Programming the Sampler for Level Control

4.9 Programming the Sampler for Level Control

Put the sampler in a standby state (i.e. "Program Halted", "Ready To Start" or 
"Program Complete") hold 

CLEAR ENTRY

 for two seconds. This will cause the 

display to read 

"

Level Control?". The following messages may be displayed in 

the level control program:

Note: 

In addition to the liquid level actuator, a Multi-purpose Half cable (Cat. No. 941 

or 541) may be used to connect the sampler to any controlling device which 
produces a dry contact closure upon a predetermined conditions. For example, a 
pH meter with adjustable high/low set points can be used to control sampler 
operation in any of the three level/auxiliary control modes. 

4.10 Other Displayed Messages

Item 

Number

Message

Description of Message

1

LEVEL CONTROL

If you press 

NO

SPECIAL OUTPUT?

 will appear on the display. If you press 

YES

, the 

display shows item 2.

2

LEVEL START?

If you press 

YES

, the sampler operates as described in Mode 1. If you press 

NO

, the 

display shows item number 4.

3

LEVEL START/STOP

If you press 

YES,

 the sampler operates as described in Mode 2. If you press 

NO

, the 

display shows item number 5.

4

SAMPLE ON LEVEL?

If you press 

YES, 

the sampler operates as described in Mode 3. If you press 

NO

, the 

display shows item number 1.

Message

Description of Message

YES/NO or * TO 

BACKUP

This message appears if a key other than 

YES

 or 

NO

 is pressed and the displayed message is 

followed by a question mark while in the Parameter Entry, Programming Running, or Time Set 
modes.

WHILE RUNNING, KEY 

0 - 1 - 2 ONLY

This message appears if a key other than 

0

1

, or 

2

 is pressed while the program is running. The 

program must be halted (press the 

2

 key) before any manual operations are performed, such as 

Bottle Advance, Pump, Purge, Take Sample, New Program, etc.

REVIEW RESULTS?

This message appears if 

DISPLAY FEEDBACK

 is pressed and the sample program has been halted. 

YES

 response will cause the sample program results to appear. A 

NO

 response causes 

REVIEW 

PROGRAM?

 to appear.

REVIEW PROGRAM?

If you press the 

YES

 key, all programmed settings will appear on the display. Pressing the 

NO

 key 

will place the sampler back into the standby state.

LOW BATTERY

This message indicates that the external battery is low and requires recharging.

MEMORY POWER LOW

This message indicates that the internal lithium battery is low and needs to be replaced.

PROGRAM STOPPED 

DUE TO FULL BOTTLE

During composite mode, this message appears upon a full bottle condition. The program cannot 
be restarted until the composite container is emptied and the float switch is in the lowered position.

BOTTLE IS FULL

If your instrument is programmed for single container composite sampling, this message appears 
if the 

START PROGRAM

 key is pressed and the float switch is not in the lowered position. Because 

the same circuit path is used for the composite full bottle float switch (composite mode) and 
#1 bottle indicator (multiple bottle mode), this message will also appear if the 

START PROGRAM

 key 

is pressed with the sampler programmed for single container composite with the distributor 
assembly in place, and the arm in the #1 bottle position. To correct this situation, remove the 
distributor, install the composite container retainer, and press the 

START PROGRAM

 key.

FULL BOTTLE or 

PROGRAM HALTED 

DUE TO FULL BOTTLE

The full bottle float switch is in the raised position due to a full composite container. To resume the 
program, exchange the full container with an empty one. With the float switch in the lowered 
position, the program will resume within one minute.

Содержание SIGMA 900

Страница 1: ...Catalog Number 8544 Sigma 900 Composite Sampler INSTRUMENT MANUAL Hach Company 2003 All rights reserved Printed in the U S A eac 07 03 2ed...

Страница 2: ...Visit http www hach com...

Страница 3: ...ng the Intake Line 14 3 4 1 Vinyl Intake Tubing 14 3 4 2 Teflon Lined Polyethylene Intake Line 14 3 5 Setting Up the Intake Line and Strainer 15 3 6 Choosing Bottles 15 3 7 Installing the Full Bottle...

Страница 4: ...34 5 8 Internal Desiccant Module 35 5 9 Circuit Board Identification 35 5 10 Motor Gear Box 36 5 11 Internal Case Humidity Indicator 36 5 12 Internal Case Humidity Indicator 36 5 13 Memory Battery 36...

Страница 5: ...on membrane switch keypad with 18 character alphanumeric liquid crystal display Self prompting menu driven program Internal Clock Indicated real time and date 0 007 time base accuracy Programmable Del...

Страница 6: ...Visit http www hach com...

Страница 7: ...es a high speed peristaltic pump for collection of the sample liquid The pump body is constructed of impact and corrosion resistant plastic Tubing is held firmly in place by the pump body haves no cla...

Страница 8: ...gram settings 2 3 2 Humidity Indicator The internal case humidity indicator Cat No 2660 turns pink when the internal case humidity exceeds 60 percent The sampler is equipped with an internal desiccant...

Страница 9: ...ts Failure to do so could result in serious injury to the operator or damage to the equipment To ensure the protection provided by this equipment is not impaired do not use or install this equipment i...

Страница 10: ...cations must be per the manufacturer s control drawing specifications Any modification to the instrumentation or the installation is not recommended and may result in life threatening injury and or da...

Страница 11: ...Electrical Code speziell gepr ft und zugelassen Dies gilt jedoch keineswegs f r das gesamte Hach Produktangebot Die Entscheidung ob ein Produkt f r den Einsatz in explosionsgef hrden Bereichen geeigne...

Страница 12: ...or fibers Toxic materials which upon contact or inhalation could result in injury impairment of health or death Confined spaces are not designed for human occupancy They have restricted entry and con...

Страница 13: ...50 C 32 to 120 F Follow the guidelines below and Figure 3 to allow complete drainage of the intake line and prevent cross contamination between samples Install the sampler as close to the source as si...

Страница 14: ...n to the stainless steel fitting connector end that points away from the pump control housing 3 4 2 Teflon Lined Polyethylene Intake Line The Connection Kit for Teflon lined Tubing Cat No 2186 contain...

Страница 15: ...could freeze and plug the line Place the sample intake and strainer in the mainstream of the sampling source in an area of turbulent and well mixed flow Also take in account the vertical location of t...

Страница 16: ...y and nickel cadmium Ni Cad battery Cat No 1416 The 120 V ac to 12 V dc power converter Cat No 1440 includes a built in lead acid battery charger The Power Converter Cat No 1443 contains a charger for...

Страница 17: ...y interface connector is a general purpose input output port Each signal is described in Table 2 Table 1 12 V dc Connector Pin Assignments Pin Signal Description A ground B 12 to 17 V dc unregulated T...

Страница 18: ...ace unit is available Cat No 2021 which converts a 4 20 mA flow meter output into a 12 V dc pulses At 20 mA 100 flow rate the interface sends ten pulses per minute at 4 mA 0 flow rate the interface se...

Страница 19: ...Interface This splitter interface Cat No 939 can be used if access to more than one of the Auxiliary Receptacle functions is required simultaneously Connecting the interface to the 6 pin connector on...

Страница 20: ...Visit http www hach com...

Страница 21: ...iew of the program CHANGE HALT Stops sampler program and places in standby state When started press this key prior to making any program change and before pressing keys 4 5 6 7 8 9 and NEW PROGRAM RES...

Страница 22: ...ete or Ready to Start The program status changes to Program Complete when sampling is complete and there are no samples to be taken or the optional full bottle shut off is actuated When the program is...

Страница 23: ...7 After the correct year has been entered the display will read Synchronize Time enter at Time Pressing YES will start the clock The display will read Clock is Now Set 4 4 Bottle Setup Mode When prog...

Страница 24: ...s To determine the appropriate number refer to the programming instructions for flow proportional operation in section 4 11 9 OTHER CHANGES Press NO to retain the previously entered program entries fo...

Страница 25: ...art Program Halted or Program Complete 2 To operate the pump press either the PUMP or PURGE key This will cause Manual Mode Pump Purge Clear to appear on the display Pressing the PUMP key causes the p...

Страница 26: ...terval automatically resets to the programmed interval If the liquid falls below the predetermined level the sampler continues to collect samples at the programmed time or flow interval until the prog...

Страница 27: ...ted press the 2 key before any manual operations are performed such as Bottle Advance Pump Purge Take Sample New Program etc REVIEW RESULTS This message appears if DISPLAY FEEDBACK is pressed and the...

Страница 28: ..._ _ CNTS during setup 1 Determine f where f represents the flow increment between samples and n represents the total number of samples to be collected 2 Multiply the flow increment between samples f b...

Страница 29: ...les over a given period of time when sampling in proportion to the flow rate To calculate the number of 12 V dc output pulses between samples determine the total number of samples represented by n in...

Страница 30: ...low proportional basis Average Flow Rate 1 75 mgd average flow rate over the 24 hour period Maximum Flow Rate 3 mgd a 1440 minutes 1 Calculate Q 2 Calculate t 3 Multiply Q x t Therefore the value ente...

