69
Basic programming setup
17-C.
Select a method for switching bottle sets. Select either clock time that switches
bottle sets every 24 hours or select duration that sets the bottles sets in minutes and
hours.
The number of bottles that were selected in bottles-per-sample mode are used as the
bottle set size. Two bottles per-sample means two bottles per Timed Bottle Set.
17-D.
Enable or Disable Continuous Mode using the
CHANGE CHOICE
key. If Continuous
Mode is enabled the program will continuously run and switch bottle sets after every
specified duration elapses, until manually stopped. If Continuous Mode is disabled,
then the sampling halts when the last bottle set in the tray is full.
17-E.
Press
ACCEPT
to continue to Upset Sample.
Step 18 - Upset sample
Upset Sampling analyzes the collected samples to determine when samples meet or
exceed specified upset limits. The sampler will swing the distributor arm around to a
sequestered set of bottles and grab an upset bottle.
The upset sample bottles are always the last bottles in the tray. For example, if there are
24 bottles in the tray, 4 first flush bottles and 4 upset bottles, then the bottles assignment
is 1–4 first flush, 5–20 main program, and 21–24 upset samples.
Note: Unlike Setpoint Sampling, Upset Sampling can be enabled while the sampler is performing a
regular sampling program. There must be more than one bottle in the sampler to perform Upset
Sampling.
Upset Samples are collected even when there are no more regular samples to be taken;
when the status screen says Program Complete, but the bottom line of the display says
Program Running.
Refer to the
Quick Start Guides on page 123
for more setup information for individual
channels.
18-A.
Highlight Upset Sampling using the
UP
and
DOWN
keys on the Advanced Sampling
Menu. Press
SELECT
to continue.
18-B.
Enable or Disable Upset Sampling using the
CHANGE CHOICE
key. Press
ACCEPT
to continue.
11:00 AM 21 - APR - 01
TIMED BOTTLE SETS
ACCEPT
USE CLOCK TIME OR
CHANGE
CHOICE
DURATION:
DURATION (hh:mm)
CANCEL
11:00 AM 21 - APR - 01
ADVANCED SAMPLING
SELECT
START/STOP TIMES
STORM WATER
TIMED BOTTLE SETS
UPSET SAMPLING
RETURN
VARIABLE INTERVALS
VARIABLE VOLUME