Option
Description
Export data log
Exports measurement data from the system to an external device. Preview data selection
before it is exported. Make sure that an external device is connected to the instrument
(e.g., a USB key, external hard drive, etc.).
Delete data log
Removes measurement data from the system. Previews data selection before it is
removed.
3.
Specify the data selection parameters. Push the left and right arrow keys to make a selection.
Push the up and down arrow keys to select an option.
Option
Description
Result type
Sets the type of result available.
Application
Sets the available applications.
Date
Sets the date range.
Operator
Sets the available operators.
4.
If
View data log
was the selected option, push
View
to see the selected data.
• Push the up and down arrow keys to select a line of data and push
Detail
to see more data
• If
Electrode calibration
is selected as the
Result type
, push the left and right arrow keys to
select the sensor if more than one sensor is installed. Push the up and down arrow keys to
select a line of data, then push the left and right arrow keys to see related graphs
• If only one
Application
is selected push the up and down arrow keys to select a line of data
and push
Detail
to see more data, or push the left and right arrow keys to see related graphs
5.
If
Export data log
or
Delete data log
was the selected option, push
Preview
to see the selected
data, then push
Export
or
Delete
to start the procedure.
Purge
Use this procedure to remove air bubbles from the system. Refer to
on page 19 for instructions.
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired,
contact the manufacturer.
Clean the instrument
N O T I C E
Never use flammable or corrosive solvents to clean any part of the instrument. Use of these solvents can degrade
the environmental protection of the instrument and may void the warranty.
Clean the exterior surface with a moist cloth or with a mixture of water and mild detergent. Dry with a
soft cloth.
Clean the sensor
Refer to the documentation delivered with the sensor.
English
23
Содержание TitraLab AT1000 Series
Страница 10: ...10 English ...
Страница 24: ...Maintenance menu Refer to the Full User Manual which is available as a download from our website 24 English ...
Страница 32: ...32 Deutsch ...
Страница 77: ...Français 77 ...
Страница 144: ...144 Čeština ...
Страница 166: ...166 Dansk ...
Страница 188: ...188 Nederlands ...
Страница 232: ...232 Svenska ...
Страница 254: ...254 Suomi ...
Страница 299: ...Magyar 299 ...
Страница 321: ...Română 321 ...
Страница 343: ...Türkçe 343 ...
Страница 365: ...Slovenský jazyk 365 ...
Страница 387: ...Slovenski 387 ...
Страница 409: ...Hrvatski 409 ...
Страница 431: ...Ελληνικά 431 ...
Страница 447: ......