Страница 31: ...rate over the 48 hour period Maximum Flow Rate 1 cfs a 2 880 minutes 1 Calculate Q 2 Calculate t 3 Multiply Q x t Therefore the value entered for INTV _ _ _ _ CNTS is 24 You will notice from the prev...

Страница 32: ...Visit http www hach com...

Страница 33: ...L MODE soft key 5 1 4 No Lubrication Required The 900 Composite Sampler does not require routine lubrication Do not lubricate the pump tube with petroleum jelly silicone grease oil or any other lubric...

Страница 34: ...g into the wrong connector can cause extensive damage Note The front panel gasket has a light coating of grease to help assure a watertight seal Do not contaminate the grease or gasket area with dirt...

Страница 35: ...desiccant module is located under the CPU board Note The desiccant module cannot be recharged by heating Do not attempt to bake the desiccant module in an oven to remove the moisture as this could be...

Страница 36: ...s power at all times If power is removed the data stored in the RAM chip is lost Therefore you cannot power the RAM chips from the 900 Composite Sampler power supply because you would lose data and pr...

Страница 37: ...In the U S A call 1 800 635 1230 Outside the U S A and Europe send E mail to intl hach com Repair Service Authorization must be obtained from Hach Company before sending any items for repair Please c...

Страница 38: ...Hvidovre Tel 45 0 36 77 29 11 Fax 45 0 36 77 49 11 info hach lange dk The Netherlands Dr Lange Nederland B V Laan an Westroijen 2a NL 4003 AZ Tiel Tel 31 0 3 44 63 11 30 Fax 31 0 3 44 63 11 50 info h...

Страница 39: ...r improper application or installation Damage caused by any repair or attempted repair not authorized by Hach Company Any product not used in accordance with the instructions furnished by Hach Company...

Страница 40: ...Visit us at www hach com...

Страница 41: ...el 6 0 in long x 0 406 in OD 2071 Strainer All 316 Stainless Steel 7 94 in long x 1 0 in OD 2070 Strainer Stainless Steel 3 9 in long x 0 406 in OD 4652 Strainer Teflon Stainless Steel 5 5 in long x 0...

Страница 42: ...Visit http www hach com...

Страница 43: ...Set Hour Minute AM PM Month Day Year Timed Mode NO Flow Mode NO YES Set Time Interval Between Samples YES Other Changes YES Composite Mode YES NO Set up Complete Ready To Start Continuous Mode NO YES...

Страница 44: ...Visit http www hach com...

Страница 45: ...the electrical capacity that can be taken out of a battery decreases However excessive heat destroys batteries Avoid placing batteries near heat sources of any kind To maximize battery life operated...

Страница 46: ...erature the self discharge rate of nickel cadmium batteries can be as high as 2 per day When charged cells have been stored for a long period of time or at an elevated temperature a change starts to t...

Страница 47: ...rawings Appendix C Exploded View Drawings Exploded View 1 of 6 Item Number Description QTY Part Number 1 Pump Assembly 1 1075 2 Overlay 1 906 3 Pump Tube Insert 1 2441 4 Hex Nut 4 SE 403 5 Threadlock...

Страница 48: ...engdraw fm Appendix C Exploded View 2 of 6 Item Number Description QTY Part Number 1 Latch 1 2279 2 Keypad Cable Assembly 1 8586 3 Screw 6 SE 340 4 CPU PCB Assembly 1 8921 5 900 Controller Cover 1 861...

Страница 49: ...229 6 Threadlock 2 SE 957 21 Slow Blow Fuse 5A 250 V 1 8753 7 Ultrasonic Sensor Gasket 1 8609 22 O ring 1 3321 8 Retainer Plate 1 2548 23 Fuse Holder 5 x 20 mm 1 3320 9 Retainer Plate Gasket 1 2550 2...

Страница 50: ...r Item Number Description QTY Part Number 1 5 Amp Fuse Label 1 8537 6 Hex Nut tighten to 16 in lb 17 SE 301 2 Overlay 1 8807 7 Distributor Cable Assembly 1 8539 3 Screw 1 SE 343 8 900 Controller Cover...

Страница 51: ...nt Card Bracket 1 2594 4 Internal Power Cable Assy 1 8917 15 Humidity Indicator Card 1 2660 5 Connector Mounting Plate 1 2555 16 Retainer Plate Gasket 1 2557 6 Mounting Plate Gasket 1 2556 17 Retainer...

Страница 52: ...1 2 Distributor J5 3 Auxiliary J6 4 Pin 2 J3 5 Power J2 6 Cable Assembly Fuse Holder 3146 7 Fuse Holder 5 x 20 mm 3320 8 Gearbox Assembly 8910 9 Motor Cable Assembly 8859 10 Keypad J9 1 2 3 4 5 6 7 8...

Отзывы: