Troubleshooting
Appendix A
Troubleshooting
Problem
Possible Cause
Solution
Sampler control panel and
refrigerator inoperative.
Blown main fuse.
Replace fuse/reset circuit breaker.
Sampler disconnected from main power
source.
Reconnect.
Sampler control panel
inoperative, refrigeration is
okay.
Blown front panel fuse.
Replace fuse.
Control panel harness connector
unplugged.
Reconnect.
Fuse blowing.
Improper fuse size or type.
Replace fuse with 1 amp fuse
(Cat. No. 3318).
Bad transformer.
Check for shorted connections.
Faulty power supply or component.
Isolate and replace bad component.
Defective power ON/OFF switch.
Check switch for short. Replace if necessary.
Control panel operative, but
power ON does not light.
Bulb burned out.
Replace bulb.
Sampler does not work in
flow proportional mode
(4–20 mA option).
Polarity incorrect or wires not connected.
Check for proper polarity. Check wiring for good
non-resistant connections.
4–20 mA signal grounded in series loop.
Check for complete current loop. Check for loss
of signal due to grounding.
Voltage level too high on 4–20 mA.
Voltage on 4–20 loop must not exceed
15 V dc—otherwise damage may occur.
EXT-MIN toggle switch is in wrong position.
Place toggle switch in left EXT position.
Sampler does not respond
to contact closure signal
from flow meter.
A voltage or ground is present on contact
closure input leads.
Check input wires for presence of voltage or
ground, the sampler counts on release of dry
contact closure.
Interval Between Samples thumbwheel
switch is set on a higher number.
Double check the setting on 4-digit Interval
Between Samples thumbwheel switch.
EXT-MIN toggle switch is in wrong position.
Place toggle switch in left EXT position.
Flow meter not sending signal to sampler.
Check the signal and make sure it has a
minimum of 25 milliseconds duration.
Refrigerator not running.
Thermostat disconnected.
Check that the thermostat wire leads are
connected.
Compressor will not start.
Line disconnect switch open.
Close disconnect switch.
Fuse removed or blown.
Replace fuse.
Unit runs in short cycles.
High pressure cutout due to insufficient air
cooling.
Vacuum dust/debris from condenser coils.
Low pressure cut out due to:
a. Compressor valve leak.
b. System refrigerant undercharged.
c. Restriction in expansion device.
a. Replace valve.
b. Fix leak, add refrigerant.
c. Replace device.
Compressor runs long or
continuously.
Shortage of refrigerant due to leak.
Fix leak.
Control contacts are stuck together.
Defrost and make sure connecting tube between
sample chamber and sample container is in
place.
Dirty condenser.
Clean condenser.
Refrigerator is noisy while
operating.
Loose parts or mountings.
Tighten and secure loose parts and mounting.
Tubing rattle.
Reform tubing away from the obstruction.
Bent fan blade.
Replace fan.
Worn fan motor bearings.
Replace motor.
Содержание Sigma 1600
Страница 2: ...Visit http www hach com ...
Страница 12: ...Visit http www hach com ...
Страница 16: ...Page 16 Electrical Connections 6507installation fm Section 2 Figure 4 Sigma 1600 Automatic Sampler Wiring ...
Страница 17: ...Section 2 Page 17 6507installation fm Electrical Connections Figure 5 Control Panel Wiring ...
Страница 22: ...Visit http www hach com ...
Страница 32: ...Visit http www hach com ...
Страница 34: ...Visit http www hach com ...
Страница 36: ...Visit http www hach com ...
Страница 38: ...Visit http www hach com ...
Страница 44: ...Visit http www hach com Figure 14 Dipper Housing ...
Страница 46: ...Visit http www hach com ...
Страница 50: ...Visit http www hach com ...