOR setup 18...

Страница 4: ...Clean the measurement windows 25 5 4 Replace the wiper profile 26 Section 6 Troubleshooting 27 6 1 Error messages 27 6 2 Warnings 27 Section 7 Replacement parts and accessories 29 7 1 Replacement parts 29 7 2 Accessories 29 Section 8 Warranty and liability 31 Appendix A Modbus register 33 ...

Страница 5: ...able Calibration Turbidity TRB Calibrated before shipping Solid matter TS To be calibrated on site by the customer Zero point Calibrated permanently before shipping Environmental conditions Pressure range TSS sc 10 bar or 100 m 145 PSI TSS W sc 6 bar or 60 m 87 PSI TSS HT sc 10 bar or 100 m 145 PSI TSS VARI sc 16 bar or 160 m 232 PSI TSS XL sc 16 bar or 160 m 232 PSI TSS TITANIUM2 sc 10 bar or 100...

Страница 6: ...WG 22 12 V DC twisted cable pair 1 AWG 24 data twisted cable pair shared cable screen TSS HT sc TSS TITANIUM sc Sensor connection cable permanently connected Teflon PTFE 1 AWG 22 12 V DC twisted cable pair 1 AWG 22 data twisted cable pair shared cable shield Cable gland TSS sc TSS W sc TSS HT sc TSS VARI sc TSS XL sc Stainless steel 1 4305 TSS TITANIUM2 sc Grade 2 titanium TSS TITANIUM7 sc Grade 7...

Страница 7: ...7 Technical data 1 1 Dimensions Figure 1 Dimensions 1 Basin sensor 3 TSS XL sc 2 Installation sensor TriClamp 4 TSS VARI sc ...

Страница 8: ...8 Technical data ...

Страница 9: ... such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction DANGER Explosion hazard This product is not suitable for use in hazardous areas DANGER Indicates a potentially or imminently hazardous situation that if not avoided can r...

Страница 10: ...de from stainless steel for color independent measurement of highly concentrated sludges This sensor connects to VARIVENT piping systems without wiper refer to 6 in Figure 2 Overview 2 2 4 TSS XL sc 0 001 to 9999 FNU 0 001 to 500 g L Extremely accurate turbidity and solid sensors made from stainless steel for color independent measurement of highly concentrated sludges This sensor connects to TriC...

Страница 11: ...nts can be defined for a highly fluctuating medium The combined multiple beam alternating light technique records solids in the medium with even greater accuracy 2 4 Handling Do not subject the sensor to any hard mechanical impacts 2 5 Scope of delivery TSS sc sensor Protective cap for sensor tip depending on model Test log User manual TSS sc wiper kit for 5 changes including screws and screwdrive...

Страница 12: ...12 General information ...

Страница 13: ...asin sensor Note This system must be installed by qualified personnel Figure 3 Installation example with optional accessories DANGER Explosion hazard The TSS sc sensors are not suitable for use in hazardous locations CAUTION Personal injury hazard The installation of this system must only be carried out by qualified experts in accordance with all local safety regulations ...

Страница 14: ...bar LZU300 99 000001 1 x identifier for the nominal diameter of the pipe 2 y identifier for the material selection of the associated connecting flange 4 TSS XL sc with XL LZU304 99 100x01 measuring tube 2 TSS sc Inline TSS W sc Inline TSS HT sc Inline with LZY630 00 1y0002 safety installation fitting max operating pressure 6 bar 5 TSS VARI sc with VARIVENT LZU304 99 000x01 measuring tube 3 TSS sc ...

Страница 15: ...lug and push the plug into the socket 3 Tighten the nut by hand Note Extension cables are available in various lengths refer to Section 7 Replacement parts and accessories Maximum cable length 100 m 328 ft Figure 5 Connect the sensor plug to the controller CAUTION Personal injury hazard Always lay cables and hoses so that they are straight and do not pose a tripping hazard ...

Страница 16: ...re 6 Pin configuration Number Description Standard cable cable color Teflon cable cable color 1 12 VDC Brown Pink 2 Earth Black Gray 3 Data Blue Brown 4 Data White White 5 Shield Shield gray Shield gray 6 Guide 6 1 2 5 3 4 ...

Страница 17: ...PARAMETERS CLEAN INTERVAL RESPONSE TIME LOGGER INTERVAL 7 Go back to the MAIN MENU or the measurement mode display 4 3 Sensor data logger There is a data log and event log available for each sensor The data memory is used to store measurement data at preset intervals the event memory stores events such as configuration changes alarms and warning conditions Both logs can be exported to CSV format r...

Страница 18: ...corded values for all points POINT 1 Current calibration for point 1 POINT 2 Current calibration for point 2 POINT 3 Current calibration for point 3 SET CAL DEFLT Security prompt reset to default calibration CALIBRATE TS content SET OUTMODE Behavior of the outputs during calibration and zero point adjustment HOLD ACTIVE SET TRANSFER SELECTION SENSOR MEASURE Current uncorrected measurement value FA...

Страница 19: ...NFIG TRB CLEAN INTERVAL 15 min 30 min 1 h 4 h 12 h 1 day 3 days 7 days FACTORY CONFIG 4 h RESPONSE TIME 1 to 300 s DEFAULT CONFIG 60 s LOGGER INTERVAL 10 s 30 s 1 min 2 min 3 min 4 min 5 min 6 min 10 min 15 min 30 min FACTORY CONFIG 10 min SET DEFAULTS Security prompt reset to default configuration for all menu options listed above 4 4 2 SENSOR setup SELECT SENSOR if there is more than one sensor ...

Страница 20: ... FNU SOLID 0 001 to 500 g L MODEL NUMBER Item no Sensor CODE VERS Sensor software PROFILE PROFILE COUNTER Counter 20 000 backwards RESET CONFIG MANUAL RESET security prompt COUNTERS MANUAL RESET PRESS ENTER security prompt TEST MAINT COUNTER X DAYS BACKWARDS GASKET GASK COUNTER X DAYS BACKWARDS TOTAL OPERATING HOURS COUNTER MOTOR WIPE CYCLE COUNTER INTERVAL Default for maintenance counter SERVICE ...

Страница 21: ...URBIDITY TRB parameter Before the sensor can be calibrated to the TURBIDITY TRB parameter the parameter must be selected 4 5 1 1 Select the TURBIDITY TRB parameter 1 Open the MAIN MENU 2 Select SENSOR SETUP and confirm 3 Select the corresponding sensor and confirm 4 Select CONFIGURE and confirm 5 Select PARAMETERS and confirm 6 Select the TRB parameter and confirm 7 Go back to the MAIN MENU or the...

Страница 22: ... being used 8 Select the recorded POINT and confirm 9 Enter the laboratory comparison value and confirm To record more calibration points repeat steps 6 to 9 10 Go back to the MAIN MENU or the measurement mode display The instrument automatically sorts the saved calibration points according to the size of the calibration values irrespective of the sequence in which the calibration points were reco...

Страница 23: ...parison is made from a grab sample and not a known standard Once the calibration point has been recorded by the probe a mark is shown after the point or points that has have been recorded for approximately 3 seconds Note If the Calibrate menu is closed and then opened again before the calibration is complete the mark is shown again This indicates that the calibration for this point these points ha...

Страница 24: ...points for approximately 3 seconds 6 Select MEMORY and confirm 7 Select CLEAR MEMORY and confirm The sensor will continue working with the old calibration values 8 Go back to the MAIN MENU or the measurement mode display 4 5 3 2 Clear a calibration point An individual calibration point can be cleared at any time by entering the value 0 0 for the concentration 1 Open the MAIN MENU 2 Select SENSOR S...

Страница 25: ...e manufacturer CAUTION Personal injury hazard Only qualified personnel should conduct the tasks described in this section of the manual NOTICE The gaskets on the wiper shaft must be replaced yearly Failure to replace the gaskets regularly could result in moisture ingress into the sensor head and lead to irreparable damage to the device Maintenance task Maintenance interval Visual inspection Monthl...

Страница 26: ...ENSOR SETUP and confirm 3 Select the corresponding sensor and confirm 4 Select TEST MAINT and confirm 5 Select PROFILE replace the wiper profile as described in Figure 7 Note Make sure that the wiper is within the tolerance range shown 6 Select RESET and confirm 7 Confirm MANUAL RESET ARE YOU SURE 8 Go back to the MAIN MENU or the measurement mode display Figure 7 Wiper replacement 1 Sensor 4 Scre...

Страница 27: ... requires another calibration point in a higher concentration ZERO Calibration is too close to the zero point Calibrate again with higher concentration CAL REQUIRED No existing calibration Calibrate probe EE RSRVD ERR Error in the probe electronics Call the manufacturer s customer service department ERROR PROBE Error in the probe electronics Call the manufacturer s customer service department LED ...

Страница 28: ...28 Troubleshooting ...

Страница 29: ...sor bracket including 90 adapter LZX414 00 10000 Consisting of Base ATS010 Mounting attachment HPL061 Holding clamp 2 LZX200 Assembly pipe 2 m BRO075 HS small parts set LZX416 1 8 m extension pipe LZY414 1 0 m extension pipe LZY413 Second attachment point including holding clamp LZX456 90 sensor adapter AHA034 Set of small parts for securing sensor LZX417 90 base ATS011 Retractable ball valve fitt...

Страница 30: ...0 Unprocessed welding connector for TSS XL sc LZU302 99 10000 DN65 welding connector for TSS XL sc LZU302 99 10010 DN80 welding connector for TSS XL sc LZU302 99 10020 DN100 welding connector for TSS XL sc LZU302 99 10030 DN125 welding connector for TSS XL sc LZU302 99 10040 DN150 welding connector for TSS XL sc LZU302 99 10050 DN200 welding connector for TSS XL sc LZU302 99 10060 DN250 welding co...

Страница 31: ...iscovery of the fault If the customer fails to notify the supplier the product is considered approved despite the defect Further liability for indirect or direct damages is not accepted If device specific maintenance or inspection work prescribed by the supplier is to be performed within the guarantee period by the customer maintenance or by the supplier inspection and these requirements are not m...

Страница 32: ...32 Warranty and liability ...

Страница 33: ...ekal Service 40025 Unsigned integer 1 R W Output mode Calibrate Outputmodesrv Service 40026 Unsigned integer 1 R W Output mode Service EDITED NAME Configuration 40027 String 8 R W Measurement location PROFILE COUNTER Configuration 40035 Unsigned integer 1 R W Profile counter SERIAL NUMBER Configuration 40036 String 6 R W Serial number CAL DATE Configuration 40042 Time2 2 R Date of factory calibrat...

Страница 34: ...ance 40079 Float 2 R Signal LED S6E1 S6E3 Maintenance 40081 Float 2 R Signal LED S6E3 S5E2 Maintenance 40083 Float 2 R Signal LED S5E2 S5E4 Maintenance 40085 Float 2 R Signal LED S5E4 S6E2 Maintenance 40087 Float 2 R Signal LED S6E2 S6E4 Maintenance 40089 Float 2 R Signal LED S6E4 Table 3 Sensor Modbus registers Continued Tag name Group name Register Data type Length R W Description ...

Страница 35: ......

Страница 36: ...hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2009 2012 2018 All rights reserved Printed in Germany 09 2018 Edition 3 ...

Отзывы: