Basic programming setup
After three attempts and no sample liquid has been collected, the cycle is aborted, a
Missed Sample is reported to the Sample History Log, and the sampler begins a new
sample interval.
10-B.
Press
ACCEPT
to continue to Site ID.
Step 11 - Site ID
11-A.
Enter a site identification number of up to 8 digits. This Site ID will appear on all data
printouts. This feature is useful when multiple sites are monitored using a single flow
meter or if data readings from multiple flow meters are collected. Press
ACCEPT
.
11-B.
After Site ID is accepted the following menu will appear:
11-C.
If
NO
, the basic setup program is complete.
Note: To return to the Advanced Sampling options at a later time, from the Main Menu select,
SETUP, MODIFY SELECTED ITEMS, PROGRAM ENTRIES.
4.
Highlight Advanced Sampling and press
SELECT
. In the Advanced Sampling menu
highlight the an item using the
UP
and
DOWN
keys and press
SELECT
.
11-D.
If
YES
, continue to Advanced Sampling,
.
11:00 AM 21 - APR - 01
SITE ID
ACCEPT
SITE ID:
BACKUP
00000000
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01
ADVANCED SAMPLING
DO YOU WISH TO
ACCESS THE ADVANCED
SAMPLING FEATURES?
YES
NO
ENTER: 1 — 999
11:00 AM 21 - APR - 01
PROGRAM ENTRIES
SELECT
INTAKE RINSES
SAMPLE RETRIES
SITE ID
ADVANCED SAMPLING
RETURN